TC-325 (Hemospray) is a mineral-based hemostatic powder that is increasingly used for nonvariceal upper gastrointestinal bleeding, particularly diffuse or ma
An 84-year-old man with metastatic jejunal mucinous adenocarcinoma presented with hematemesis and a hemoglobin of 7.5 g/dL. Upper endoscopy with push enteroscopy identified a 3-cm cratered gastric body ulcer with stigmata of recent bleeding (Forrest class IIc) and the known obstructing jejunal tumor. TC-325 was applied to the ulcer with the endoscope in partial retroflexion. On attempted withdrawal the endoscope was fixed: Hydrated powder had coated the insertion tube and bonded it to the mucosa at the gastroesophageal junction. Water irriga
TC-325 applied in retroflexion can bond the endoscope to the mucosa. Early recognition, avoidance of forceful traction, irrigation, controlled de-retroflexion and, when these fail, rescue with a parallel ultrathin endoscope can free the instrument safely.
Core Tip: Gastrointestinal bleeding can be challenging to treat even in the most experienced hands. The modalities available to gastroenterologists for achieving hemostasis have expanded over the years, and this area continues to evolve. Hemostatic powders are one such valuable innovation; they have become widely popular and are often used. However, awareness of unique risks, such as failure to withdraw the endoscope because of mucosal adherence, and knowledge of how to prevent and overcome this challenge are crucial to continued use.
- Citation: Gangu K, Daglilar E, Chela HK. Stuck scope: Tackling the adherent scope situation after hemostatic powder application: A case report and review of literature. World J Gastrointest Endosc 2026; 18(11): 125797
- URL: https://www.wjgnet.com/1948-5190/full/v18/i11/125797.htm
- DOI: https://dx.doi.org/10.4253/wjge.125797
Acute upper gastrointestinal (GI) bleeding remains a leading cause of hospital admission. Although mortality has de
Reported adverse events are uncommon and include catheter or working-channel blockage, transient biliary obstruction, perforation and, rarely, difficulty withdrawing the endoscope[8,13,14]. In May 2023 the manufacturer issued an urgent medical device correction warning that Hemospray powder may adhere to the distal endoscope and result in difficulty or inability to maneuver or remove the instrument, with potential for mucosal tear, perforation, hemorrhage and death; in some, but not all, reported instances the powder had been sprayed with the endoscope in a retroflexed position[15]. Retroflexion, routinely used to inspect and treat lesions tangential to the forward view such as the gastric cardia and fundus or the distal rectum[16], is itself low-risk but can cause perforation even in experienced hands[17,18].
Endoscope adhesion after TC-325 application has been described previously, including during retroflexion, and prior reports were resolved by traction or torque, prolonged irrigation, delayed removal after the powder had sloughed or, in one report, a fiberoptic bronchoscope advanced alongside the entrapped gastroscope (Table 1)[19-27]. The contribution of the present report is therefore not the complication itself but its management: After irrigation, lubrication, torque and forceps clearance through the entrapped endoscope had failed, the instrument was freed by advancing a second ultrathin gastroscope alongside it and stripping the adherent powder under direct vision. To our knowledge, the use of a second endoscope with a working channel, through which the adherent powder is stripped from the entrapped instrument under direct vision, has not previously been reported for powder-related entrapment. In the only prior report in which a second instrument was used, Crowley[23] freed an entrapped gastroscope with a fiberoptic bronchoscope advanced alongside it, dislodging the adhesive with gentle back-and-forth movements of the bronchoscope; a second endoscope was mentioned as a theoretical option in one pediatric report but was not attempted[25], and the double-endoscopy tech
| Ref. | Report type | Hemostatic agent | Indication/bleeding source | How the scope became stuck | How it was freed/managed | Complications/outcome |
| Smith et al[19] and Smith et al[20] | Conference abstract (SEAL prospective registry interim data, n = 82), with the same two events reported in the subsequent full registry publication (n = 63) | TC-325 | Nonvariceal upper GI bleeding (UGIB), various lesions | Endoscope became adherent to esophageal mucosa after application in retroflexion (2 patients) | Not detailed | Not detailed; listed as technical complications |
| Yii et al[21] | Full case report | TC-325 (Hemospray) | Bleeding gastric cardia malignancy (lymphoma) | Powder adhesion after application; endoscope retained at the gastroesophageal junction | Endoscope left in situ under intensive care and withdrawn after approximately 48 hours once the powder had sloughed | Prolonged retention (approximately 48 hours); mechanical ventilation/critical care |
| Shaikh and Parikh[22] | Conference abstract | TC-325 | Bleeding of gastric cardia and fundus | Adhesion after retroflexed application; endoscope retained approximately 6 hours | Traction after transfer to a tertiary center | Scope freed; further outcome not detailed |
| Crowley[23] | Full case report (anesthesia journal) | TC-325 | Oozing esophageal mass (nonvariceal UGIB; hemoglobin 6.1 g/dL) | Gastroscope could be neither advanced nor withdrawn after application and was found adherent to the esophagus; position of the gastroscope during spraying (retroflexion) not stated | Sedation stopped to protect the airway; fiberoptic video bronchoscope passed orally alongside the gastroscope, secretions suctioned and the adhesive dislodged from the gastroscope with gentle back-and-forth movements; freed in under 15 minutes | No complication; procedure completed; report written from the anesthesia (airway-management) perspective |
| Syed et al[24] | Full case report | TC-325 | Bleeding gastric GIST near the gastroesophageal junction (fundus in the abstract; lesser curve in the case description), treated in retroflexion | Powder coated the scope at the gastroesophageal junction and adhered to mucosa | Torque maneuvering plus withdrawal force (described as moderate in the abstract and as significant in the case description) | Scope freed; no recognized injury, although the authors could not exclude tissue trauma from the force applied; started on imatinib |
| Edwards et al[25] | Full case report (pediatric) | TC-325 | Pediatric upper GI bleeding (22-month-old) | Transient adhesion of the distal endoscope to the esophagus after diffuse application in retroflexion | Constant traction for several seconds with gentle side-to-side rotation; a second endoscope was considered but not feasible because of patient size | No esophageal trauma on re-intubation and no immediate or long-term adverse effects; normal repeat endoscopy at 3 months |
| Guzik et al[26] | Conference abstract | TC-325 | Retained endoscope, case series (upper GI bleeding); number of patients not stated in abstract | Powder adhesion | Not detailed (conference abstract) | Not detailed |
| Echeverria et al[27] | Full case report | TC-325 (with PuraStat gel) | Gastric cardia/fundus AVMs (arteriovenous malformations), upper GI bleeding, INR 7 (retroflexion) | Powder adhesion after retroflexed application | Aggressive water irrigation; scope removed after 90 minutes | Prolonged retention (90 minutes); no further GI bleeding documented; lost to outpatient follow-up |
| Present case | Full case report | TC-325 | Bleeding gastric body ulcer (partial retroflexion) | Insertion tube coated with hydrated powder and bonded to cardia/gastroesophageal junction mucosa after application in partial retroflexion | Irrigation, viscous lidocaine, torque and forceps through the entrapped scope failed; second ultrathin (XP) gastroscope advanced alongside and forceps used to clear the adherent powder under direct vision | Scope freed after 30 minutes; no mucosal injury; no clinically evident rebleeding during admission; discharged to hospice |
An 84-year-old man presented to the emergency department with several episodes of hematemesis as well as melena of a few days’ duration.
The patient reported several episodes of sudden-onset hematemesis as well as melena of a few days’ duration. He was not on nonsteroidal anti-inflammatory drugs and he was not taking antiplatelet or anticoagulant therapy.
He had a history of colon cancer (in the remote past, with details unknown) and of jejunal mucinous adenocarcinoma diagnosed a few months earlier. At that time he had undergone exploratory laparotomy, which revealed a large jejunal mass and extensive peritoneal carcinomatosis with metastatic deposits up to 10 cm; an internal bypass from the jejunum to the mid-transverse colon had been created to relieve obstruction. He had subsequently started palliative chemotherapy with FOLFOX and had received one cycle.
He had no history of tobacco or alcohol use. No known family history of GI malignancy or bleeding diathesis.
On admission the patient was tachycardic (110 beats/minute) and hypotensive (83/64 mmHg). He appeared pale on evaluation and abdominal examination revealed mid-abdominal and epigastric tenderness.
Hemoglobin was 7.5 g/dL compared with a recent baseline of 9 g/dL. The platelet count of 405 × 103/μL and international normalized ratio of 1.1 were unremarkable. He was also noted to have acute-on-chronic kidney injury (creatinine 1.4 mg/dL at baseline to 1.9 mg/dL). He received 3 units of packed red blood cells before endoscopy and was started on intravenous proton pump inhibitor therapy in the form of a loading dose 80 mg intravenously followed by continuous infusion at a rate of 8 mg/hour. The high-dose acid suppression therapy was initiated on presentation and continued for 48 hours post procedure before transitioning to 40 mg twice daily.
Computed tomography of the abdomen and pelvis showed no active contrast extravasation but demonstrated a large intra-abdominal mass involving the lumen of the duodenojejunal junction, consistent with the known malignancy (Figure 1).
The procedure was performed on the day of presentation to the hospital within about 7 hours of admission under moni
The decision to treat the gastric ulcer with TC-325 rather than, or in addition to, conventional injection, thermal or mechanical therapy was based on the following considerations: The lesion was large with a broad necrotic base and friable heaped-up margins suspicious for malignant infiltration, there was no single targetable vessel, and the patient's advanced malignancy and goals of care favored a rapid, non-contact approach. TC-325 was delivered through the Hemo
Immediately after application, attempted withdrawal met firm resistance and the endoscope could be neither advanced nor withdrawn. In the retroflexed view the insertion tube was coated with hydrated powder that bonded it to the mucosa of the cardia and gastroesophageal junction (Figure 6; the adhesion site is indicated by an arrow). Entrapment was recognized within a few minutes, forceful traction was deliberately avoided, insufflation was minimized, and the ane
A second, ultrathin gastroscope (Olympus GIF-XP190N) was then passed orally alongside the entrapped pediatric colonoscope. The ultrathin scope provided an independent view of the adhesion site from the esophageal side, and biopsy forceps passed through its working channel were used to strip the adherent powder from the insertion tube and the adjacent mucosa under direct vision (Figure 8). Care was taken to grasp powder rather than mucosa and to avoid contact with the lens and bending section of the entrapped instrument. Once the bond was cleared, the pediatric colono
The ultrathin gastroscope was then used to re-examine the entire esophagus, the gastroesophageal junction in forward and retroflexed views, and the stomach; there was no mucosal tear, laceration, hematoma or perforation, and the powder remained in place over the ulcer.
Bleeding gastric body ulcer, moderately differentiated adenocarcinoma on biopsy, in a patient with metastatic jejunal mucinous adenocarcinoma, complicated by TC-325-related endoscope entrapment during partial retroflexion. The patho
Treatment consisted of endoscopic hemostasis with TC-325 monotherapy, intravenous proton pump inhibitor therapy with continuous infusion for 72 hours, transfusion of a total of three units of packed red blood cells, and rescue of the entrapped endoscope as described above. No further endoscopic therapy was required.
The patient had no further hematemesis or melena, and hemoglobin remained stable at 10 g/dL over the remaining 5 days of hospitalization; no repeat transfusion was required. In view of his advanced malignancy he was discharged to home hospice on the last hospital day. Follow-up consisted of close monitoring with home hospice during which no clinically evident rebleeding was reported. Because hospice enrollment precluded surveillance endoscopy or routine laboratory testing, we characterize the outcome as absence of clinically evident rebleeding during the follow-up period rather than confirmed durable hemostasis.
TC-325 achieves hemostasis through the physical property that also makes it hazardous: On contact with moisture it transforms from a free-flowing powder into a cohesive, adherent mass[3,5]. The same coalescence that seals a bleeding surface can bind to the insertion tube, bending section or distal cap of the endoscope. Edwards et al[25] proposed that inactive powder layered onto the endoscope surface becomes activated when the straightened instrument is withdrawn against moist esophageal mucosa. In retroflexion this risk is amplified: The shaft is acutely angulated within a confined lumen, the tip and shaft lie in close apposition to the mucosa and to each other, and powder dispersed in a retroflexed position settles onto the scope rather than away from it. A layer of hydrating powder bridging the angulated scope and the surrounding wall can mechanically lock the instrument. Overapplication, spraying close to the shaft and aspiration of powder into the working channel (which can also block suction) probably compound the effect. In our case the endoscope was in partial rather than full retroflexion, indicating that partial retroflexion does not eliminate the risk. These obser
Endoscope entrapment after TC-325 has now been described in at least 10 patients across five full case reports and three conference abstracts (Table 1)[19-27]. Of these, eight patients are individually documented (two in the SEAL registry[19,20] and one in each of six single-patient reports[21-25,27]); the remaining conference abstract describes a case series without stating the number of patients[26], so the total of at least 10 assumes a minimum of two patients in that series. The complication also appears among events reported to the MAUDE database[13,14] and in the manufacturer’s 2023 device correction[15]. The reported rescues have relied on traction or torque[22,24,25], prolonged irrigation for up to 90 minutes[27], or deferred removal under intensive care after approximately 48 hours once the powder had sloughed[21], with attendant airway and critical-care implications when an upper endoscope cannot be withdrawn[23]. In the anesthe
It is important to separate the manufacturer's instructions, published evidence and our own inference. The instructions for use state that when spraying in the retroflexed position the powder may adhere to the outside of the endoscope, and the 2023 device correction adds that adhesion may also occur in the esophagus or stomach and that providers should be prepared to take immediate steps to manage it[15]. Product labeling and technical reviews advise delivering the powder in short bursts from a short distance (approximately 1-2 cm) without contact between the catheter and the target[3,8]. The published case reports support limiting the amount of powder applied in retroflexion and near the gastroesophageal junction, moistening any pale, inactivated powder visible on the endoscope before withdrawal, and treating in a straigh
We propose the following sequence for an endoscope that will not withdraw after TC-325 application, drawn from the prior reports, the manufacturer’s notice and our experience; except where referenced, it represents expert opinion rather than tested evidence: (1) Recognize entrapment immediately and stop; forceful traction risks mucosal tear, perforation and, when the endoscope is in the esophagus, airway compromise[15,23]; (2) Alert the anesthesia team at once so that the airway can be secured or maintained and the patient kept still and comfortable for what may be a prolonged rescue[23]; (3) Reduce insufflation to decompress the lumen (our suggestion); (4) Apply copious water irrigation at the interface; the powder does not truly dissolve, but hydration and mechanical loosening allowed release after prolonged irrigation in one report[27] and were suggested as a countermeasure in another[25]; (5) Apply gentle torque and attempt controlled de-retroflexion under direct vision, as reported by Syed et al[24], rather than abrupt pulling; (6) Attempt mechanical clearance of the bond with forceps through the entrapped endoscope; (7) If these fail, advance a second ultrathin or pediatric endoscope alongside the entrapped instrument to clear the bond under direct vision, as in the present case, or advance a fiberoptic bronchoscope alongside the gastroscope to dislodge the adhesive, as reported by Crowley[23]; con
This complication should be weighed against the independent, low but real perforation risk of retroflexion[17,18] and against the established efficacy and generally favorable safety profile of TC-325 across upper, lower and malignant GI bleeding[8-12]. The lesson is not to avoid the powder but to apply it with deliberate technique and to recognize entrap
TC-325 applied with the endoscope in retroflexion, even partial retroflexion, can bond the instrument to the mucosa at the gastroesophageal junction. The practical learning points are early recognition, avoidance of forceful traction, prompt involvement of anesthesia, copious irrigation and controlled de-retroflexion under direct vision, with a second ultrathin endoscope advanced alongside the entrapped instrument as a rescue option when simpler measures fail.
| 1. | Lanas A, Dumonceau JM, Hunt RH, Fujishiro M, Scheiman JM, Gralnek IM, Campbell HE, Rostom A, Villanueva C, Sung JJY. Non-variceal upper gastrointestinal bleeding. Nat Rev Dis Primers. 2018;4:18020. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 140] [Cited by in RCA: 126] [Article Influence: 15.8] [Reference Citation Analysis (8)] |
| 2. | Wuerth BA, Rockey DC. Changing Epidemiology of Upper Gastrointestinal Hemorrhage in the Last Decade: A Nationwide Analysis. Dig Dis Sci. 2018;63:1286-1293. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 236] [Cited by in RCA: 206] [Article Influence: 25.8] [Reference Citation Analysis (11)] |
| 3. | Changela K, Papafragkakis H, Ofori E, Ona MA, Krishnaiah M, Duddempudi S, Anand S. Hemostatic powder spray: a new method for managing gastrointestinal bleeding. Ther Adv Gastroenterol. 2015;8:125-135. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 36] [Cited by in RCA: 35] [Article Influence: 3.2] [Reference Citation Analysis (1)] |
| 4. | Laine L, Barkun AN, Saltzman JR, Martel M, Leontiadis GI. ACG Clinical Guideline: Upper Gastrointestinal and Ulcer Bleeding. Am J Gastroenterol. 2021;116:899-917. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 476] [Cited by in RCA: 441] [Article Influence: 88.2] [Reference Citation Analysis (4)] |
| 5. | Sung JJ, Luo D, Wu JC, Ching JY, Chan FK, Lau JY, Mack S, Ducharme R, Okolo P, Canto M, Kalloo A, Giday SA. Early clinical experience of the safety and effectiveness of Hemospray in achieving hemostasis in patients with acute peptic ulcer bleeding. Endoscopy. 2011;43:291-295. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 194] [Cited by in RCA: 159] [Article Influence: 10.6] [Reference Citation Analysis (7)] |
| 6. | Gralnek IM, Stanley AJ, Morris AJ, Camus M, Lau J, Lanas A, Laursen SB, Radaelli F, Papanikolaou IS, Cúrdia Gonçalves T, Dinis-Ribeiro M, Awadie H, Braun G, de Groot N, Udd M, Sanchez-Yague A, Neeman Z, van Hooft JE. Endoscopic diagnosis and management of nonvariceal upper gastrointestinal hemorrhage (NVUGIH): European Society of Gastrointestinal Endoscopy (ESGE) Guideline - Update 2021. Endoscopy. 2021;53:300-332. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 450] [Cited by in RCA: 392] [Article Influence: 78.4] [Reference Citation Analysis (16)] |
| 7. | Mullady DK, Wang AY, Waschke KA. AGA Clinical Practice Update on Endoscopic Therapies for Non-Variceal Upper Gastrointestinal Bleeding: Expert Review. Gastroenterology. 2020;159:1120-1128. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 145] [Cited by in RCA: 127] [Article Influence: 21.2] [Reference Citation Analysis (6)] |
| 8. | Gralnek IM, Bhandari P, Alkandari A, Alali A, Haidry RJ, Papaefthymiou A, Radaelli F, Subramaniam S, Fuccio L. Topical hemostatic agents in endoscopy: European Society of Gastrointestinal Endoscopy (ESGE) Technical and Technology Review. Endoscopy. 2025;57:1150-1173. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 11] [Cited by in RCA: 17] [Article Influence: 17.0] [Reference Citation Analysis (5)] |
| 9. | Alali AA, Pittayanon R, Martel M, Martins BC, Almadi MA, Chen YI, Barkun AN. TC-325 Superiority in Malignant Gastrointestinal Bleeding: An Individual Patient Data Meta-Analysis of Randomized Trials. Am J Gastroenterol. 2025;120:332-339. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 11] [Cited by in RCA: 11] [Article Influence: 11.0] [Reference Citation Analysis (0)] |
| 10. | Hussein M, Alzoubaidi D, O'Donnell M, de la Serna A, Bassett P, Varbobitis I, Hengehold T, Ortiz Fernandez-Sordo J, Rey JW, Hayee B, Despott EJ, Murino A, Graham D, Latorre M, Moreea S, Boger P, Dunn J, Mainie I, Mullady D, Early D, Ragunath K, Anderson J, Bhandari P, Goetz M, Kiesslich R, Coron E, Rodriguez de Santiago E, Gonda T, Gross SA, Lovat LB, Haidry R. Hemostatic powder TC-325 treatment of malignancy-related upper gastrointestinal bleeds: International registry outcomes. J Gastroenterol Hepatol. 2021;36:3027-3032. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 16] [Cited by in RCA: 18] [Article Influence: 3.6] [Reference Citation Analysis (0)] |
| 11. | Rodríguez de Santiago E, Burgos-Santamaría D, Pérez-Carazo L, Brullet E, Ciriano L, Riu Pons F, de Jorge Turrión MÁ, Prados S, Pérez-Corte D, Becerro-Gonzalez I, Martinez-Moneo E, Barturen A, Fernández-Urién I, López-Serrano A, Ferre-Aracil C, Lopez-Ibañez M, Carbonell C, Nogales O, Martínez-Bauer E, Terán Lantarón Á, Pagano G, Vázquez-Sequeiros E, Albillos A; TC-325 Collaboration Project, Endoscopy Group of the Spanish Association of Gastroenterology. Hemostatic spray powder TC-325 for GI bleeding in a nationwide study: survival and predictors of failure via competing risks analysis. Gastrointest Endosc. 2019;90:581-590.e6. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 39] [Cited by in RCA: 44] [Article Influence: 6.3] [Reference Citation Analysis (8)] |
| 12. | Chen YI, Barkun A, Nolan S. Hemostatic powder TC-325 in the management of upper and lower gastrointestinal bleeding: a two-year experience at a single institution. Endoscopy. 2015;47:167-171. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 15] [Cited by in RCA: 28] [Article Influence: 2.5] [Reference Citation Analysis (15)] |
| 13. | Ahmed K, Abdallah M, Abbas D, Jaber F, Abdalla AO, Mohamed M, McDonald N, Hanson BJ, Bilal M. Analysis of Reported Adverse Events Related to Hemospray: An FDA MAUDE Database Study. J Clin Gastroenterol. 2024;58:402-406. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 6] [Cited by in RCA: 6] [Article Influence: 3.0] [Reference Citation Analysis (0)] |
| 14. | Ali H, Dahiya DS, Gangwani MK, Patel P. Analysis of Reported Adverse Events Related to Hemospray an FDA MAUDE Database Study. J Clin Gastroenterol. 2023;57:1070-1071. [RCA] [PubMed] [DOI] [Full Text] [Reference Citation Analysis (0)] |
| 15. | Cook Medical. Urgent: Medical device correction. Hemospray Endoscopic Hemostat (061-C). May 23, 2023. Available from: https://www.sages.org/wp-content/uploads/2023/05/061-C-Customer-Letter-signed.pdf. |
| 16. | Rex DK, Khashab M. Colonoscopic polypectomy in retroflexion. Gastrointest Endosc. 2006;63:144-148. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 29] [Cited by in RCA: 28] [Article Influence: 1.4] [Reference Citation Analysis (2)] |
| 17. | Quallick MR, Brown WR. Rectal perforation during colonoscopic retroflexion: a large, prospective experience in an academic center. Gastrointest Endosc. 2009;69:960-963. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 36] [Cited by in RCA: 37] [Article Influence: 2.2] [Reference Citation Analysis (0)] |
| 18. | Katsinelos P, Kountouras J, Chatzimavroudis G, Zavos C, Pilpilidis I, Tzilves D, Paroutoglou G. Endoscopic closure of a large iatrogenic rectal perforation using endoloop/clips technique. Acta Gastroenterol Belg. 2009;72:357-359. [PubMed] |
| 19. | Smith LA, Stanley A, Morris J. OC-142 Hemospray for non-variceal upper gastrointestinal bleeding: results of the seal dataset (survey to evaluate the application of hemospray in the luminal tract). Gut. 2012;61:A61.2-A62. [DOI] [Full Text] |
| 20. | Smith LA, Stanley AJ, Bergman JJ, Kiesslich R, Hoffman A, Tjwa ET, Kuipers EJ, von Holstein CS, Oberg S, Brullet E, Schmidt PN, Iqbal T, Mangiavillano B, Masci E, Prat F, Morris AJ. Hemospray application in nonvariceal upper gastrointestinal bleeding: results of the Survey to Evaluate the Application of Hemospray in the Luminal Tract. J Clin Gastroenterol. 2014;48:e89-e92. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 99] [Cited by in RCA: 89] [Article Influence: 7.4] [Reference Citation Analysis (7)] |
| 21. | Yii RSL, Chuah KH, Poh KS, Lau PC, Ng KL, Ho SH, Mahadeva S. Retained Endoscope: An Unexpected but Serious Complication of Hemospray®. Dig Dis Sci. 2022;67:344-347. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 6] [Reference Citation Analysis (0)] |
| 22. | Shaikh A, Parikh ND. S2523 Stuck Scope: Failed Endoscopic Withdrawal After Hemospray. Am J Gastroenterol. 2022;117:e1678-e1679. [DOI] [Full Text] |
| 23. | Crowley JM. Hemostatic Spray TC-325 (Hemospray®) and Unexpected Esophageal Entrapment of Gastroscope With Airway Management Considerations: A Case Report. AANA J. 2023;91:349-352. [PubMed] |
| 24. | Syed AS, Vaughan AR, McCarthy JG, Laczek JT. Bleeding Gastrointestinal Stromal Tumor (GIST) Results in a Sticky Situation: The Good and Bad of Hemostatic Spray. Cureus. 2023;15:e46691. [RCA] [PubMed] [DOI] [Full Text] [Reference Citation Analysis (0)] |
| 25. | Edwards ST, Lerner DG, Alatorre Jimenez M, Attard TM. Don't Look Back and Spray! Risk of Endoscope Adhesion with Hemospray Application in Retroflexion in a Pediatric Patient. JPGN Rep. 2023;4:e336. [RCA] [PubMed] [DOI] [Full Text] [Reference Citation Analysis (0)] |
| 26. | Guzik P, Kabaria S, Parsa N. S2961 Help, I'm Stuck: A Case Series of Retained Endoscope Following Hemostatic Powder (TC-325) Use. Am J Gastroenterol. 2023;118:S1999-S1999. [DOI] [Full Text] |
| 27. | Echeverria C, Udaikumar J, Farha J, El Zoghbi M. Hemostatic Hold-Up: A Rare Case of Endoscopic Retention After Hemostatic Powder Application. ACG Case Rep J. 2026;13:e01981. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Reference Citation Analysis (0)] |
| 28. | Huang PM, Wang HP, Chen CN, Chen LH, Lee YC. Impaction of an upper gastrointestinal endoscope in the distal esophagus: alternative method of removal by double endoscopy without fluoroscopic assistance. Endoscopy. 2006;38:293. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 4] [Cited by in RCA: 4] [Article Influence: 0.2] [Reference Citation Analysis (0)] |
| 29. | Gagnier JJ, Kienle G, Altman DG, Moher D, Sox H, Riley D; CARE Group. The CARE guidelines: consensus-based clinical case reporting guideline development. J Med Case Rep. 2013;7:223. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 157] [Cited by in RCA: 426] [Article Influence: 32.8] [Reference Citation Analysis (0)] |
| 30. | Riley DS, Barber MS, Kienle GS, Aronson JK, von Schoen-Angerer T, Tugwell P, Kiene H, Helfand M, Altman DG, Sox H, Werthmann PG, Moher D, Rison RA, Shamseer L, Koch CA, Sun GH, Hanaway P, Sudak NL, Kaszkin-Bettag M, Carpenter JE, Gagnier JJ. CARE guidelines for case reports: explanation and elaboration document. J Clin Epidemiol. 2017;89:218-235. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1559] [Cited by in RCA: 1387] [Article Influence: 154.1] [Reference Citation Analysis (3)] |