Table 1 Simplified lumbar morphometric parameters used for implant design and finite element construct modelling
| Parameter | Mean value (mm) | Design relevance | Application in FEA model |
| ABVH | 28.97 | Reference for anterior body height | Used to define anterior support height |
| PBVH | 29.24 | Reference for posterior body height | Used to adjust posterior construct profile |
| VW | 41.13 | Reference for vertebral width | Used to determine implant footprint width |
| VD | 30.57 | Reference for vertebral depth | Used to determine implant anteroposterior depth |
| PL | 48.96 | Reference for pedicle length | Used to guide pedicle screw length selection |
| PW | 10.56 | Reference for pedicle width | Used to guide pedicle screw diameter selection |
Table 2 Material properties
| Component | Material | Young’s modulus (MPa) | Poisson’s ratio | Yield strength (MPa) | Model context |
| Polyethylene block | Polyethylene | 690 | 0.46 | 20 | ASTM-based model |
| Cortical bone | Bone | 17000 | 0.30 | 50 | Vertebra-based model |
| Cancellous bone | Bone | 100 | 0.20 | 1.67 | Vertebra-based model |
| Conventional mesh cage | Titanium | 110000 | 0.36 | 793 | Vertebra-based and ASTM-based models |
| Pedicle screws | Titanium | 110000 | 0.36 | 793 | Vertebra-based and ASTM-based models |
| Rods | Titanium | 110000 | 0.36 | 793 | Vertebra-based and ASTM-based models |
| VCR-MSS components | Titanium alloy | 110000 | 0.36 | 793 | Vertebra-based and ASTM-based models |