World J Clin Pediatr. Sep 19, 2020; 9(2): 17-28
Published online Sep 19, 2020. doi: 10.5409/wjcp.v9.i2.17
Does carrier fluid reduce low flow drug infusion error from syringe size?
Zachary C Madson, Sitaram Vangala, Grace T Sund, James A Lin
Zachary C Madson, Pediatric Hospitalist Medicine, Lutheran Children's Hospital, Fort Wayne, IN 46804, United States
Sitaram Vangala, Medicine Statistics Core, David Geffen School of Medicine at UCLA, Los Angeles, CA 90095, United States
Grace T Sund, Department of Nursing, UCLA Mattel Children's Hospital, Los Angeles, CA 90095, United States
James A Lin, Department of Pediatrics, David Geffen School of Medicine at UCLA, Los Angeles, CA 90095, United States
Author contributions: Madson ZC and Lin JA conceived and designed and personally conducted all the experiments, made observations, analyzed results, and wrote the initial draft and revisions of the manuscript; Sund GT helped to conceive and design and interpret the experiments; Vangala S provided statistical analysis, analyzed results, and helped draft and revise portions of the manuscript. All authors reviewed and approved of the manuscript.
Supported by NIH National Center for Advancing Translational Sciences (NCATS) UCLA CTSI, No. UL1TR001881.
Institutional review board statement: This work was performed at UCLA Medical Center, Santa Monica and the UCLA Department of Medicine Statistics Core.
Conflict-of-interest statement: The authors report no financial relationships relevant to this article.
Data sharing statement: No additional data are available.
Open-Access: This article is an open-access article that was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution NonCommercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Corresponding author: James A Lin, BSc, MD, Assistant Professor, Attending Doctor, Department of Pediatrics, David Geffen School of Medicine at UCLA, 10833 Le Conte Ave, A2-383 MDCC, Los Angeles, CA 90095, United States. jameslin@mednet.ucla.edu
Received: April 30, 2020
Peer-review started: April 30, 2020
First decision: May 24, 2020
Revised: June 7, 2020
Accepted: August 31, 2020
Article in press: August 31, 2020
Published online: September 19, 2020
Abstract
BACKGROUND

Critically ill neonates and pediatric patients commonly require multiple low flow infusions. Volume limitations are imposed by small body habitus and co-morbidities like cardiopulmonary disease, renal failure, or fluid overload. Vascular access is limited by diminutive veins. Maintenance fluids or parenteral nutrition in conjunction with actively titrated infusions such as insulin, fentanyl, prostaglandins, inotropes and vasopressors may necessitate simultaneous infusions using a single lumen to maintain vascular catheter patency. This requirement for multiple titratable infusions requires concentrated medications at low flows, rather than more dilute drugs at higher flows that in combination may volume overload small infants.

AIM

To determine whether carrier fluid reduces variability that variability of low flow drug infusions is proportional to syringe size in pediatric critical care.

METHODS

We assessed concentrations of orange “drug” in a 0.2 mL/h low flow clinical model with blue dyed carrier fluid at 5 mL/h, using 3-, 10-, or 60-mL syringes. A graduated volumetric pipette was used to measure total flow. Mean time to target concentration was 30, 21, and 46 min in 3-, 10-, and 60-mL syringes, respectively (P = 0.42). After achieving target concentration, more dilute drug was delivered by 60-mL (P < 0.001) and 10-mL syringes (P = 0.04) compared to 3-mL syringes. Drug overdoses were observed during the initial 45 min of infusion in 10-and 60-mL syringes. Total volumes infused after target concentration were less in the 60-mL condition compared to 3-mL (P < 0.01) and 10-mL (P < 0.001) syringes.

RESULTS

Linear mixed effects models demonstrated lesser delivered drug concentrations in the initial 30 min by 3-mL compared to 10-and 60-mL syringes (P = 0.005 and P < 0.001, respectively) but greater drug concentrations and total infused drug in the subsequent 30-60 and 60-90 min intervals with the 3- and 10-mL compared to 60-mL syringes.

CONCLUSION

With carrier fluid, larger syringes were associated with significantly less drug delivery, less total volume delivered, and other flow problems in our low flow drug model. Carrier fluid should not be used to compensate for inappropriately large syringes in critical low flow drug infusions.

Keywords: Infusion pumps, Intensive care, Neonatal, Nursing research, Patient safety, Spectrophotometry, Syringes

Core Tip: Infusions of critical drugs in infants frequently require low flow rates. We previously observed errors in low flow infusions that were directly proportional to syringe size. Because low flow infusions in clinical practice are essentially always co-infused with a primary carrier fluid, we now use a similar model to test whether carrier fluid improves accuracy and flow continuity of low flow drug from large compared to smaller syringes. We report that despite carrier fluid, larger syringes were associated with less overall drug and fluid volumes delivered, worse flow continuity, and other flow problems in low flow infusions compared to smaller syringe sizes. Carrier fluid should not be used to compensate for errors introduced by syringe size in critical low flow drug infusions. Syringe size should be matched to the rate of infusion.