Table 1 Interpretation of various chromatogram regions and commonly encountered peaks in various high-performance liquid chromatography windows (retention time ranges obtained from Bio-Rad kit inserts)
Region/windowRetention timeInterpretation
Pre-integration region< 1 minBilirubin, Hb H, Hb Barts, modified HbF
P1 peak0.74A minuscule peak usually found in specimens with increased HbF
F window0.98-1.22HbF, Hb Okayama
P2 window1.28-1.50Glycated HbA
P3 window1.50-1.90Aged samples, HbJ-Meerut, modified HbE
A window1.90-3.10HbA, Glycated HbS, intact Hb Koln
A2 window3.30-3.90HbE, Hb D-Iran, Hb Lepore, Hb G-Koushatta, Hb Zurich, Hb Korle Bu
D window3.90-4.30Hb D-Punjab, Hb G-Philadelphia
S window4.30-4.70HbS, Hb Q-Thailand, Hb Manitoba
C window4.90-5.30HbC, Hb Constant Spring, Hb Agenogi
Further unknown peaks> 5.30Hb Q-India