Systematic Reviews
Copyright ©The Author(s) 2019.
World J Orthop. Oct 28, 2019; 10(11): 394-403
Published online Oct 28, 2019. doi: 10.5312/wjo.v10.i11.394
Table 1 Summary of patient and surgery characteristics for identified patients undergoing magnetically controlled growing rods instrumentation, n (%)
Items
Curve etiology
Idiopathic89 (32.8)
Congenital43 (15.9)
Syndromic68 (25.1)
Neuromuscular63 (23.2)
Neurofibromatosis8 (2.9)
Type of surgery
Primary195 (74.7)
Conversion66 (25.3)
Unspecified10
Type of instrumentation
Single rod64 (23.6)
Dual rod207 (76.4)
Table 2 Summary of hardware related complications following magnetically controlled growing rods instrumentation for early onset scoliosis
Complication rateWithout keeper plateWith keeper plate
Overall complication rate/study35.6% (n = 115)80.61% (n = 38)35.65% (n = 77)
Major complicationsn = 95n = 32n = 63
Cumulative Complications
Distraction failure14.0% (n = 38)40.8% (n = 20)8.1% (n = 18)
Implant failure8.86% (n = 24)18.36% (n = 9)6.76% (n = 15)
Screw pull-out8.12% (n = 22)4.1% (n = 2)9.0% (n = 20)
Infection2.2% (n = 6)2.04% (n = 1)2.25% (n = 5)
Prominent hardware2.58% (n = 7)14.28% (n = 7)0% (n = 0)
Proximal junctional kyphosis2.58% (n = 7)0% (n = 0)3.15% (n = 7)
Wound dehiscence0.74% (n = 2)0% (n = 0)0.9% (n = 2)
Table 3 Summary of articles included for analysis
First authorYrKeeper plate?# Of patientsPrimary surgeriesRevisions% MaleAge at surgery (yr)Curve magnitude
Hickey[8]2014Y84475%4.559.25
Akbarnia[6]2014N1212042%6.859
Lebon[4]2017Y3025553%9.166
Akbarnia[2]2013N1414050%8.8360
Thompson[17]2016Y1911853%9.162
Heydar[14]2017Y1616037.5%7.8362
Heydar[3]2016Y1818039%7.368
Yılmaz[18]2016Y85325%10.6---
Keskinen[16]2016Y50272338.4%55.2
Hosseini[15]2016N2315829.2%7.4555.35
La Rosa[21]2017Y1010050%7.264.7
Teoh[11]2016Y844---8.260
Rolton[22]2016Y21101152%7.854
Nnadi[23]2018Y1010050%6.257.7
Ridderbusch[5]2017Y2424033%8.963