Revised: October 13, 2026
Accepted: November 2, 2026
Published online: November 28, 2026
Processing time: 100 Days and 9.1 Hours
Published evidence on paediatric computed tomography (CT) dose indicators and diagnostic reference levels in Palestine is limited, particularly for children with cancer who may undergo repeated imaging. Local diagnostic reference levels provide practical optimisation benchmarks where national paediatric reference levels are unavailable. We therefore examined whether examination-specific and size-specific dose distributions could identify actionable local optimisation priorities.
To establish paediatric oncology CT local diagnostic reference levels and identify actionable dose-optimisation priorities at Gaza’s sole specialist paediatric referral hospital.
This retrospective cross-sectional examination-level dose survey analysed 193 paediatric oncology CT examinations performed between March 2019 and December 2019. Age-specific and protocol-specific P75 values of scanner-recorded CTDIvol and dose-length product (DLP) were proposed as local dose audit benchmarks; P50 values and non-parametric bootstrap 95%CIs were also calculated. P75 estimates from groups of 4-9 examinations were treated as provisional. Additional descriptive analyses assessed patient-size grouping, scan range and protocol variability. Model-derived size-specific dose estimates, effective-dose and organ-dose estimates were not included because their computational inputs and conversion factors could not be recovered.
The principal CTDIvol P75 values were 23.46-23.73 mGy for brain CT, 2.44-3.15 mGy for chest CT, 4.68-6.78 mGy for abdomen CT and 4.66-7.50 mGy for chest/abdomen/pelvis CT in children aged 0-10 years. Provisional values for older small subgroups are reported separately. Chest/abdomen/pelvis CT demonstrated substantial DLP variability and extensive anatomical coverage, although brain CT had higher DLP in the two younger age groups. Confidence intervals were wider in smaller groups; an examination-level audit identified some discordance between recorded scan length and DLP/CTDIvol-derived length. Secondary associations were descriptive rather than evidence of avoidable dose.
These scanner-specific CT dose benchmarks provide a basis for local dose audit. Review of patient-size adaptation and scan range should be informed by clinical indication and image quality. Small-subgroup estimates remain provisional and patient-level dependence could not be quantified.
Core Tip: This 2019 retrospective cross-sectional examination-level survey reports local computed tomography dose reference values from Gaza’s specialist paediatric oncology service (193 examinations). The conventional P75 values are complemented by P50 benchmarks, bootstrap precision estimates and descriptive protocol variability. Chest/abdomen/pelvis computed tomography warrants prioritised clinical audit, but the lack of image-quality, indication and phase-level data prevents categorising dose variation as avoidable.
- Citation: Alshokry A, Alajerami Y, Najim A, Almukbel R, Mokbel K. Local diagnostic reference levels and optimisation priorities for paediatric oncology computed tomography at Gaza’s sole specialist paediatric hospital. World J Radiol 2026; 18(11): 127030
- URL: https://www.wjgnet.com/1949-8470/full/v18/i11/127030.htm
- DOI: https://dx.doi.org/10.4329/wjr.127030
Computed tomography (CT) is central to paediatric oncology imaging, including diagnosis, staging, treatment-response assessment, complication detection and follow-up surveillance. In children with cancer, CT may be performed repeatedly during treatment and survivorship, making radiation dose optimisation particularly important. Although CT provides substantial clinical benefit when justified, it contributes a higher radiation dose than many other diagnostic imaging modalities[1]. This is especially relevant in children because of their greater radiosensitivity and longer remaining lifespan during which radiation-related stochastic effects may theoretically develop[2-6].
Paediatric oncology CT must balance timely diagnosis, treatment assessment and follow-up with radiation protection, particularly where access to alternative imaging, modern scanner technology or harmonised protocols is limited[7]. Optimisation requires child-size protocol adaptation, avoidance of unnecessary phases and restriction of anatomical coverage while preserving diagnostic adequacy[8-10]. Diagnostic reference levels are investigation and optimisation benchmarks rather than dose limits; repeated or systematic exceedance should prompt review of equipment and clinical practice[11]. The service context is unusual because Al-Rantisi is the sole specialist paediatric referral hospital in Gaza and provides specialist paediatric and oncology services to children from across the territory[12,13]. For CT, CTDIvol reflects scanner output, whereas dose-length product (DLP) also incorporates scan length[11,14]. Because body size changes substantially during childhood, European guidance recommends age grouping for head CT and weight grouping for body CT[14]. The Gaza Strip has been described as facing a growing cancer burden, providing important local context for oncology service planning and imaging optimisation[15]. Local diagnostic reference levels (LDRLs) are especially important where national paediatric DRLs are unavailable. Consequently, although this is a single-institution study, the institution serves a territory-wide specialist paediatric population rather than a narrowly local catchment. International DRLs are useful comparators, but they may not reflect local scanner technology, patient population, maintenance constraints, radiographer practice or protocol design. A recent Gaza-based study in Radiological Physics and Technology established LDRLs for adult oncology trunk CT and showed dose variation linked to modifiable scan parameters, including tube current, tube voltage and scan length[16]. That adult study provides an important local precedent, while broader evidence shows substantial variation in CT dose between institutions and countries[17]. However, adult trunk CT values cannot be used as paediatric benchmarks because children require size-based optimisation and different radiation-protection considerations.
There is limited published evidence on paediatric CT dose indicators and local reference values in Palestine, particularly for paediatric oncology CT in Gaza. This study analysed examinations from Gaza’s specialist paediatric referral hospital. The primary objective was to establish examination-level P75 benchmarks and P50 typical values for brain CT, chest CT, abdomen CT without contrast and chest/abdomen/pelvis CT using scanner-recorded CTDIvol and DLP. Secondary aims were to describe acquisition characteristics and examine weight-stratified indicators, scan range, size adaptation and within-protocol variability. The study was descriptive; no formal prespecified inferential hypothesis was defined. The standard methods used to derive percentile benchmarks are not themselves novel; the principal contribution is the examination-level data from this specialist paediatric oncology service and locally relevant audit priorities.
This was a retrospective cross-sectional CT dose survey of paediatric oncology CT examinations performed at Dr Abd Al-Aziz Al-Rantisi Specialized Hospital for Children in the Gaza Strip, Palestine. Al-Rantisi Hospital is the sole specialist paediatric referral hospital in the Gaza Strip and provides tertiary paediatric services to children referred from across Gaza[12,13]. Its CT service therefore represents the principal dedicated paediatric CT service for the population served by the Gaza health system, rather than one of several equivalent specialist paediatric centres. The clinical examination dates were March 15 to December 19, 2019, within the March-December 2019 study period. All available eligible examination records identified during this period were screened; no sampling of eligible target-protocol records was undertaken. Clinical and scanner-dose information was entered into the study spreadsheet during 2019; individual research-abstraction dates were not documented separately from the clinical examination dates.
All examinations were acquired on a six-slice SOMATOM Emotion CT scanner (Siemens Healthineers, Erlangen, Germany), installed at the institution in 2008. Study documentation identifies syngo CT 2012E software (VB41A); this software designation could not be independently verified from the available examination records. CARE Dose4D automatic tube-current modulation was active during the examinations. CARE Dose4D adapts current according to patient size and attenuation, so a fixed-current acquisition should not be assumed. Scanner-reported CTDIvol and DLP indicators were extracted from post-acquisition dose reports, not pre-acquisition estimates. Siemens SOMATOM Emotion application guidance describes the effective mAs displayed after acquisition as the mean actually applied during the scan, and the Emotion instructions describe an examination-end dose report containing CTDIvol and DLP (technical documentation is detailed in the Supplementary material). This supports the interpretation that the recorded values were post-acquisition dose indicators rather than pre-scan maximum estimates. The recorded CTDIvol was interpreted as an acquisition-average value, although the individual scanner dose-report objects were unavailable to verify the precise software-specific calculation algorithm. Whether optional iterative reconstruction or artificial intelligence-based noise reduction was installed cannot be established from the available examination records. DRLs were treated as local audit tools rather than dose limits, in line with International Commission on Radiological Protection guidance[11].
The analytical unit was the CT examination rather than the individual child. Eligible records were all paediatric oncology CT examinations performed between March and December 2019 in children aged ≤ 15 years and recorded as one of four predefined target protocols: (1) Brain CT; (2) Chest CT; (3) Abdomen CT without contrast; and (4) Chest/abdomen/pelvis CT. All available eligible target-protocol examinations during the study period were included. Examinations recorded as neck/chest/abdomen/pelvis were excluded because this protocol was outside the predefined target examinations and involved a substantially different anatomical scan range.
A total of 207 examination records were screened. Thirteen neck/chest/abdomen/pelvis examinations were excluded because they were outside the predefined target protocol groups. Of the remaining 194 target-protocol examinations, one examination was excluded from the primary analysis because its protocol label was inconsistent with its acquisition and dose characteristics. The final analytical cohort comprised 193 CT examinations. Hospital identifiers were used to establish the count of 149 children contributing 194 target-protocol examinations before anonymisation. The retained analytical dataset contains no patient identifiers or linkage key, so the exact number of distinct children among the 193 included examinations cannot be confirmed. The flow of CT examination records through screening, exclusion and inclusion is summarised in Figure 1. The impact of this protocol-classification uncertainty was assessed in sensitivity analysis. Because this was a retrospective census of all available eligible target-protocol examinations during the study period, no a priori sample-size calculation was performed. The study size was determined by the number of eligible examination records available during that period. Age groups were assigned using recorded numerical ages rather than three inconsistent pre-existing age-group labels. This approach preserved the examination counts and percentile estimates used in the analysis. As each examination is an analytical observation, any child contributing repeated examinations could influence the empirical percentile distribution more than a child examined once. The resulting P75 is an examination-weighted distributional benchmark, not a patient-weighted percentile. Patient-level cluster resampling and cumulative-dose analysis were not possible without linkage identifiers.
For each eligible CT examination, demographic and examination information was abstracted from local clinical records. Acquisition parameters and radiation-dose indicators were obtained from scanner-generated post-acquisition dose reports rather than pre-scan estimates, as documented in the study records. CTDIvol was referenced to a 16-cm CTDI phantom for brain (head) CT and a 32-cm phantom for body CT. Scanner-reported CTDIvol and DLP values were retained without phantom conversion. CARE Dose4D automatic exposure control was active, and the recorded current and effective mAs variables should not be interpreted as proof of a fixed current during acquisition. Variables included age, sex, height, weight, protocol label, contrast use, kVp, tube current, effective mAs, pitch, rotation time, number of slices, recorded scan range, volumetric CT dose index (CTDIvol) and DLP. Each spreadsheet row contains one recorded CT examination with one scanner-reported DLP value. No additional acquisition phases belonging to the same clinical examination were separately entered elsewhere in the spreadsheet or knowingly omitted. However, some clinical examinations may have involved single or multiple acquisition phases, and the worksheets contain no phase counts or individual phase-dose entries. The individual scanner reports were unavailable, preventing independent determination of whether the recorded DLP was acquisition-specific or aggregated across phases in multiphase examinations. The scan-range field also cannot be classified reliably as planned, irradiated or reconstructed length. No phase totals or scan-length corrections were inferred.
The study data also included model-derived size-specific dose estimates (SSDE), effective-dose and selected organ-dose outputs associated with the ImPACT CT Patient Dosimetry Calculator version 1.0.4[18]. The ImPACT input files, scanner-matching and phantom settings, scan-region and phase entries, tissue-weighting convention, and documented SSDE patient-size measurements and conversion-factor tables were unavailable. The 16-cm head and 32-cm body CTDI reference-phantom conventions alone cannot validate head SSDE estimates without the requisite size metrics and conversion factors[19,20]. Model-derived SSDE, organ-dose and effective-dose results could not be independently verified and were therefore excluded from the present analysis. No substitute model-derived values were generated. The scanner-reported CTDIvol and DLP are independent primary indicators and remain the basis of all retained local benchmarks; the model-derived outputs were not substituted for those measured indices.
Clinical information was abstracted from hospital records, while technical parameters, including CTDIvol, DLP and scan range, were obtained from scanner-generated dose reports and entered into Excel. Data entry took place during 2019, although research-abstraction dates for individual examinations were not documented. Extracted entries were checked for completeness, inconsistencies and unusual values, with clinical or scanner source records consulted where necessary. No independent second review was reported. The examination categories followed original hospital or scanner protocol names, with anatomical coverage checked as necessary; one discordantly classified record could not be assigned reliably and was excluded from the primary analysis. Numerical ages originated from clinical records and were used to correct three inconsistent pre-existing age-group labels without changing any age or dose measurement. During a subsequent data-consistency assessment, we checked the retained 193 examinations against the eligibility rules and compared recorded scan range with 10 × DLP/CTDIvol (mm). Seven records differed by more than 50 mm. It could not be established whether these seven discrepancies had been individually investigated during data extraction. No scanner report or dose value was retrospectively altered, and sensitivity scenarios are reported separately. The precise scanner definition of recorded scan range could not be established, and the individual scanner dose-report objects were unavailable (Supplementary material).
The analysis used four predefined recorded categories: (1) Brain CT; (2) Chest CT; (3) Abdomen CT without contrast; and (4) Chest/abdomen/pelvis CT. Protocol categories were extracted from the original clinical or scanner labels, with anatomical coverage checked when necessary. In the retained study worksheet, contrast was recorded using a binary label (with/without contrast); positive entries denoted intravenous contrast. Among the 193 included examinations, 1/44 brain CT records and 48/51 chest/abdomen/pelvis CT records were labelled with contrast, while all 33 chest and 65 abdomen-without-contrast records were labelled without contrast. These source-data labels do not provide per-phase administration times or clinical indications. Clinical examinations included single-phase and multiphase acquisitions, although phase counts and protocol-specific patterns could not be reconstructed. Each examination was represented by one spreadsheet row, with no additional phases recorded separately or knowingly omitted. We retained the single reported DLP value per row without assuming how the scanner aggregated dose across any multiphase acquisitions. Accordingly, the P75 values are local benchmarks for the recorded examination categories, not harmonised single-phase protocol benchmarks; the uncertainty around phase-level source reporting remains a limitation. Further protocol-field and contrast-field details are provided in the Supplementary material.
The primary outcomes were protocol-specific and age-specific P75 values of scanner-reported CTDIvol and DLP. Secondary descriptive outcomes were P50, scan-range summaries, protocol variability, technical associations and weight-stratified CTDIvol/DLP distributions. No individual patient absorbed dose or effective dose is claimed, and no causal effect of technical settings on dose was estimated.
Age was analysed using the study categories of 0-5 years, 6-10 years and > 10-15 years. For the additional RP185-aligned size-stratified analysis, brain CT examinations were grouped as < 6 years and ≥ 6 years, while body CT examinations were grouped by weight as 5-15 kg, 15-30 kg and 30-50 kg[14]. Other continuous acquisition variables were analysed on their original scales. Age groups were assigned from recorded numerical ages because pre-existing age-group labels disagreed with numerical ages in three examinations.
LDRLs were calculated separately by examination type and age group as the 75th percentiles of the CTDIvol and DLP distributions. The P25, median, mean, minimum and maximum were also reported, with quartiles calculated using the exclusive percentile method. P25 and P75 values were not calculated for subgroups containing fewer than four examinations. For subgroups containing four to nine examinations, P75 values were retained as provisional LDRLs and interpreted cautiously because of the limited precision associated with small samples. Values derived from subgroups containing at least ten examinations were treated as the LDRLs. No statistical trimming was applied after the predefined protocol-based exclusions. The 75th percentiles were calculated using Excel PERCENTILE.EXC (the Weibull quantile definition). Non-parametric examination-level percentile bootstrap 95%CIs for P75 used 50000 within-stratum resamples, a fixed random seed (127030), and the same exclusive quantile estimator. Confidence intervals describe examination-level resampling uncertainty and may understate uncertainty when repeated examinations were obtained from the same child. P50 values were labelled descriptive achievable-dose indicators and were not interpreted as patient-level dose limits (full bootstrap results are supplied in the Supplementary material). These are examination-level percentile intervals: If children had multiple scans, resampling examinations independently could underestimate sampling variability under positive within-child dependence. The magnitude cannot be estimated without patient linkage; intervals are descriptive precision summaries and do not quantify independent-child uncertainty.
Additional analyses were undertaken to identify priorities for local review. First, an RP185-aligned size-stratified analysis summarised P50 and P75 CTDIvol/DLP values for brain CT by age and body CT by recorded weight; groups smaller than ten were flagged for cautious interpretation. Second, median, P75 and range of recorded scan range and median DLP/CTDIvol were summarised by examination type, with descriptive Pearson correlations between DLP and scan range or CTDIvol. Third, protocol-specific Spearman correlations examined the relationship of patient weight with CTDIvol and DLP. Fourth, coefficients of variation and maximum-to-P75 ratios described dose-index variability. Model-derived SSDE, effective-dose and organ-dose endpoints were excluded because their original inputs were unavailable for independent verification.
Exploratory Pearson correlations were calculated to describe associations of CTDIvol and DLP with technical parameters (reported in the Supplementary material). These were considered descriptive, particularly because DLP is mathematically related to CTDIvol and irradiated length. Exploratory multivariable models were not retained for interpretation because diagnostic assessment identified model misspecification, non-normal residuals, heteroscedasticity and influential observations; robust standard errors alone would not resolve those limitations.
To reduce selection and analytical bias, all available eligible target-protocol examinations during the study period were included rather than sampled, and no statistical trimming of dose values was undertaken after the predefined protocol-based exclusions. One examination with discordant protocol labelling and acquisition characteristics was excluded from the primary analysis and examined under alternative classifications in sensitivity analysis. Estimates from small subgroups were explicitly designated as provisional or were not calculated where quartile estimation was considered insufficiently reliable. A further data-consistency assessment compared recorded scan length with 10 × DLP/CTDIvol (in mm), solely as a diagnostic check rather than an exact physical identity. Because scanner acquisition and dose-report metadata were unavailable, discrepant values were flagged rather than corrected or removed. Sensitivity analyses excluded these records only illustratively, not as a revised primary cohort (reported in the Supplementary material).
Descriptive analyses and percentile summaries were carried out using the study examination dataset; bootstrap resampling and consistency checks were conducted programmatically using Python/NumPy. Exploratory correlation analyses were performed using Microsoft Excel and Stata/SE version 19 (StataCorp LLC, College Station, TX, United States). Age-stratified and weight-stratified analyses were descriptive and no interaction analyses were undertaken. Correlation P values, where reported, were unadjusted and interpreted as exploratory. No imputation was performed. Required variables were complete for each reported analysis, but no unique patient identifier was provided and repeated examinations could not be linked.
The final analytical cohort comprised 193 paediatric oncology CT examinations: (1) 44 brain CT examinations (22.8%); (2) 33 chest CT examinations (17.1%); (3) 65 abdomen CT examinations without contrast (33.7%); and (4) 51 chest/abdomen/pelvis CT examinations (26.4%) (Table 1). Overall, 93 examinations (48.2%) were performed in children aged 0-5 years, 79 (40.9%) in those aged 6-10 years and 21 (10.9%) in those aged > 10 years. Eighty-three examinations (43.0%) were performed in female patients and 110 (57.0%) in male patients (Table 1). No observations were excluded from the reported analyses because of missing values in the variables required for those analyses. Three pre-existing age-group labels were inconsistent with the recorded numerical ages. Classification using numerical age preserved the analytical cohort totals and protocol-age subgroup sizes.
| Examination | 0-5 years | 6-10 years | > 10 years | Female | Male | Total |
| Brain CT | 16 (36.4) | 23 (52.3) | 5 (11.4) | 18 (40.9) | 26 (59.1) | 44 (22.8) |
| Chest CT | 18 (54.5) | 12 (36.4) | 3 (9.1) | 15 (45.5) | 18 (54.5) | 33 (17.1) |
| Abdomen CT without contrast | 31 (47.7) | 26 (40.0) | 8 (12.3) | 28 (43.1) | 37 (56.9) | 65 (33.7) |
| Chest/abdomen/pelvis CT | 28 (54.9) | 18 (35.3) | 5 (9.8) | 22 (43.1) | 29 (56.9) | 51 (26.4) |
| Total | 93 (48.2) | 79 (40.9) | 21 (10.9) | 83 (43.0) | 110 (57.0) | 193 (100.0) |
Acquisition parameters varied by examination type (Table 2). Brain CT was performed at a fixed tube voltage of 110 kVp and pitch of 0.90, with a mean tube current of 114.8 mA and mean scan range of 167.2 mm. Chest CT had the lowest mean tube current (81.6 mA) and a mean scan range of 204.6 mm. Abdomen CT without contrast had a mean tube current of 148.6 mA and mean scan range of 169.1 mm, although scan range varied from 64.0 mm to 405.0 mm. Chest/abdomen/pelvis CT involved the greatest anatomical coverage, with a mean scan range of 428.4 mm and a range of 142.0-660.0 mm. CARE Dose4D was active, so the reported acquisition-level tube-current values are compatible with automated adaptation to patient size and attenuation. The dataset does not preserve the time-resolved modulation profile within each acquisition (Table 2).
| Examination | n | Tube voltage, kVp | Tube current, mA | Effective mAs | Pitch | Rotation time, seconds | Scan range, mm |
| Brain CT | 44 | 110.0 (110.0-110.0) | 114.8 (79.0-199.0) | 126.6 (107.0-143.0) | 0.90 (0.90-0.90) | 0.99 (0.60-1.00) | 167.2 (132.0-191.0) |
| Chest CT | 33 | 110.6 (110.0-130.0) | 81.6 (55.0-220.0) | 32.5 (22.0-89.0) | 1.50 (1.50-1.50) | 0.60 (0.60-0.60) | 204.6 (125.0-261.0) |
| Abdomen CT without contrast | 65 | 111.2 (110.0-130.0) | 148.6 (70.0-275.0) | 59.6 (28.0-110.0) | 1.50 (1.20-1.50) | 0.60 (0.60-0.60) | 169.1 (64.0-405.0) |
| Chest/abdomen/pelvis CT | 51 | 110.8 (110.0-130.0) | 152.1 (60.0-287.0) | 61.6 (24.0-115.0) | 1.49 (0.75-1.50) | 0.60 (0.60-0.60) | 428.4 (142.0-660.0) |
CTDIvol values and proposed LDRLs are shown in Table 3 and Figure 2. For brain CT, the principal CTDIvol LDRLs were similar in the 0-5-year and 6-10-year groups, at 23.46 mGy and 23.73 mGy, respectively. The provisional CTDIvol LDRL for the > 10-year brain CT subgroup was 23.70 mGy (n = 5). This narrow spread is consistent with the relatively fixed brain CT protocol. For chest CT, the CTDIvol LDRLs were 2.44 mGy for children aged 0-5 years and 3.15 mGy for those aged 6-10 years. A CTDIvol LDRL was not reported for the > 10-year chest CT subgroup because only three examinations were available. For abdomen CT without contrast, the principal CTDIvol LDRLs increased from 4.68 mGy in the 0-5-year group to 6.78 mGy in the 6-10-year group; the provisional LDRL for the > 10-year subgroup was 7.11 mGy (n = 8). Chest/abdomen/pelvis CT showed a similar age-related increase, with principal CTDIvol LDRLs of 4.66 mGy and 7.50 mGy in the 0-5-year and 6-10-year groups, respectively, and a provisional LDRL of 8.62 mGy in the > 10-year subgroup (n = 5) (Table 3 and Figure 2). The P50 values and bootstrap P75 confidence intervals for each age-protocol subgroup are shown in Table 3, with full resampling details in the Supplementary material. For example, the 6-10-year brain P75 of 23.73 mGy had a bootstrap interval of 23.22-24.10 mGy, whereas the provisional > 10-year chest/abdomen/pelvis P75 of 8.62 mGy had an interval of 6.80-10.02 mGy. These intervals should not be interpreted as independent-patient uncertainty estimates.
| Examination | Age group | n | Minimum | P25 | Mean | P50 (median) | P75 LDRL (95%CI) | Maximum |
| Brain CT | 0-5 years | 16 | 19.12 | 20.61 | 22.15 | 21.87 | 23.46 (21.99-25.36) | 26.00 |
| 6-10 years | 23 | 20.48 | 21.74 | 22.80 | 23.21 | 23.73 (23.22-24.10) | 24.64 | |
| > 10 years | 5 | 19.27 | 20.32 | 22.23 | 23.11 | 23.70 (22.24-23.91)1 | 23.91 | |
| Chest CT | 0-5 years | 18 | 1.98 | 2.12 | 2.35 | 2.25 | 2.44 (2.27-2.63) | 3.87 |
| 6-10 years | 12 | 2.21 | 2.75 | 2.96 | 3.00 | 3.15 (3.00-3.69) | 3.86 | |
| > 10 years | 3 | 2.29 | NA | 3.41 | 3.21 | NA | 4.72 | |
| Abdomen CT without contrast | 0-5 years | 31 | 3.27 | 4.25 | 4.42 | 4.40 | 4.68 (4.60-5.00) | 5.19 |
| 6-10 years | 26 | 2.75 | 4.49 | 5.71 | 5.63 | 6.78 (5.92-7.71) | 9.28 | |
| > 10 years | 8 | 5.17 | 5.41 | 6.28 | 6.37 | 7.11 (6.24-7.40)1 | 7.40 | |
| Chest/abdomen/pelvis CT | 0-5 years | 28 | 2.10 | 3.76 | 4.14 | 4.32 | 4.66 (4.40-4.94) | 5.51 |
| 6-10 years | 18 | 3.47 | 4.45 | 6.15 | 6.19 | 7.50 (6.29-8.53) | 9.67 | |
| > 10 years | 5 | 6.02 | 6.41 | 7.37 | 6.80 | 8.62 (6.80-10.02)1 | 10.02 |
DLP values and proposed DLP LDRLs are shown in Table 4 and Figure 3. For brain CT, the principal DLP LDRLs increased from 379.5 mGy·cm in the 0-5-year group to 411.0 mGy·cm in the 6-10-year group. The provisional DLP LDRL for the > 10-year subgroup was 418.0 mGy·cm (n = 5). The greater age-related variation in DLP than in CTDIvol suggests that anatomical coverage contributed to total examination dose. For chest CT, the DLP LDRLs were 48.3 mGy·cm and 76.3 mGy·cm in the 0-5-year and 6-10-year groups, respectively. No DLP LDRL was reported for the > 10-year subgroup because only three examinations were available. For abdomen CT without contrast, the principal DLP LDRLs increased from 72.0 mGy·cm to 119.8 mGy·cm in the 0-5-year and 6-10-year groups, respectively; the provisional LDRL for the > 10-year subgroup was 142.0 mGy·cm (n = 8). Chest/abdomen/pelvis CT showed substantial DLP values and variation, with LDRLs of 186.8 and 352.8 mGy·cm in the 0-5-year and 6-10-year groups, respectively, and a provisional LDRL of 431.5 mGy·cm in the > 10-year subgroup (n = 5) (Table 4 and Figure 3). The corresponding P50 and P75 confidence interval results for DLP are shown in Table 4 and detailed further in the Supplementary material.
| Examination | Age group | n | Minimum | P25 | Mean | P50 (median) | P75 LDRL (95%CI) | Maximum |
| Brain CT | 0-5 years | 16 | 290 | 325.2 | 352.2 | 340.5 | 379.5 (341.0-418.0) | 419 |
| 6-10 years | 23 | 340 | 371.0 | 390.2 | 386.0 | 411.0 (389.0-421.0) | 449 | |
| > 10 years | 5 | 356 | 379.5 | 400.0 | 405.0 | 418.0 (404.0-418.0)1 | 418 | |
| Chest CT | 0-5 years | 18 | 30 | 37.8 | 44.2 | 41.5 | 48.3 (42.0-62.0) | 81 |
| 6-10 years | 12 | 49 | 60.5 | 67.9 | 69.5 | 76.3 (69.8-80.2) | 81 | |
| > 10 years | 3 | 52 | NA | 79.7 | 72.0 | NA | 115 | |
| Abdomen CT without contrast | 0-5 years | 31 | 32 | 57.0 | 65.1 | 65.0 | 72.0 (67.0-80.0) | 99 |
| 6-10 years | 26 | 40 | 85.5 | 103.0 | 98.0 | 119.8 (102.8-142.2) | 156 | |
| > 10 years | 8 | 103 | 111.0 | 127.6 | 117.5 | 142.0 (117.0-181.0)1 | 181 | |
| Chest/abdomen/pelvis CT | 0-5 years | 28 | 29 | 136.5 | 149.5 | 159.5 | 186.8 (165.8-191.2) | 202 |
| 6-10 years | 18 | 163 | 226.3 | 307.8 | 293.5 | 352.8 (302.0-487.0) | 519 | |
| > 10 years | 5 | 170 | 244.0 | 339.0 | 344.0 | 431.5 (331.0-477.0)1 | 477 |
Overall protocol-level CTDIvol and DLP summaries, irrespective of age group, are provided in Supplementary Table 1. The overall CTDIvol P75 values were 23.69 mGy for brain CT, 3.05 mGy for chest CT, 6.02 mGy for abdomen CT without contrast and 6.26 mGy for chest/abdomen/pelvis CT. The corresponding overall DLP P75 values were 410.25 mGy·cm, 71.50 mGy·cm, 112.00 mGy·cm and 297.00 mGy·cm, respectively (Supplementary Table 1).
The numbers and percentages of examinations exceeding the principal or provisional age-specific LDRLs are summarised in Supplementary Table 2. Across most subgroups with reportable P75 values, approximately one-fifth to one-quarter of examinations exceeded the CTDIvol or DLP LDRL, as expected when the 75th percentiles are used as the reference level. The proportion exceeding either the CTDIvol or DLP LDRL was highest in the 6-10-year chest CT group, at 50.0%. In the provisional > 10-year chest/abdomen/pelvis CT subgroup, 40.0% exceeded either LDRL, although this finding should be interpreted cautiously because the subgroup contained only five examinations (Supplementary Table
The size-stratified analysis is shown in Table 5. For brain CT, age-stratified CTDIvol P75 values were similar in children aged < 6 years and ≥ 6 years, at 23.46 mGy and 23.69 mGy, respectively. DLP P75 was higher in the older group, increasing from 379.5 mGy·cm to 416.3 mGy·cm. These values refer to scanner output and recorded DLP, not estimated tissue absorbed dose.
| Protocol | Grouping parameter | n | CTDIvol median | CTDIvol P75 | DLP median | DLP P75 | Interpretation |
| Brain CT | Age < 6 years | 16 | 21.87 | 23.46 | 340.5 | 379.5 | Principal estimate (n ≥ 10) |
| Age ≥ 6 years | 28 | 23.16 | 23.69 | 389.5 | 416.3 | Principal estimate (n ≥ 10) | |
| Chest CT | Weight 5-15 kg | 8 | 2.15 | 2.35 | 38.0 | 40.8 | Interpret cautiously (n < 10) |
| Weight 15-30 kg | 20 | 2.62 | 3.07 | 58.5 | 73.5 | Principal estimate (n ≥ 10) | |
| Weight 30-50 kg | 5 | 3.20 | 3.97 | 68.0 | 93.5 | Interpret cautiously (n < 10) | |
| Abdomen CT without contrast | Weight 5-15 kg | 14 | 4.31 | 4.41 | 56.0 | 64.3 | Principal estimate (n ≥ 10) |
| Weight 15-30 kg | 34 | 4.68 | 5.33 | 77.5 | 98.0 | Principal estimate (n ≥ 10) | |
| Weight 30-50 kg | 17 | 6.56 | 7.33 | 118.0 | 145.0 | Principal estimate (n ≥ 10) | |
| Chest/abdomen/pelvis CT | Weight 5-15 kg | 12 | 4.31 | 4.65 | 137.0 | 162.8 | Principal estimate (n ≥ 10) |
| Weight 15-30 kg | 28 | 4.48 | 5.41 | 188.0 | 235.3 | Principal estimate (n ≥ 10) | |
| Weight 30-50 kg | 11 | 6.80 | 8.53 | 361.0 | 477.0 | Principal estimate (n ≥ 10) |
For body CT, weight stratification showed clearer dose-index patterns. In abdomen CT without contrast, CTDIvol P75 increased from 4.41 mGy in the 5-15 kg group to 5.33 mGy in the 15-30 kg group and 7.33 mGy in the 30-50 kg group. DLP P75 increased from 64.3 mGy·cm to 98.0 mGy·cm and 145.0 mGy·cm. In chest/abdomen/pelvis CT, DLP P75 increased from 162.8 mGy·cm to 235.3 mGy·cm and 477.0 mGy·cm across the same weight groups (Table 5). Some chest CT weight groups were small and should be interpreted descriptively.
Scan-range indicators are shown in Supplementary Table 3. Chest/abdomen/pelvis CT had the largest median scan range at 430.0 mm, P75 scan range 522.0 mm and maximum scan range 660.0 mm. Median DLP/CTDIvol, used as an approximate irradiated length, was also highest for chest/abdomen/pelvis CT at 41.2 cm. DLP was significantly correlated with scan range across all examination types, with Pearson r values of 0.68 for brain CT, 0.70 for chest CT, 0.65 for abdomen CT and 0.73 for chest/abdomen/pelvis CT, all P < 0.001 (Supplementary Table 3). Seven records differed by more than 50 mm between recorded scan range and the DLP/CTDIvol implied length, including three chest/abdomen/pelvis records by more than 100 mm. No corrected values were inferred. Removing those three records in a diagnostic, non-primary sensitivity calculation changed the chest/abdomen/pelvis DLP-scan-range correlation from 0.731 (n = 51) to 0.780 (n = 48), but the exclusions were not clinically validated (Supplementary material).
Weight associations using only scanner-recorded dose indicators are presented in Supplementary Table 4. Brain CTDIvol was not significantly correlated with weight. However, active CARE Dose4D means this finding cannot be used to infer a fixed tube-current protocol. In body CT, recorded weight was positively correlated with CTDIvol and DLP. These descriptive relationships do not demonstrate that the selected exposure settings were clinically optimal.
Dose variability indicators are shown in Supplementary Table 5. Brain CT had the lowest variability, with CTDIvol coefficient of variation 7.3% and DLP coefficient of variation 9.7%. Body CT protocols showed greater variability, particularly chest/abdomen/pelvis CT, where DLP coefficient of variation was 49.6% and maximum-to-P75 DLP ratio was 1.75 (Supplementary Table 5).
Protocol-specific correlations are shown in Supplementary Table 6. In brain CT, tube current correlated with CTDIvol, while scan range correlated strongly with DLP. In abdomen and chest/abdomen/pelvis CT, tube current correlated strongly with both CTDIvol and DLP, and scan range correlated significantly with DLP (Supplementary Table 6).
Multivariable regression findings were not retained because of severe residual-diagnostic problems and limited independent interpretability. The descriptive reference levels and size-stratified dose distributions do not depend on those models. The correlations in Supplementary Table 6 are exploratory and were not corrected for multiplicity.
Sensitivity analysis for the uncertain protocol classification showed little change in the main LDRLs (Supplementary Table 7). Full P50/P75 distributions and examination-level bootstrap intervals are documented in Supplementary Tables 8 and 9. A further examination-data consistency assessment documented the discordant length records without silently changing their dose fields (Supplementary Table 10).
This study established LDRLs for paediatric oncology CT examinations at Gaza’s sole specialist paediatric referral hospital. The findings show that dose indicators varied by examination type and patient age group. Brain CT showed relatively stable CTDIvol values across age groups despite active tube-current modulation, whereas abdomen CT and chest/abdomen/pelvis CT showed clearer increases in dose indicators with increasing age and body size. Although data were collected at one institution, Al-Rantisi is not one of several equivalent paediatric CT centres. It is the only specialised paediatric hospital in the Gaza Strip and provides tertiary referral CT services for children from across Gaza[12,13]. The study therefore captures dose practice at the territory’s principal specialist paediatric CT service. Nevertheless, the values remain specific to the scanner and protocols used during the study period and should be updated if equipment or protocols change.
The main optimisation message is that CTDIvol and DLP should be interpreted together. CTDIvol reflects scanner output, while DLP also incorporates scan length. In this study, brain CT showed relatively stable CTDIvol values but increasing DLP with age group, indicating that anatomical coverage contributed to total examination dose. For body CT, particularly chest/abdomen/pelvis CT, both dose level and dose variability were more prominent. This suggests that optimisation should prioritise patient-size adaptation, scan-range control and review of high-dose examinations rather than relying on scanner-output indices alone.
Chest/abdomen/pelvis CT emerged as the main local optimisation priority. This pattern is expected because the protocol covers a larger anatomical region than isolated chest or abdomen CT. However, the important issue is not only that the DLP values were higher, but that the protocol showed greater variability. This suggests that some examinations may involve longer scan ranges or higher technical exposure than necessary for the clinical question. Local review of anatomical coverage, scan start and end points, and protocol selection is therefore warranted. Because no indication-level information, acquisition phase count or diagnostic-quality evaluation was available, chest/abdomen/pelvis CT variation should be regarded as a reason for case review, not evidence that the recorded exposure was excessive. Wider anatomical coverage can be justified, for example when disseminated malignancy or clinically suspected complications require assessment of multiple body regions; these examples are clinical considerations, not recorded indications for the study examinations.
The observed DLP variability is therefore interpreted as an institution-level protocol-optimisation signal, particularly in relation to anatomical coverage and protocol selection, rather than as evidence of inter-institutional or regional dose variation.
International Commission on Radiological Protection guidance emphasises that DRLs are optimisation tools, not dose limits, and that local values should be interpreted in relation to clinical context[11]. This is particularly important in paediatric oncology, where CT examinations may be repeated during diagnosis, staging, treatment-response assessment and follow-up. The proposed LDRLs should therefore be used as audit and investigation levels rather than thresholds for judging individual examinations as safe or unsafe.
European paediatric DRL guidance recommends age grouping for head CT and weight grouping for body CT[14]. The present findings support this approach. Brain CT showed relatively stable scanner-output values across age groups, whereas body CT showed clearer dose-size patterns. This suggests that local body CT optimisation should consider patient weight or size directly, rather than relying only on broad age categories.
Patient size is relevant to interpreting CTDIvol, which is a scanner-output index rather than a measurement of tissue absorbed dose[19,20]. Size-specific dose estimation can support paediatric dose assessment where a validated patient-size metric and conversion-factor framework are available. The patient-size measurements and conversion factors required to verify the SSDE estimates were unavailable; those model-derived values were therefore excluded. The analysis instead compares recorded CTDIvol and DLP across age and weight groups, while recognising that identical CTDIvol values do not imply identical absorbed doses in different-sized children.
The findings complement recent Gaza-based adult oncology CT work, which established LDRLs for adult trunk CT and linked dose variation to modifiable scan parameters[16]. Adult trunk CT LDRLs cannot be used as paediatric benchmarks, but together the adult and paediatric findings support the need for a Gaza-specific CT dose-optimisation framework. This aligns with international evidence showing that CT dose varies substantially between institutions and countries, often because of protocol design, scanner operation and scan coverage rather than patient characteristics alone[17].
The study records also contained numerical ImPACT-based organ-dose and effective-dose estimates. ImPACT provides model-based population reference estimates, not direct measurements of an individual child’s organ dose[18,21-23]. The model input files, scanner-matching and phantom assumptions, tissue-weighting settings, and any paediatric adaptations could not be recovered. Because the necessary inputs cannot be verified, those model-derived outputs were excluded from the present analysis. Scanner-reported CTDIvol used 16-cm and 32-cm reference phantoms for head and body CT, respectively; no phantom conversion was applied.
Published paediatric CT DRLs provide context only when reference phantom size[24-26], anatomical coverage, patient-size grouping[27-30], indications and acquisition-phase definitions align[31-34]. The present study used a 16-cm head phantom and 32-cm body phantom, but the clinical indication and acquisition-phase structure could not be matched consistently across the external reports. We therefore avoid ranking the hospital against other services or inferring relative optimisation from unmatched numerical values.
The Egyptian national paediatric CT survey provides a broad geographical reference for brain and chest imaging[33], but the available methodological details do not establish complete comparability with this oncology-specific single-scanner dataset. Differences in CTDIvol or DLP may arise from study populations, acquisition choices and scan coverage. We therefore do not calculate percentage differences between these datasets.
Abdomen CT without contrast in the present dataset should not be treated as directly equivalent to an abdomen-pelvis protocol reported elsewhere[33]. Inclusion of the pelvis alters anatomical coverage and typically influences DLP; contrast-enhancement and phase structure may also differ. These distinctions preclude a valid numerical judgement about which protocol was better optimised.
The US Dose Index Registry provides body CTDIvol values based on a 32-cm phantom and reports chest/abdomen/pelvis examinations[24]. This shared phantom convention is useful but insufficient to ensure like-for-like comparison: The registry and present study do not have identical age boundaries, verified contrast and phase definitions, or oncology indications. We therefore present this source as descriptive context without percentage excess or superiority claims.
The principal conclusion from the local data is unchanged: Chest/abdomen/pelvis CT demonstrated relatively extensive coverage and considerable within-protocol DLP variation. These observations support a local review of scan start and end points, patient-size adaptation, image quality and clinical indication. They do not show that any individual examination was inappropriate or that its dose should match an externally reported benchmark.
Any future formal external benchmarking should prospectively harmonise patient size, dose-reference phantom, anatomical regions, contrast and phase structure, and required diagnostic quality. Until these details are available, the audit priorities drawn from within-hospital distributions are more defensible than cross-study percentage comparisons.
The size-stratified scanner-output results add information beyond pooled age-based LDRLs. Previous paediatric CT work emphasises the importance of body-size metrics when assessing radiation output[35-38]. Children of similar age may vary considerably in size, and fixed technical protocols may therefore merit local review. The study data support comparison of scanner CTDIvol and DLP by recorded age and weight, not individual absorbed-dose estimation.
Recent paediatric head CT studies emphasise the need to consider patient size, diagnostic adequacy and suitable dose metrics[37,38]. This is relevant to the present finding that brain CTDIvol varied relatively little with age. CARE Dose4D automatic tube-current modulation was active, so relatively stable brain CTDIvol across age groups should not be interpreted as evidence of a fixed tube current. Assessment of modulation performance and opportunities for opti
Scan range is a central optimisation target. Its association with DLP is expected because DLP is derived from CTDIvol and scan length, but this technical relationship remains clinically important. In paediatric oncology, where repeated imaging may occur, avoiding unnecessary anatomical coverage is one of the most practical ways to reduce radiation burden without compromising diagnostic quality. Chest/abdomen/pelvis CT should therefore be prioritised for review of scan start and end points, clinical indication and protocol selection.
Protocol variability also provides a practical ranking of optimisation priorities. Low variability may suggest standardisation, while high variability may indicate inconsistent protocol use, variable anatomical coverage or a subgroup of high-dose examinations. The findings suggest that chest/abdomen/pelvis CT contributes most to the high-dose tail and should be reviewed first.
The exploratory technical correlations should not be interpreted as independent dose determinants because scan length and CTDIvol are components of DLP. Available model diagnostics did not support interpretable multivariable inference. Established technical literature addresses tube current, tube voltage, pitch and scan length as CT dose determinants[39-43], while positioning and exposure-control systems can introduce additional variation[44,45]. None of these general relationships allows the present retrospective dataset to quantify how much of the observed dose variation could safely have been avoided.
For paediatric oncology CT, optimisation should not be understood as dose reduction alone. The goal is justified, diagnostically adequate imaging at the lowest radiation dose consistent with the clinical task. This requires balancing radiation protection with the need for accurate diagnosis, staging, treatment-response assessment and follow-up.
This examination-level survey provides local scanner-dose benchmarks from an underrepresented paediatric oncology service. Its verifiable outputs are the CTDIvol/DLP distributions, age-stratified and weight-stratified summaries, P50 typical values, examination-level bootstrap precision intervals and descriptive scan-range and protocol-variability analyses. Model-derived SSDE, organ-dose and effective-dose estimates were excluded because their computational inputs could not be independently verified.
Several limitations apply. Data came from one hospital and a single six-slice CT scanner, restricting transferability to other technology and practice. The scanner used a 16-cm CTDI reference phantom for brain CT and a 32-cm phantom for body CT, with CARE Dose4D tube-current modulation active during the examinations. CTDIvol and DLP values were obtained from post-acquisition scanner reports. The reported CTDIvol was interpreted as an acquisition-average value under active tube-current modulation, although the precise software-specific calculation could not be independently verified. The precise reconstruction algorithm and optional iterative or AI-based processing cannot be established. Al-Rantisi was Gaza’s specialist paediatric referral service, so the survey provides a study-period local reference, but it cannot provide regional or national CT reference levels. Proposed technical optimisation targets require diagnostic-quality and clinical-indication review before implementation.
Some age-specific subgroups were small. No LDRLs were calculated for subgroups containing fewer than four examinations, while P75 values from subgroups containing four to nine examinations were retained as provisional LDRLs and should be interpreted with greater uncertainty. The dataset was examination-level rather than patient-level, so cumulative dose across repeated examinations in individual children could not be assessed. Repeated examinations from the same child could also contribute separately to the dataset, meaning that children undergoing more frequent imaging may have contributed disproportionately to the dose distributions; the direction and magnitude of this potential weighting could not be quantified. The initial study cohort was reported as 149 children across 194 target-protocol examinations. However, the analytical dataset contained no individual identifiers to establish whether 148 or 149 distinct children remained after exclusion of one examination. The final unique-child count therefore cannot be verified for the 193-examination sample. Therefore, examination-level bootstrap intervals cannot correct for within-child clustering and may be too narrow. In particular, the reported P75 estimates describe the distribution of examinations rather than a distribution with one observation per child; the possible influence of more frequently imaged children on subgroup P75 values cannot be determined. Unlinked clustering may also affect precision and the true coverage of the nominal 95% bootstrap intervals.
Image quality, diagnostic adequacy, detailed clinical indication, scan-phase number and patient centring were not recorded reliably in the available examination dataset, although broad contrast labels were available. Consequently, a higher DLP or greater protocol variability cannot establish that an examination was unnecessary or technically inappropriate. Acquisition-specific DICOM dose reports were unavailable, and seven records showed recorded scan-range vs DLP/CTDIvol-derived-length discrepancies greater than 50 mm; scanner-recorded values were retained without alteration, and a sensitivity audit is provided. Model-derived SSDE, effective-dose and organ-dose values were omitted because the original input settings and correction factors cannot be retrieved. The precise patient-size conversion and any paediatric modifications therefore remain unknown, and no patient-specific absorbed dose is estimated. The retrospective QC audit was restricted to the accessible analytical data, rather than independent reconciliation against original DICOM files. The available study documentation does not establish a formal contemporaneous double-abstraction or external source-record verification procedure.
The numerical LDRLs are therefore most directly applicable to this hospital and to services using comparable scanner technology and paediatric CT protocols with automatic tube-current modulation, and should not be extrapolated directly to other institutions, general paediatric populations or newer CT systems. Nevertheless, the broader optimisation priorities identified, particularly patient-size adaptation and control of anatomical scan range, may be relevant to other resource-constrained paediatric CT services.
Three pre-existing age-group labels disagreed with numerical patient age; analytic age groups were assigned from numerical age while preserving scanner-reported dose values. The examination with uncertain protocol classification was excluded from the primary analysis, with alternative classifications evaluated in sensitivity analysis. Discrepant scan-range records were retained unchanged in the primary analysis. The CTDI reference phantoms were 16 cm for head CT and 32 cm for body CT; unavailable metadata include scan-phase structure and the inputs needed for secondary dose calculations.
This retrospective dose survey established paediatric oncology CT LDRLs at Gaza’s sole specialist paediatric referral hospital, providing, to our knowledge, the first such benchmarks for the territory’s specialist paediatric oncology CT service. Brain CT showed relatively consistent CTDIvol values across age groups despite active CARE Dose4D modulation, whereas abdomen and chest/abdomen/pelvis CT showed clearer increases in dose indicators with age and body size. Estimates derived from small subgroups should be interpreted as provisional.
Chest/abdomen/pelvis CT exhibited substantial within-protocol DLP variability and extensive anatomical coverage and remains a priority for clinical audit, not evidence of inappropriate scanning. Any changes to anatomical coverage or CARE Dose4D quality-reference and modulation settings require examination-specific indication review and confir
| 1. | Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007;357:2277-2284. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 7180] [Cited by in RCA: 5772] [Article Influence: 303.8] [Reference Citation Analysis (12)] |
| 2. | Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP, Howe NL, Ronckers CM, Rajaraman P, Sir Craft AW, Parker L, Berrington de González A. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet. 2012;380:499-505. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 3124] [Cited by in RCA: 2755] [Article Influence: 196.8] [Reference Citation Analysis (0)] |
| 3. | Miglioretti DL, Johnson E, Williams A, Greenlee RT, Weinmann S, Solberg LI, Feigelson HS, Roblin D, Flynn MJ, Vanneman N, Smith-Bindman R. The use of computed tomography in pediatrics and the associated radiation exposure and estimated cancer risk. JAMA Pediatr. 2013;167:700-707. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1226] [Cited by in RCA: 1105] [Article Influence: 85.0] [Reference Citation Analysis (0)] |
| 4. | Mathews JD, Forsythe AV, Brady Z, Butler MW, Goergen SK, Byrnes GB, Giles GG, Wallace AB, Anderson PR, Guiver TA, McGale P, Cain TM, Dowty JG, Bickerstaffe AC, Darby SC. Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ. 2013;346:f2360. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1598] [Cited by in RCA: 1440] [Article Influence: 110.8] [Reference Citation Analysis (5)] |
| 5. | Meulepas JM, Ronckers CM, Smets AMJB, Nievelstein RAJ, Gradowska P, Lee C, Jahnen A, van Straten M, de Wit MY, Zonnenberg B, Klein WM, Merks JH, Visser O, van Leeuwen FE, Hauptmann M. Radiation Exposure From Pediatric CT Scans and Subsequent Cancer Risk in the Netherlands. J Natl Cancer Inst. 2019;111:256-263. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 129] [Cited by in RCA: 261] [Article Influence: 43.5] [Reference Citation Analysis (0)] |
| 6. | Huang R, Liu X, He L, Zhou PK. Radiation Exposure Associated With Computed Tomography in Childhood and the Subsequent Risk of Cancer: A Meta-Analysis of Cohort Studies. Dose Response. 2020;18:1559325820923828. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 10] [Cited by in RCA: 36] [Article Influence: 6.0] [Reference Citation Analysis (0)] |
| 7. | World Health Organization. Global Initiative for Childhood Cancer [Internet]. Available from: https://www.who.int/initiatives/the-global-initiative-for-childhood-cancer. |
| 8. | Strauss KJ, Goske MJ, Kaste SC, Bulas D, Frush DP, Butler P, Morrison G, Callahan MJ, Applegate KE. Image gently: Ten steps you can take to optimize image quality and lower CT dose for pediatric patients. AJR Am J Roentgenol. 2010;194:868-873. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 205] [Cited by in RCA: 208] [Article Influence: 13.0] [Reference Citation Analysis (1)] |
| 9. | Greenwood TJ, Lopez-Costa RI, Rhoades PD, Ramírez-Giraldo JC, Starr M, Street M, Duncan J, McKinstry RC. CT Dose Optimization in Pediatric Radiology: A Multiyear Effort to Preserve the Benefits of Imaging While Reducing the Risks. Radiographics. 2015;35:1539-1554. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 44] [Cited by in RCA: 30] [Article Influence: 2.7] [Reference Citation Analysis (0)] |
| 10. | Gottumukkala RV, Kalra MK, Tabari A, Otrakji A, Gee MS. Advanced CT Techniques for Decreasing Radiation Dose, Reducing Sedation Requirements, and Optimizing Image Quality in Children. Radiographics. 2019;39:709-726. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 32] [Cited by in RCA: 60] [Article Influence: 8.6] [Reference Citation Analysis (0)] |
| 11. | Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, Ortiz-López P, Mattsson S, Padovani R, Rogers A; Authors on behalf of ICRP. ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging. Ann ICRP. 2017;46:1-144. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 806] [Cited by in RCA: 653] [Article Influence: 72.6] [Reference Citation Analysis (0)] |
| 12. | Palestinian Ministry of Health. Dr Abd Al-Aziz Al-Rantisi Specialized Hospital for Children [Internet]. 2009. Available from: https://www.moh.gov.ps/portal/%D9%85%D8%B3%D8%AA%D8%B4%D9%81%D9%89-%D8%AF-%D8%B9%D8%A8%D8%AF-%D8%A7%D9%84%D8%B9%D8%B2%D9%8A%D8%B2-%D8%A7%D9%84%D8%B1%D9%86%D8%AA%D9%8A%D8%B3%D9%8A-%D8%A7%D9%84%D8%AA%D8%AE%D8%B5%D8%B5%D9%8A-%D9%84/. |
| 13. | Palestinian Ministry of Health. Dr. Musa and Suhaila Nasir Pediatric Cancer Department provides integrated services for children in Gaza [Internet]. 2020. Available from: https://www.moh.gov.ps/portal/dr-musa-and-suhaila-nasir-pediatric-cancer-department-provides-integrated-services-for-children-in-gaza/. |
| 14. | European Commission. Radiation Protection No. 185. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Luxembourg: Publications Office of the European Union, 2018. [DOI] [Full Text] |
| 15. | Abdalla B, Mansour M, Ghanim M, Aia B, Yassin M. The growing burden of cancer in the Gaza Strip. Lancet Oncol. 2019;20:1054-1056. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 9] [Cited by in RCA: 13] [Article Influence: 1.9] [Reference Citation Analysis (0)] |
| 16. | Ayyad A, Alajerami Y, Najim A, Mansour HH, Karim MKA, Alghamdi F, Mokbel K. Establishment of local diagnostic reference levels for trunk computed tomography examinations at governmental hospitals in the Gaza Strip: a cross-sectional study. Radiol Phys Technol. 2026;19:782-791. [RCA] [PubMed] [DOI] [Full Text] [Reference Citation Analysis (0)] |
| 17. | Smith-Bindman R, Wang Y, Chu P, Chung R, Einstein AJ, Balcombe J, Cocker M, Das M, Delman BN, Flynn M, Gould R, Lee RK, Nelson TR, Schindera S, Seibert A, Starkey J, Suntharalingam S, Wetter A, Wildberger JE, Miglioretti DL. International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ. 2019;364:k4931. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 91] [Cited by in RCA: 123] [Article Influence: 17.6] [Reference Citation Analysis (0)] |
| 18. | ImPACT. ImPACT CT Patient Dosimetry Calculator [computer program], version 1.0.4 (27 May 2011). Hosted by the UK Health Security Agency Medical Dosimetry Group [Internet]. Available from: https://www.ukhsa-protectionservices.org.uk/mdg/patientdosimetry/patientdoseestimationtool. |
| 19. | American Association of Physicists in Medicine. AAPM Report No. 204. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. College Park: AAPM, 2011. Available from: https://www.aapm.org/pubs/reports/rpt_204.pdf. |
| 20. | American Association of Physicists in Medicine. Report No. 293 - Size Specific Dose Estimate (SSDE) for Head CT. The Report of AAPM Task Group 293 (2019). Available from: https://www.aapm.org/pubs/reports/detail.asp?docid=185. |
| 21. | The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37:1-332. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 443] [Cited by in RCA: 1214] [Article Influence: 63.9] [Reference Citation Analysis (0)] |
| 22. | Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology. 2008;248:254-263. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1427] [Cited by in RCA: 1360] [Article Influence: 75.6] [Reference Citation Analysis (5)] |
| 23. | Christner JA, Kofler JM, McCollough CH. Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection publication 103 or dual-energy scanning. AJR Am J Roentgenol. 2010;194:881-889. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 433] [Cited by in RCA: 376] [Article Influence: 23.5] [Reference Citation Analysis (1)] |
| 24. | Kanal KM, Butler PF, Chatfield MB, Wells J, Samei E, Simanowith M, Golden D, Gress DA, Burleson J, Sensakovic WF, Strauss KJ, Frush D. U.S. Diagnostic Reference Levels and Achievable Doses for 10 Pediatric CT Examinations. Radiology. 2022;302:164-174. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 5] [Cited by in RCA: 72] [Article Influence: 14.4] [Reference Citation Analysis (1)] |
| 25. | Vassileva J, Rehani M, Kostova-Lefterova D, Al-Naemi HM, Al Suwaidi JS, Arandjic D, Bashier EH, Kodlulovich Renha S, El-Nachef L, Aguilar JG, Gershan V, Gershkevitsh E, Gruppetta E, Hustuc A, Jauhari A, Kharita MH, Khelassi-Toutaoui N, Khosravi HR, Khoury H, Kralik I, Mahere S, Mazuoliene J, Mora P, Muhogora W, Muthuvelu P, Nikodemova D, Novak L, Pallewatte A, Pekarovič D, Shaaban M, Shelly E, Stepanyan K, Thelsy N, Visrutaratna P, Zaman A. A study to establish international diagnostic reference levels for paediatric computed tomography. Radiat Prot Dosimetry. 2015;165:70-80. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 38] [Cited by in RCA: 41] [Article Influence: 3.7] [Reference Citation Analysis (0)] |
| 26. | Almén A, Guðjónsdóttir J, Heimland N, Højgaard B, Waltenburg H, Widmark A. Paediatric diagnostic reference levels for common radiological examinations using the European guidelines. Br J Radiol. 2022;95:20210700. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 21] [Reference Citation Analysis (0)] |
| 27. | Bouchareb Y, Al Kharusi M, Al Maqbali A, Al Maimani A, Al Maskari H, Sirasanagandla SR, Al Jabri A, Al Kindi F, Al Shabibi S, Baawain S. Establishing Diagnostic Reference Levels for Paediatric CT Imaging: A Multi-Centre Study. Healthcare (Basel). 2025;13:1728. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 11] [Cited by in RCA: 6] [Article Influence: 6.0] [Reference Citation Analysis (0)] |
| 28. | Bos D, Wang Y, Stewart C, Chu PW, Luong J, Haubold J, Schröer B, Zensen S, Duong PT, Schindera S, Jeukens CRLPN, Das M, Einstein AJ, Cervantes L, Smith-Bindman R. Diagnostic reference levels for indication-based CT categories in pediatric CT: data from an international registry. Eur Radiol. 2025;35:7985-7999. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 5] [Reference Citation Analysis (1)] |
| 29. | Bani Yaseen AB, Trapp J, Fontanarosa D. Proposing diagnostic reference levels for paediatric computed tomography in Jordan: a national multicentre analysis. Phys Eng Sci Med. 2026;49:1179-1190. [RCA] [PubMed] [DOI] [Full Text] [Reference Citation Analysis (0)] |
| 30. | Garba I, Engel-Hills P. Paediatric diagnostic reference levels for common computed tomography procedures: A systematic review. Radiography (Lond). 2025;31:27-35. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 1] [Article Influence: 1.0] [Reference Citation Analysis (0)] |
| 31. | Public Health England. Doses from computed tomography (CT) examinations in the UK: 2011 review. 2014. Available from: https://assets.publishing.service.gov.uk/media/5a7e0bdc40f0b62302688d1a/PHE_CRCE_013.pdf. |
| 32. | The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Current Australian national diagnostic reference levels for multi detector computed tomography. 2025. Available from: https://www.arpansa.gov.au/research-and-expertise/surveys/national-diagnostic-reference-level-service/current-australian-drls/mdct. |
| 33. | Abdou SE, Salama DH, Ahmad KA, Sallam AM, El-Sayed EM, Talaat MS, El-Farrash RA, Vassileva J. 2021 National diagnostic reference levels for paediatric computed tomography in Egypt. Radiat Prot Dosimetry. 2022;198:423-433. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 3] [Reference Citation Analysis (0)] |
| 34. | Gyan E, Subaar C, Edusei G, Antwi Nyarko L. Paediatric computed tomography diagnostic reference levels in Africa: A systematic review. J Med Radiat Sci. 2025;72:139-147. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 2] [Reference Citation Analysis (0)] |
| 35. | Menke J. Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology. 2005;236:565-571. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 131] [Cited by in RCA: 120] [Article Influence: 5.7] [Reference Citation Analysis (0)] |
| 36. | Özsoykal İ, Yurt A, Akgüngör K. Size-specific dose estimates in chest, abdomen, and pelvis CT examinations of pediatric patients. Diagn Interv Radiol. 2018;24:243-248. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 17] [Cited by in RCA: 20] [Article Influence: 2.5] [Reference Citation Analysis (0)] |
| 37. | Sapignoli S, Roggio A, Boschini A, Guida F, Merlo C, Paiusco M, Zorz A, De Monte F. Size-specific dose estimates for pediatric head CT protocols based on the AAPM report TG-293. Phys Med. 2022;100:26-30. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 3] [Reference Citation Analysis (1)] |
| 38. | Chen T, Kong X, Peng W, Liao T, Hu H, Pan N, Yuan Z. Applying the AAPM 293 report to estimate the absorbed dose during head computed tomography: using head circumference for rapid dose estimation. Quant Imaging Med Surg. 2023;13:3140-3149. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 5] [Article Influence: 1.7] [Reference Citation Analysis (0)] |
| 39. | McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J. Strategies for reducing radiation dose in CT. Radiol Clin North Am. 2009;47:27-40. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 691] [Cited by in RCA: 548] [Article Influence: 32.2] [Reference Citation Analysis (5)] |
| 40. | Nelson TR. Practical strategies to reduce pediatric CT radiation dose. J Am Coll Radiol. 2014;11:292-299. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 36] [Cited by in RCA: 44] [Article Influence: 3.7] [Reference Citation Analysis (0)] |
| 41. | Kalra MK, Maher MM, Toth TL, Kamath RS, Halpern EF, Saini S. Comparison of Z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology. 2004;232:347-353. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 146] [Cited by in RCA: 117] [Article Influence: 5.3] [Reference Citation Analysis (0)] |
| 42. | Mahesh M, Scatarige JC, Cooper J, Fishman EK. Dose and pitch relationship for a particular multislice CT scanner. AJR Am J Roentgenol. 2001;177:1273-1275. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 101] [Cited by in RCA: 88] [Article Influence: 3.5] [Reference Citation Analysis (0)] |
| 43. | Lira D, Padole A, Kalra MK, Singh S. Tube potential and CT radiation dose optimization. AJR Am J Roentgenol. 2015;204:W4-10. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 76] [Cited by in RCA: 54] [Article Influence: 4.9] [Reference Citation Analysis (0)] |
| 44. | Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK. Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol. 2007;188:547-552. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 180] [Cited by in RCA: 141] [Article Influence: 7.4] [Reference Citation Analysis (1)] |
| 45. | Matsubara K, Koshida K, Ichikawa K, Suzuki M, Takata T, Yamamoto T, Matsui O. Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies. AJR Am J Roentgenol. 2009;192:862-865. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 75] [Cited by in RCA: 66] [Article Influence: 3.9] [Reference Citation Analysis (2)] |