Basic Study Open Access
Copyright ©The Author(s) 2021. Published by Baishideng Publishing Group Inc. All rights reserved.
World J Orthop. Aug 18, 2021; 12(8): 548-554
Published online Aug 18, 2021. doi: 10.5312/wjo.v12.i8.548
Fibula pro-tibia vs standard locking plate fixation in an ankle fracture saw bone model
Tosan Okoro, Department of Arthroplasty, Robert Jones and Agnes Hunt Orthopaedic Hospital NHS Foundation Trust, Oswestry SY10 7AG, United Kingdom
Kar Hao Teoh, Department of Orthopaedic Surgery, Princess Alexandra Hospital NHS Trust, Harlow CM20 1QX, United Kingdom
Hiro Tanaka, Department of Orthopaedics, Royal Gwent Hospital, Newport NP20 2UB, United Kingdom
ORCID number: Tosan Okoro (0000-0003-0432-0936); Kar Hao Teoh (0000-0002-1538-8760); Hiro Tanaka (0000-0002-5693-983X).
Author contributions: Okoro T contributed study design, performance of biomechanical testing, analysis of results, drafting and proof reading of manuscript; Teoh KH contributed study design, analysis of results, drafting and proof-reading of manuscript; Tanaka H contributed study design, supervision of study, analysis of results, drafting and proof reading of manuscript.
Supported by Research Grant Provided by AOUK.
Institutional review board statement: This was a biomechanical study and did not involve human and the use of animals.
Conflict-of-interest statement: The authors have no conflict of interest to declare.
Data sharing statement: No additional data are available.
Open-Access: This article is an open-access article that was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution NonCommercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/Licenses/by-nc/4.0/
Corresponding author: Tosan Okoro, BSc, FRCS, MBChB, PhD, Doctor, Surgeon, Department of Arthroplasty, Robert Jones and Agnes Hunt Orthopaedic Hospital NHS Foundation Trust, Gobowen, Oswestry SY10 7AG, United Kingdom. tosanwumi@hotmail.com
Received: February 9, 2021
Peer-review started: February 9, 2021
First decision: June 25, 2021
Revised: July 2, 2021
Accepted: August 2, 2021
Article in press: August 2, 2021
Published online: August 18, 2021

Abstract
BACKGROUND

Locking plate fixation in osteoporotic ankle fractures may fail due to cut-out or metalwork failure. Fibula pro-tibia fixation was a technique prior to the advent of locking plates that was used to enhance stability in ankle fractures by achieving tri or tetra-cortical fixation. With locking plates, the strength of this fixation construct can be further enhanced. There is lack of evidence currently on the merits of tibia-pro-fibula augmented locking plate fixation of unstable ankle fractures.

AIM

To assess if there is increased strength to failure, in an ankle fracture saw bone model, with a fibula pro-tibia construct when compared with standard locking plate fixation.

METHODS

Ten osteoporotic saw bones with simulated supination external rotation injuries were used. Five saw bones were fixed with standard locking plates whilst the other 5 saw bones were fixed with locking plates in a fibula pro-tibia construct. The fibula pro-tibia construct involved fixation with 3 consecutive locking screws applied across 3 cortices proximally from the level of the syndesmosis. All fixations were tested in axial external rotation to failure on an electromagnetic test frame (MTS 858 Mini-Bionix test machine, MTS Corp, Eden Praire, MN, United States). Torque at 30 degrees external rotation, failure torque, and external rotation angle at failure were compared between both groups and statistically analyzed.

RESULTS

The fibula pro-tibia construct demonstrated a statistically higher torque at 30 degrees external rotation (4.421 ± 0.796 N/m vs 1.451 ± 0.467 N/m; t-test P = 0.000), as well as maximum torque at failure (5.079 ± 0.694N/m vs 2.299 ± 0.931 N/m; t-test P = 0.001) compared to the standard locking plate construct. The fibula pro-tibia construct also had a lower external rotation angle at failure (54.7 ± 14.5 vs 67.7 ± 22.9).

CONCLUSION

The fibula pro-tibia locking plate construct demonstrates biomechanical superiority to standard locking plates in fixation of unstable ankle fractures in this saw bone model. There is merit in the use of this construct in patients with unstable osteoporotic ankle fractures as it may aid improved clinical outcomes.

Key Words: Unstable ankle fractures, Pro-tibia fixation, Improved stability, Simulated biomechanical analysis, Osteoporotic fractures, Ankle injuries

Core Tip: Locking plate fixation in osteoporotic ankle fractures may fail due to cut-out or metalwork failure. This study compared a fibula pro-tibia construct to standard locking plate fixation in an ankle fracture saw bone model. The fibula pro-tibia construct demonstrated biomechanical superiority and there is merit to consideration of its use in patients with unstable osteoporotic ankle fractures.



INTRODUCTION

In osteoporotic bone, there is unsatisfactory fixation strength with uni-cortical cancellous fixation for distal fibula fractures[1,2], which can lead to loss of fixation as well as delayed or non-union[3]. The ways to try to obviate these risks include the use of locking, posterior plating, or non-locking constructs with adjunct fixation. One such example of the latter is the use of tri- or tetra-cortical fixation with fibula pro-tibia (syndesmotic) screws[1]. In comparison to the same construct without additional screws, fibula pro-tibia fixation has demonstrated a 9% increase in torque to failure, 24% increase ability to withstand external rotation, and a 34% increase in energy before failure of the construct[4]. This technique adds little operative time, is inexpensive, and is a technically straightforward method to increase the stability of the construct[4].

In unstable bi-malleolar ankle fractures, the talus remains attached to the lateral malleous[5]. Reducing the medial malleolus alone may prevent anatomical repositioning of the talus, as in some cases the lateral malleolus cannot be accurately reduced when it impinges on the proximal fibular fragment. Repositioning of the talus can be achieved by forcibly internally rotating the ankle in such cases, but this stretches the fibular collateral ligament. When external immobilization is discontinued, the lateral ligaments remain in a stretched position and slight to moderate talar instability, which predisposes to development of late degenerative arthritis, may be the result[5].

The lateral malleolus appears therefore to be the key to the anatomical reduction of displaced bi-malleolar fractures, and restoring the integrity of the lateral malleolus restores the integrity of the ankle[5].

Being able to maintain the integrity of the lateral malleolar fixation in osteoporotic bone in therefore important. This study aims to biomechanically assess whether there is an increased strength to failure with a fibula pro-tibia construct when compared with standard locking plate fixation for ankle fractures in an ankle fracture saw bone model.

MATERIALS AND METHODS
Materials

Ten osteoporotic saw bones (Pacific Research Laboratories Inc., Vashon, WA, United States) with simulated supination external rotation injuries were used in this study.

Fracture simulation

A lateral malleolar ankle fracture was simulated with an osteotomy at the lateral malleolus (oblique orientation, starting medially at the level of the tibial plafond), and extending distally and laterally at a 45° angle[4]. Figure 1 illustrates the simulated lateral malleolar fracture.

Figure 1
Figure 1 Simulated lateral malleolar sawbone ankle fracture.
Fracture fixation

Five of the lateral malleolar osteotomies were fixed in a standard fashion using a fibular locking plate (Stryker Variax locking plate; Stryker, Mahwah, NJ, United States; Figure 2A) whilst fibula pro-tibia fixation was utilized in the other 5 models [fibular locking plate (Stryker Variax locking plate; Mahwah, NJ, United States)] with tri-cortical fixation. Tri-cortical fibula pro-tibia fixation entailed the use of 3 consecutive fully threaded cortical 3.5mm locking screws placed proximal to the lateral malleolar osteotomy at the level of the tibio-fibular syndesmosis (Figure 2B).

Figure 2
Figure 2 Sawbone lateral malleolar fixation construct. A: Sawbone lateral malleolar fracture treated with standard locking plate; B: Sawbone lateral malleolar fracture treated with locking plate in fibula pro-tibia configuration.
Biomechanical testing

Each model was then subjected to biomechanical analysis after being mounted on a resin and tested on an electromagnetic test frame (MTS 858 Mini-Bionix test machine, MTS Corp, Eden Praire, MN, United States), Figure 3, with measurement of torque (N/m) at 30 degrees external rotation, maximum failure torque (N/m) and external rotation angle (°) at failure.

Figure 3
Figure 3 Fibula pro-tibia construct of ankle fracture sawbone model mounted on the electromagnetic test frame (MTS 858 Mini-Bionix test machine, MTS Corp, Eden Praire, MN, United States).
Statistical analysis

The student’s t-test was used to analyze differences between both groups with a P value < 0.05 taken as statistically significant.

RESULTS
Fibula pro-tibia vs standard locking plates (torque assessment)

The mean torque for the fibula pro-tibia constructs at 30° external rotation was 4.421 ± 0.796 N/m, which was significantly higher than that obtained for the standard locking plate fixation constructs 1.451 ± 0.467 N/m (t test P = 0.000). This difference was also noted in the maximum torque to failure (tibia-pro-fibula 5.079 ± 0.694 N/m vs standard locking plate fixation 2.298 ± 0.931 N/m; t test P = 0.001).

The torque values of each construct for the above parameters are detailed in Tables 1 and 2.

Table 1 Torque (N/m) recorded for fibula pro-tibia and standard locking plate constructs (torque at 30 degrees external rotation).
Construct
Fixation type
Torque (N/m)
1Fibula pro-tibia3.723
2Fibula pro-tibia5.692
3Fibula pro-tibia4.695
4Fibula pro-tibia4.043
5Fibula pro-tibia3.954
6Standard locking plate0.829
7Standard locking plate1.709
8Standard locking plate1.539
9Standard locking plate1.155
10Standard locking plate2.022
Table 2 Torque (N/m) recorded for fibula pro-tibia and standard locking plate constructs (maximum torque at failure).
Construct
Fixation type
Maximum torque at failure (N/m)
1Fibula pro-tibia4.270
2Fibula pro-tibia5.970
3Fibula pro-tibia5.176
4Fibula pro-tibia5.468
5Fibula pro-tibia4.513
6Standard locking plate1.187
7Standard locking plate2.869
8Standard locking plate1.519
9Standard locking plate2.497
10Standard locking plate3.422
Fibula pro-tibia vs standard locking plates (angle to failure)

There was a lower mean external rotation angle to failure for the fibula pro-tibia construct (54.7° ± 14.5°) compared to standard locking plate fixation 67.7° ± 22.9°, but this was not statistically significant; t test P = 0.313.

DISCUSSION

Open reduction and internal fixation for an unstable ankle fracture in young patients is relatively predictable with excellent outcomes[6]. However the management of ankle fractures in the elderly remains less predictable, secondary to the various comorbidities associated with elderly patients such as osteoporosis, diabetes, cardiovascular, and peripheral vascular disease[3].

A recent trial demonstrated superiority of tibio-talo-calcaneal (TTC) nailing over standard locking plate fixation in the elderly in terms of having a low risk of complications, an earlier return to previous level of mobility, and the allowance of an immediate return to full weight bearing[6]. A limitation to the use of the TTC nail in routine practice is the risk of proximal peri-prosthetic fractures, as well as the need for the availability of a senior trauma surgeon or foot and ankle specialist to obtain optimal outcomes[7].

This study demonstrates that the fibula pro-tibia locking plate construct has biomechanical superiority to standard locking plates in a saw-bone model. There is increased torque at 30 degrees external rotation as well as a higher torque at maximum failure of the construct. The reduced maximal external rotation angle at failure of the construct is most likely due to the increased rigidity of the fixation, which is potentially beneficial for osteoporotic bone.

A limitation of this study is that it was performed on sawbones not cadaveric bone. Such data may therefore not be readily transferable to a clinical scenario, as we have performed an isolated analysis of a lateral malleolar fracture. The data we have shown however gives an objective assessment of the difference in biomechanical properties between the two constructs. Another limitation of the study is that we have not used a model that accounts for a bi-malleolar fracture pattern. The lateral malleolus is key to the anatomical reduction of displaced bi-malleolar fractures, and restoring the integrity of the lateral malleolus restores the integrity of the ankle[5]. The use of fibula pro-tibia fixation in this study demonstrates that there is up to approximately 3 times the level of torque achieved at 30 degrees external rotation and twice the failure torque in comparison to standard locking plate fixation. Initiating use of the adjunctive technique whilst performing such fixation is inexpensive, adds little operative time, and is not technically demanding. We propose the 3, 3, 3 rule for use of this adjunctive technique; Fixation with 3 screws across 3 cortices starting 3 cm above the tibial plafond.

The fibula pro-tibia construct utilizes the combined pull-out strength of locking screws to ensure a more biomechanical stronger construct. By using a tricortical fixation, it also ensures that the fixation is not as rigid as non-locking tetracortical fixation and provides some syndesmosis micro movement. There is therefore not a need for removal before weight bearing of the patient.

The increased biomechanical strength of the fibula pro-tibia construct demonstrates that there is merit to its use in patients with unstable osteoporotic ankle fractures. Future research is required to evaluate if its use would aid improved clinical outcomes in this important group of patients.

CONCLUSION

This study compared a fibula pro-tibia construct to standard locking plate fixation in an ankle fracture saw bone model. The fibula pro-tibia construct demonstrated biomechanical superiority and there is merit to consideration of its use in patients with unstable osteoporotic ankle fractures.

ARTICLE HIGHLIGHTS
Research background

The lateral malleolus is key to the anatomical reduction of displaced bi-malleolar fractures, and restoring its structural integrity restores the integrity of the ankle. Various fixation techniques have been utilized in osteoporotic bone to ensure lateral malleolar integrity.

Research motivation

Biomechanical assessment of whether there is an increased strength to failure with a fibula pro-tibia construct when compared with standard locking plate fixation for ankle fractures in an ankle fracture saw bone model.

Research objectives

To compare a fibula pro-tibia construct to standard locking plate fixation in a saw bone model using biomechanical parameters.

Research methods

After simulation of supination/external rotation injuries in a series of n = 10 sawbones, n = 5 were fixed with the fibula pro-tibia construct and n = 5 were fixed with the standard locking plate. Biomechanical analysis was performed to assess torque (N/m) at 30 degrees external rotation, maximum failure torque (N/m) and external rotation angle (°) at failure. Students t test was used for comparison of both groups.

Research results

The fibula pro-tibia construct was biomechanically superior to the standard locking plate in torque at 30 degrees external rotation, and maximum failure torque. There was no statistically significant difference in the external rotation angle at failure.

Research conclusions

There is merit to considering the use of the fibula pro-tibia construct in fixation of bimalleolar ankle fractures in view of its biomechanical superiority over standard locking plates.

Research perspectives

Future research should evaluate the clinical significance of these findings.

ACKNOWLEDGEMENTS

We would like to thank King I and Lane H of Cardiff University School of Engineering for their assistance in providing access to the biomechanical testing equipment.

Footnotes

Manuscript source: Invited manuscript

Specialty type: Orthopedics

Country/Territory of origin: United Kingdom

Peer-review report’s scientific quality classification

Grade A (Excellent): 0

Grade B (Very good): 0

Grade C (Good): C

Grade D (Fair): 0

Grade E (Poor): 0

P-Reviewer: Liu J S-Editor: Gao CC L-Editor: A P-Editor: Xing YX

References
1.  Koval KJ, Petraco DM, Kummer FJ, Bharam S. A new technique for complex fibula fracture fixation in the elderly: a clinical and biomechanical evaluation. J Orthop Trauma. 1997;11:28-33.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 58]  [Cited by in F6Publishing: 51]  [Article Influence: 1.9]  [Reference Citation Analysis (0)]
2.  McKean J, Cuellar DO, Hak D, Mauffrey C. Osteoporotic ankle fractures: an approach to operative management. Orthopedics. 2013;36:936-940.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 31]  [Cited by in F6Publishing: 30]  [Article Influence: 2.7]  [Reference Citation Analysis (0)]
3.  Beauchamp CG, Clay NR, Thexton PW. Displaced ankle fractures in patients over 50 years of age. J Bone Joint Surg Br. 1983;65:329-332.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 135]  [Cited by in F6Publishing: 139]  [Article Influence: 3.4]  [Reference Citation Analysis (0)]
4.  Panchbhavi VK, Vallurupalli S, Morris R. Comparison of augmentation methods for internal fixation of osteoporotic ankle fractures. Foot Ankle Int. 2009;30:696-703.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 36]  [Cited by in F6Publishing: 31]  [Article Influence: 2.1]  [Reference Citation Analysis (0)]
5.  Yablon IG, Heller FG, Shouse L. The key role of the lateral malleolus in displaced fractures of the ankle. J Bone Joint Surg Am. 1977;59:169-173.  [PubMed]  [DOI]  [Cited in This Article: ]
6.  Georgiannos D, Lampridis V, Bisbinas I. Fragility fractures of the ankle in the elderly: Open reduction and internal fixation vs tibio-talo-calcaneal nailing: Short-term results of a prospective randomized-controlled study. Injury. 2017;48:519-524.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 50]  [Cited by in F6Publishing: 58]  [Article Influence: 8.3]  [Reference Citation Analysis (0)]
7.  Baker G, Mayne AIW, Andrews C. Fixation of unstable ankle fractures using a long hindfoot nail. Injury. 2018;49:2083-2086.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 13]  [Cited by in F6Publishing: 15]  [Article Influence: 2.5]  [Reference Citation Analysis (0)]