Case Report Open Access
Copyright ©The Author(s) 2017. Published by Baishideng Publishing Group Inc. All rights reserved.
World J Cardiol. Jun 26, 2017; 9(6): 558-561
Published online Jun 26, 2017. doi: 10.4330/wjc.v9.i6.558
Inadvertent cardiac phlebography
Konstantinos Aznaouridis, Constantina Masoura, Stylianos Kastellanos, Albert Alahmar, Cardiology Department, Castle Hill Hospital, Hull and East Yorkshire NHS Trust, Cottingham HU16 5JQ, United Kingdom
Author contributions: Aznaouridis K and Alahmar A designed the report and revised the drafted manuscript; Masoura C collected the clinical data and drafted the manuscript; Kastellanos S collected the clinical data and drafted the manuscript.
Institutional review board statement: This case report was exempt from the Institutional Review Board standards at Hull and East Yorkshire NHS Trust, Hull, United Kingdom.
Informed consent statement: The involved patient gave her informed consent prior to study inclusion.
Conflict-of-interest statement: None.
Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Correspondence to: Konstantinos Aznaouridis, PhD, Cardiology Department, Castle Hill Hospital, Hull and East Yorkshire NHS Trust, Castle Rd, Cottingham HU16 5JQ, United Kingdom. konstantinos.aznaouridis@hey.nhs.uk
Telephone: +44-1482-622294
Received: October 28, 2016
Peer-review started: October 29, 2016
First decision: January 14, 2017
Revised: February 10, 2017
Accepted: April 6, 2017
Article in press: April 10, 2017
Published online: June 26, 2017

Abstract

We are reporting a case of a 80-year-old lady with effort angina who underwent coronary angiography through the right radial artery, using a dedicated radial multipurpose 5 French Optitorque Tiger catheter. The catheter was advanced into the left ventricle and a left ventriculogram was obtained, while the catheter appeared optimally placed at the centre of the ventricle and the pressure waveform was normal. A large posterior interventricular vein draining into the right atrium was opacified, presumably because the catheter’s end hole inadvertently cannulated an endocardial opening of a small thebesian vein, with subsequent retrograde filling of the epicardial vein. Our case suggests that caution is needed when a dedicated radial catheter with both an end-hole and a side hole is used for a ventriculogram, as a normal left ventricular pressure waveform does not exclude malposition of the end-hole against the ventricular wall.

Key Words: Thebesian vein, Radial access, Transradial, Cardiac phlebography

Core tip: Use of a dedicated radial catheter with both an end-hole and a side hole to perform a left ventriculogram, can result in inadvertent cannulation of a small Thebesian vein and subsequent opacification of a large epicardial vein. When such catheters are used for ventriculogram, a normal ventricular pressure waveform does not exclude malposition of the end-hole against the ventricular wall and extra caution is needed in order to prevent iatrogenic myocardial injury. We review current literature on myocardial injury induced by end-hole catheters used for left ventriculograms.



INTRODUCTION

In the last years, the use of radial artery as an arterial access site for coronary procedures has gained increasing popularity, as it is considered safer compared to transfemoral procedures. Recent data from large randomized trials suggest that the radial access is associated with a reduction of major adverse events in patients with acute coronary syndrome undergoing invasive management[1]. Furthermore, there is expanding use of dedicated “multipurpose” radial catheters, which enable the operator to cannulate both coronary arteries, and also to perform chamber injection and left ventriculogram. For transradial procedures, this single-catheter approach has been shown to decrease radiation exposure, fluoroscopy time, contrast volume and total procedure time compared with standard Judkins catheters[2]. Using a single catheter also reduces the risk of spasm of the radial artery. On the other hand, those dedicated radial catheters may rarely cause myocardial injury when used for a left ventriculogram. We present a rare angiographic finding in a patient who underwent cardiac catheterization through the radial artery using a dedicated radial catheter.

CASE REPORT

A 80-year-old lady with effort angina underwent coronary angiography through the right radial artery, using a dedicated radial “multipurpose” 5 French Optitorque Tiger catheter (Terumo, Somerset, New Jersey). Coronary angiogram of the left and right coronary arteries demonstrated a significant stenosis in the ostium of a modest sized intermediate artery. The catheter was then advanced into the left ventricle (LV) and appeared optimally placed at the centre of the LV, while the LV pressure waveform was normal. A left ventriculogram was obtained after delivering 10 mL of contrast with a vigorous hand injection. Few ectopics were noticed at the beginning of the injection, followed by a small deflection of the catheter’s tip, a minor subendocardial staining (arrowhead in panel A) and visualization of a large posterior interventricular vein, which seemed to drain directly into the right atrium (arrows in Figure 1 and supplementary Video 1).

Figure 1
Figure 1 Inadvertent phlebography of the posterior interventricular vein. Right anterior oblique image taken at time of left ventriculography. Arrows show the posterior interventricular vein in systole (A) and diastole (B). Arrowhead shows a minor subendocardial staining (A).

We observed no persistent staining of the myocardium and the patient did not experience any discomfort, arrhythmia or electrocardiographic changes. The patient was monitored and was discharged few hours later.

DISCUSSION

Dedicated radial “multipurpose” catheters such as the Tiger catheter have been specifically designed to minimise catheter exchange with ability to access both coronary ostia from the radial approach, and also provide ability to perform chamber injection with the presence of both an end hole and a single side hole. In our case, we assume that the catheter’s end hole cannulated an endocardial opening of a small Thebesian vein, with subsequent retrograde filling of the epicardial vein. Venae cordis minimae (Thebesian veins) are small valveless venous conduits that connect the coronary arteries, veins or capillaries with the cardiac chambers. Most Thebesian veins of the ventricles are connected to the cardiac venous system[3], as was the case in our patient.

We screened the available literature and we identified a total of 7 reports with 8 cases of myocardial injury following contrast injection with end-hole catheters for left ventriculogram[4-10]. The characteristics of the patients and procedures and the type of myocardial injury and outcome are shown in Table 1. In 5 of those cases, a dedicated transradial end-hole catheter with one side hole near the tip (radial Tiger catheter) or two side holes (radial Jacky catheter) was used[4,6,7,9,10]. Traditional multipurpose (MPA) catheters with an end-hole and 2 side holes near the tip were used in 2 cases[4,5], whereas a catheter with a single end-hole (Judkins right 4) was used from femoral access in 1 case[8]. The Thebesian venous network and/or cardiac veins were visualized in 4 cases[4,5,8,9]. A high-pressure power injection had been performed in most cases describing myocardial laceration/dissection and persistent myocardial staining[6,7,9,10]. Complete myocardial “perforation” with presence of contrast in the pericardial space was confirmed in 2 of those cases[6,10], and emergency pericardiocentesis due to tamponade was performed in one patient[6]. No fatalities were reported (Table 1).

Table 1 Patient and procedure-related characteristics and outcomes of published cases describing myocardial laceration or cannulation of Thebesian veins following left ventriculogram with end-hole catheters.
CaseRef.DemographicsCatheter, accessInjection characteristicsComplicationClinical findings/outcome
1Judkins et al[4]72 yr, woman, aortic stenosisMultipurpose-1 (right radial access)Not providedOpacification of Thebesian veins, coronary veins and coronary sinusNot provided
2Judkins et al[4]77 yr, woman, chest painOptitorque Tiger (right radial access)Not providedOpacification of Thebesian veins and coronary veinsNot provided
3Singhal et al[5]46 yr, man, hypertrophic cardiomyopathyMultipurpose-2 (femoral access)Power injection, 25 mL of contrast, 10 mL/sOpacification of Thebesian veins, coronary veins and coronary sinusVentricular tachycardia requiring cardioversion/uneventful recovery and next day discharge
4Frizzell et al[6]76 yr, woman, myocardial infarctionOptitorque Tiger (radial access)Power injection, 30 mL of contrast over 10 sLaceration/dissection of anterolateral myocardium and pericardial stainingChest discomfort, pericardial effusion and cardiac tamponade/pericardiocentesis, uneventful recovery
5Rossington et al[7]71 yr, woman, anginaOptitorque Tiger (right radial access)Power injection, 25 mL of contrast, 8 mL/s, 600 psiLaceration/dissection of anterolateral myocardiumChest discomfort, transient bundle branch block/uneventful course and next day discharge
6Aqel et al[8]50 yr, woman, chest painJudkins right 4 (femoral access)Hand injectionOpacification of Thebesian veins, coronary veins and coronary sinusNot provided
7Kang et al[9]66 yr, woman, anginaOptitorque Jacky radial (radial access)Power injection, 30 mL of contrast over 2 s, 600 psiLaceration/dissection of anterior myocardium and opacification of anterior interventricular vein and coronary sinusNot provided
8Basit et al[10]69 yr, man, inferior wall ischemiaOptitorque Jacky radial (radial access)Not providedLaceration/dissection of myocardium with pericardial opacificationChest pain, trivial pericardial effusion/uneventful recovery

Even a pigtail catheter can rarely cause severe myocardial injury during ventriculography when its tip is inappropriately positioned[11]. However, apposition of the pigtail catheter’s end-hole against the endocardium or cannulation of Thebesian veins is extremely unlikely, and therefore this catheter should be the preferred option for ventriculograms.

Caution is needed when a dedicated radial catheter with both an end-hole and side holes is used for a ventriculogram, as a normal left ventricular pressure waveform (likely from the side hole) does not exclude unsafe positioning of the end-hole against the ventricular wall[6,7]. This malposition of the catheter may result in cannulation and injection in a Thebesian vein[4,5,8], or injection against the endocardial layer. In our case, only a small volume of contrast was delivered in an endocardial opening of the Thebesian network with a hand injection, and this may partly explain the relatively “benign” outcome of opacifying an epicardial vein without causing any major myocardial injury. In this scenario, it seems that the injected contrast drains through the Thebesian network into the cardiac veins and therefore no significant intramyocardial shearing forces are generated. However, serious complications such as laceration/dissection of the myocardium or even catastrophic “perforation” of the ventricular wall[6,7,9,10] may occur when the contrast is injected against the endocardium, or when of a large contrast volume is injected in a Thebesian vein with high-pressure (with an automated power injector), as in this case the small Thebesian network would likely be unable to accommodate the forcefully injected large volume of contrast. Hence, we believe that our case indirectly supports the common practice of avoiding the use of radial multipurpose catheters with automated high-pressure power injectors and large volume of contrast for ventriculograms. Therefore, additional care should be taken to confirm that the catheter is optimally positioned at the centre of the left ventricular cavity and that the catheter’s tip is free before contrast injection, and the operator should not rely only on a normal waveform of ventricular pressure. Finally, the injection of contrast must stop immediately when subendocardial or myocardial staining or opacification of an epicardial vein occurs during a ventriculogram with a dedicated radial multipurpose catheter, as this invariably indicates myocardial injury due to malposition of the catheter’s end hole against the endocardium.

COMMENTS
Case characteristics

This case shows that using a dedicated radial catheter with both an end-hole and a side hole for a left ventriculogram can result in inadvertent cannulation of a small Thebesian vein and subsequent opacification of an epicardial cardiac vein. This was not related to any symptoms or adverse outcomes.

Clinical diagnosis

Minor catheter-induced endocardial staining and visualization of posterior interventricular vein.

Differential diagnosis

Catheter-induced laceration/dissection of myocardial wall.

Imaging diagnosis

Left ventriculogram with a dedicated radial Tiger catheter.

Related reports

Current literature on myocardial injury induced by end-hole catheters used for left ventriculogram is reviewed.

Experiences and lessons

When dedicated radial end-hole catheters are used for ventriculogram, a normal ventricular pressure waveform does not exclude malposition of the end-hole against the ventricular wall and extra caution is needed in order to prevent iatrogenic myocardial injury.

Peer-review

This is a case report about an unexpected cardiac phlebography. The manuscript is well written and describes an important aspect related to the use of multipurpose radial catheters.

Footnotes

Manuscript source: Invited manuscript

Specialty type: Cardiac and cardiovascular systems

Country of origin: United Kingdom

Peer-review report classification

Grade A (Excellent): A

Grade B (Very good): 0

Grade C (Good): C, C

Grade D (Fair): 0

Grade E (Poor): E

P- Reviewer: Amiya E, Barili F, Farand P, Zhang ZH S- Editor: Song XX L- Editor: A E- Editor: Li D

References
1.  Valgimigli M, Gagnor A, Calabró P, Frigoli E, Leonardi S, Zaro T, Rubartelli P, Briguori C, Andò G, Repetto A. Radial versus femoral access in patients with acute coronary syndromes undergoing invasive management: a randomised multicentre trial. Lancet. 2015;385:2465-2476.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 856]  [Cited by in F6Publishing: 869]  [Article Influence: 96.6]  [Reference Citation Analysis (0)]
2.  Chen O, Goel S, Acholonu M, Kulbak G, Verma S, Travlos E, Casazza R, Borgen E, Malik B, Friedman M. Comparison of Standard Catheters Versus Radial Artery-Specific Catheter in Patients Who Underwent Coronary Angiography Through Transradial Access. Am J Cardiol. 2016;118:357-361.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 14]  [Cited by in F6Publishing: 11]  [Article Influence: 1.4]  [Reference Citation Analysis (0)]
3.  Jain AK, Smith EJ, Rothman MT. The coronary venous system: an alternative route of access to the myocardium. J Invasive Cardiol. 2006;18:563-568.  [PubMed]  [DOI]  [Cited in This Article: ]
4.  Judkins C, Yamen E. Inadvertent thebesian vein cannulation during radial access ventriculography. JACC Cardiovasc Interv. 2013;6:e9-e10.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 2]  [Cited by in F6Publishing: 2]  [Article Influence: 0.2]  [Reference Citation Analysis (0)]
5.  Singhal S, Khoury S. Images in clinical medicine. Imaging of thebesian venous system. N Engl J Med. 2008;359:e8.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 7]  [Cited by in F6Publishing: 7]  [Article Influence: 0.4]  [Reference Citation Analysis (0)]
6.  Frizzell JD, Alkouz M, Sheldon MW. Left ventricular perforation during ventriculogram using an optitorque tiger catheter. JACC Cardiovasc Interv. 2014;7:1456-1457.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 2]  [Cited by in F6Publishing: 2]  [Article Influence: 0.2]  [Reference Citation Analysis (0)]
7.  Rossington JA, Aznaouridis K, Davison B, Oliver RM. A Tear in the Heart: Myocardial Laceration Following Left Ventriculogram With a Dedicated Radial Catheter. Am J Med Sci. 2016;352:533.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 1]  [Cited by in F6Publishing: 1]  [Article Influence: 0.1]  [Reference Citation Analysis (0)]
8.  Aqel R, Gupta R, Zoghbi GJ. Images in cardiology: Thebesian venous lake. Heart. 2005;91:1318.  [PubMed]  [DOI]  [Cited in This Article: ]
9.  Kang G, Kang K, Jurado J, Byrnes L. Left ventricular angiography with a Jacky radial catheter at high pressure leading to myocardial staining and opacification of the great cardiac vein and coronary sinus. EuroIntervention. 2012;8:410.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 2]  [Cited by in F6Publishing: 2]  [Article Influence: 0.2]  [Reference Citation Analysis (0)]
10.  Basit A, Nazir R, Hahn H. Myocardial and pericardial staining by transradial Optitorque Jacky shape catheter during left ventriculogram. J Invasive Cardiol. 2012;24:128.  [PubMed]  [DOI]  [Cited in This Article: ]
11.  Gach O, Lempereur M, Eeckhout E, Legrand V. Unintentional “ventriculo-phlebo-myo-pericardiography”. JACC Cardiovasc Interv. 2014;7:577-578.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 2]  [Cited by in F6Publishing: 2]  [Article Influence: 0.2]  [Reference Citation Analysis (0)]