Observational Study
Copyright ©The Author(s) 2016. Published by Baishideng Publishing Group Inc. All rights reserved.
World J Dermatol. Nov 2, 2016; 5(4): 129-135
Published online Nov 2, 2016. doi: 10.5314/wjd.v5.i4.129
Multi-channeling optimized radiofrequency energy: A new age in well-established radiofrequency technology
Cruzy Tagger, Inna Belenky
Cruzy Tagger, Inna Belenky, Viora Inc., Jersey City, NJ 07306, United States
Author contributions: Tagger C and Belenky I contributed equally to this work.
Institutional review board statement: The study performed according good clinical practice requirement.
Informed consent statement: All study participants provided informed written consent prior to study enrollment.
Conflict-of-interest statement: Both authors, both authors are employees in Viora Company that manufactures the RF system used in the study. However, both authors did not receive any additional financial or other benefit/interest due to this study or publication. All data provided in the study is original and not modified.
Data sharing statement: Technical appendix, statistical code, and dataset available from the corresponding author at inna@vioramed.com. Participants gave informed consent for data sharing.
Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Correspondence to: Inna Belenky, PhD, Clinical Director, Viora Inc., 392 Summit Avenue, Jersey City, NJ 07306, United States. inna@vioramed.com
Telephone: +1-972-542315266 Fax: +1-972-99551345
Received: June 8, 2016
Peer-review started: June 14, 2016
First decision: July 29, 2016
Revised: August 22, 2016
Accepted: October 1, 2016
Article in press: October 9, 2016
Published online: November 2, 2016
Core Tip

Core tip: The significant change in circumferential measurements post-V-FORM treatments can be contributed not only to volume reduction due to improved metabolic rate and enhanced natural lipolysis, but also to edema reduction due to vacuum pressure integrated in the handpiece. Moreover, this technology enables the control of radiofrequency depth penetration which allows finishing the treatments with a skin tightening effect using higher radiofrequency frequencies (1.7 and 2.45 MHz).