World J Gastrointest Endosc. Feb 16, 2013; 5(2): 62-66
Published online Feb 16, 2013. doi: 10.4253/wjge.v5.i2.62
Comparison of Pentax HiLine and Olympus Lucera systems at screening colonoscopy
Alexey Chernolesskiy, David Swain, James C Lee, Gareth D Corbett, Ewen AB Cameron, Department of Gastroenterology, Addenbrooke’s Hospital, Cambridge CB2 0QQ, United Kingdom
Author contributions: Chernolesskiy A and Cameron EAB conceived and designed this study, drafted the article and approved the final version; Swain D contributed to data collection, revised the article for important intellectual content and approved the final version; Lee JC and Corbett GD performed statistical analysis of the data, revised the article for important intellectual content and approved the final version.
Correspondence to: Dr. Ewen AB Cameron, MD, FRCP, Consultant Gastroenterologist, Department of Gastroenterology, Addenbrooke’s Hospital, Hills Road, Box 293, Cambridge CB2 0QQ, United Kingdom. ewen.cameron@addenbrookes.nhs.uk
Telephone: +44-1223-348718 Fax: +44-1223-348729
Received: June 8, 2012
Revised: September 25, 2012
Accepted: December 22, 2012
Published online: February 16, 2013

Abstract

AIM: To compare the performance characteristics of Pentax HiLine (PHL) (with i-scan) and Olympus Lucera (OL) systems in a screening population.

METHODS: Screening colonoscopies in asymptomatic guaiac faecal occult blood test-positive patients with PHL (n = 58) and OL (n = 425) colonoscopes were analysed. All procedures were performed by a single colonoscopist. PHL used white-light endoscopy (WLE) on scope insertion and contrast/surface enhancement (i-scan 1) on withdrawal, and OL utilised WLE both on insertion and withdrawal. Patient age, sex, instrument insertion and withdrawal times, nurse assessed patient comfort scores, midazolam and fentanyl doses, procedure completion and rates of lesion detection were recorded separately for each group. Comparisons between the groups were made using either Fisher’s exact test (for dichotomous variables) or Mann-Whitney U test (for ordinal and continuous variables).

RESULTS: Colonoscopy completion rates were similar in both groups: 413/425 (97.2%) for OL and 55/58 (94.9%) for PHL (P = 0.24). For complete colonoscopies, the two groups were well matched for age, sex, colonoscope insertion times (mean 11.1 min in OL vs 11.6 min in PHL, P = 0.93) and normal colonoscopy withdrawal times (mean 15.6 min in OL vs 14.7 min in PHL, P = 0.2). Patients in the PHL group experienced a small increase in discomfort (mean patient comfort scores were 0.49 in the OL and 0.95 in the PHL group, P < 0.0001). While Fentanyl doses required were similar between groups (mean 57.5 μg in OL vs 61.4 μg in PHL, P = 0.13), slightly more Midazolam was required in the PHL group (mean 2.1 mg in OL vs 2.4 mg in PHL, P = 0.035). There was no difference in polyp (58% in OL vs 67% in PHL) or adenoma (49% in OL vs 56% in PHL) detection rates between the groups. Neither the total number of polyps and adenomas, nor the characteristics of these (including size, location or presence of advanced features) were different between the two systems.

CONCLUSION: This study suggests that there is no advantage of either colonoscope system in lesion detection.

Key Words: Pentax HiLine, i-scan, Polyp, Adenoma, Colonoscopy



INTRODUCTION

Colonoscopy is the gold-standard test for the diagnosis of colorectal neoplasia. Aside from diagnosis of colorectal carcinoma (CRC), detection and removal of pre-malignant precursor lesions is an essential part of colonoscopy and has been shown to significantly reduce mortality from CRC[1].

Even with the most thorough colonoscopic examination, adenoma miss rates, as established by same-day tandem colonoscopies, remain significant and are higher in patients with at least 2 adenomas found on baseline colonoscopy with an inverse correlation with adenoma size (between 6% for large adenomas ≥ 1 cm and 27% for those ≤ 5mm)[2]. The expected incidence of CRC in one study, based on missed adenomas at screening colonoscopy, was 1.1 per 1000 persons within 5 years[3].

The effect of a variety of technical manoeuvres and mucosal enhancement technologies on adenoma detection and miss rates has been studied, including prolongation of colonoscope withdrawal time[4], systematic patient positioning[5], pancolonic dye-spray chromoendoscopy[6], cap-assisted colonoscopy[7], narrow band imaging (NBI)[8] and endoscopic trimodal imaging[9], with variable degrees of success.

The i-scan technology was developed and introduced by Pentax (Japan) and is based on digital surface, contrast and tone enhancement, designed to reveal minute mucosal irregularities and subtle changes in colour that are not visible at conventional white light endoscopy (WLE)[10]. This technology has been incorporated into the Pentax HiLine (PHL) system representing a latest generation of high definition (HD+) colonoscopes.

The efficacy of i-scan compared to standard white-light colonoscopy has been studied in prospective trials and the results are somewhat conflicting. One randomized study of the PHL system utilising surface enhancement on withdrawal demonstrated its superiority to standard Pentax video colonoscopes in neoplastic (adenomatous and cancerous) lesion detection (38% vs 13%)[11]. Another prospective study, which randomised patients into 3 groups, each using a distinct modality on withdrawal (conventional high-definition WLE vs contrast/surface enhancement vs contrast/surface/tone enhancement), concluded there was no advantage in using i-scan for adenoma detection (31.9% vs 36.5% vs 33.1%, P = 0.742) and miss (22.9% vs 19.3% vs 15.9%, P = 0.513) rates[12]. One more non-randomised study, however, suggests a superior polyp detection rate (PDR) with PHL compared to Olympus Lucera (OL) (66% vs 44%, P = 0.01) with no difference in median polyp size between the groups (3 mm vs 4 mm, P = 0.98)[13]. This study enrolled 269 colonoscopies performed in the setting of the English National Bowel Cancer Screening Programme (BCSP) at a single centre. Neither surface nor tone enhancement was used systematically for lesion detection in this study.

We sought to determine whether there were advantages in terms of lesion detection with either OL WLE or PHL with routine use of i-scan surface enhancement on withdrawal.

MATERIALS AND METHODS

The BCSP in England offers biennial guaiac based faecal occult blood testing (FOBt) between the ages of 60 and 74 and colonoscopy for those testing positive. As part of equipment familiarisation in preparation for a trial, we have intermittently utilised PHL since May 2010.

All patients undergoing baseline colonoscopy performed by a single endoscopist (EC) between receipt of the PHL system (18th May, 2010) and the beginning of the study (27th September, 2011) were identified from the Endoscopy Department reporting system. Cases were included if either PHL colonoscopes or OL Q-series colonoscopes were used. Colonoscopies utilising alternative systems, incomplete and those performed for post-polypectomy surveillance were excluded.

For OL, insertion and withdrawal was performed using WLE with NBI at the discretion of EC for lesion characterisation. For PHL, insertion was performed using WLE. On intubation of the caecal pole, the mode was switched to contrast/surface enhancement (i-scan 1) for withdrawal with utilisation of contrast/surface/tone enhancement (i-scan 2) for lesion characterisation at the discretion of the EC. Chromoendoscopy was not used in any case.

Data on patient age, sex, colonoscopy completion and limiting factors, Midazolam (mg) and Fentanyl (μg) doses, nurse-assessed patient comfort score (0 = none, 1 = minimal, 2 = mild, 3 = moderate and 4 = severe), colonoscope insertion and withdrawal times (minutes), and the total number of polyps were collected for each colonoscopy. Each polyp detected was then characterized by its location in the colon, size (mm), morphology (stalked/sessile), histology (tubular/tubulo-villous or serrated adenoma, hyperplastic, or inflammatory), and presence of dysplasia (low- or high-grade) or cancer.

Intergroup comparisons were then made for each measure using either the Fisher’s exact test or Mann-Whitney U test with the polyp data divided according to numbers of polyps and adenomas, adenomas proximal to the splenic flexure (proximal adenomas) and advanced adenomas (i.e., > 1 cm in size, villous, or containing high-grade dysplasia or cancer). Polyp (PDR) and adenoma (ADR) detection rates were calculated for each group. For withdrawal times, only normal colonoscopies were included to remove any effect of polypectomy on duration. A power calculation was performed to determine whether these cohorts were of an appropriate size to detect the differences described by Banks et al[13].

As a retrospective service evaluation, this study did not require ethics committee approval under NHS Research Governance arrangements[14] and was registered with the local Risk and Patient Safety Department.

RESULTS

Colonoscopy completion rates were similar in both groups: 413/425 (97.2%) in the OL group and 55/58 (94.9%) in the PHL group (P = 0.24) with 3/12 (OL) and 2/3 (PHL) of the incomplete cases due to obstructing cancers. The two groups were matched for age and sex, and there was no difference in cancer detection rates (OL 7.7% vs PHL 7.3%, P = 1.0).

The colonoscope insertion and normal colonoscopy withdrawal times were similar with OL and PHL (Table 1). There was a small but statistically significant increase in mean ± SD, discomfort score in the PHL group. Although the average fentanyl dose was not different between the groups, there was also a small significant increase in midazolam requirements for the PHL patients.

Table 1 Colonoscopy completion rates, patient demographics, sedation doses, procedure times and comfort scores in the Olympus Lucera and Pentax HiLine groups.
IndicatorOlympus LuceraPentax HiLineP value
n (% of total)1
Complete colonoscopies413/425 (97.2)55/58 (94.9)0.24
Male patients239/413 (57.9)35/55 (63.7)0.5
mean ± SD2
Patient age (yr)66.3 ± 4.365.8 ± 4.70.36
Fentanyl dose (μg)57.5 ± 18.061.4 ± 18.50.13
Midazolam dose (mg)2.1 ± 0.62.4 ± 0.70.035
Comfort score0.49 ± 0.60.95 ± 0.6< 0.0001
Insertion time (min)11.1 ± 6.611.6 ± 7.50.93
Normal colonoscopy withdrawal time (min)15.6 ± 8.214.7 ± 8.00.20

There were no statistically significant differences in any of the polyp or adenoma detection measures between the two systems used. The comparison included the PDR, ADR (Table 2), mean numbers of polyps (MNP), adenomas (MNA), proximal and advanced adenomas, and polyps of large (> 1cm), small (≤ 1cm) and diminutive (≤ 5 mm) size (Table 3).

Table 2 Polyp and adenoma detection rates in the Olympus Lucera and Pentax HiLine groups.
Olympus LuceraPentax HiLineP value
With ≥ 1 polyp239/413 (58%)37/55 (67%)0.19
With ≥ 1 adenoma202/413 (49%)31/55 (56%)0.3
Table 3 Polyps and adenomas of various characteristics in the Olympus Lucera and Pentax HiLine groups (mean ± SD).
Size locationOlympus LuceraPentax HiLineP value
nmean ± SDnmean ± SD
Polyps
All5861.42 ± 1.96871.58 ± 1.730.19
> 1 cm1090.26 ± 0.53160.29 ± 0.500.51
≤ 1 cm4771.15 ± 1.78701.27 ± 1.580.27
≤ 0.5 cm3680.89 ± 1.53540.98 ± 1.270.22
Adenomas
All4141.00 ± 1.54591.07 ± 1.270.27
Proximal1640.4 ± 0.9210.38 ± 0.70.74
> 1 cm970.23 ± 0.50140.25 ± 0.480.6
≤ 1 cm3170.77 ± 1.38450.82 ± 1.110.29
≤ 0.5 cm2290.55 ± 1.13310.56 ± 0.960.63
Advanced adenomas
All1450.35 ± 0.7160.29 ± 0.50.64

Our study had an 88% power to detect the differences found by Banks et al[13].

DISCUSSION

One of the challenges of modern colonoscopy is to minimise the rate of missed pathology. It has been hoped that improvements in image quality might translate into increasing adenoma detection and there have been some suggestions from previous studies[11,13] that this might be the case. In this study, however, we have not been able to identify any benefit even when subclassifying according to location and size of polyp.

Our data is discordant with the results of a previous similar study[13]. There are several possible explanations for this, although it is noteworthy that our study has several advantages over the previous work, which would support the validity of our data. Firstly, our cohort was larger than that reported previously (n = 468 vs n = 269)[13], and accordingly this study was well-powered to detect the described effect size. Secondly, we compare the performance of a single colonoscopist rather than pooling colonoscopies performed by 5 ECs using OL and comparing them to those performed by a single EC with PHL. It is therefore plausible that at least some of the difference previously identified might be explained by variation in performance between colonoscopists rather than systems. ADRs were markedly lower in this earlier study. Although not specifically reported, a PDR of 44% with OL will translate into an ADR much lower than the 49% with OL in our study. As adenoma detection increases, the potential for missed adenomas reduces and with it the potential for measurable differences between groups. It is theoretically possible that the use of i-scan 1 during colonoscope withdrawal could have reduced lesion detection compared to PHL WLE without enhancements, although the high ADR in the PHL group in our study (56%) and the results of the study by Hoffman et al[11] would argue against this. Thirdly, the earlier study by Hoffman et al[11] suggested a marked advantage of high definition Pentax colonoscopes over standard WLE. Part of this advantage might be explained by the rather low ADR in the standard colonoscopy group (13% vs 38%). This study also included a heterogeneous patient population with patients undergoing screening colonoscopy, post-polypectomy surveillance or positive for faecal occult blood. The burden of pathology in these disparate patient groups is likely to be substantially different. Whilst no statistically significant difference was demonstrated in indications or sex of patients, there were some sizeable differences (more post-polypectomy surveillance and men with HD+) which in combination could have influenced the frequency of adenomas in each study arm. In addition, patients undergoing post-polypectomy surveillance represent a very heterogeneous group both by the size and numbers of previously detected adenomas[15]. In contrast, our study comprises a homogenous group of patients (asymptomatic, FOBt-positive, aged between 60 and 74), which overcomes these limitations.

In addition to assessing the most widely used quality indicators for colonoscopy (ADR, PDR, MNA and MNP), we also analysed two further indicators. We considered the detection of proximal adenomas (i.e., those proximal to the splenic flexure) as concerns have been raised about the performance of colonoscopy in the right colon[16-18]. We also studied the detection of advanced adenomas since the detection of these lesions is most important, due to their increased likelihood to develop into CRC. One might expect that an effective image enhancement technology would significantly improve the discrimination of smaller polyps from the background mucosa to a greater degree than of larger polyps as the majority of the latter should be readily identifiable by WLE alone, unless they are hidden from view. We therefore also performed a detailed subgroup analysis of MNP/MNA of large (> 1 cm), small (≤ 1 cm) and diminutive (≤ 5 mm) size in each patient group. These sub-analyses did not detect any differences in the performance of the two endoscopy systems.

Our findings suggest that patients in the PHL group experienced small but statistically significant increases in discomfort and required slightly more Midazolam. A slight increase in discomfort with this system has been described previously[13].

Although this data has several advantages over previous reports, there are some notable limitations. These include that our evaluation was not prospective or randomised and, as a single EC experience, it may not be generalisable.

In any cohort of patients, the number of detectable adenomas is finite. With increasingly thorough and systematic examination, the proportion of detected adenomas will approach a limit beyond which the contribution of any image enhancing technology will be negligible. Whilst studies tend to be carried out by high performing ECs, consideration should be given to evaluation in average performance settings.

In this detailed retrospective cohort study on a homogenous sample of FOBt-positive patients undergoing bowel cancer screening, we were unable to demonstrate any added benefit of either the HD+ PHL or the high-resolution OL system. All studied colonoscopy quality yield indicators were similar in the two studied patient groups. A randomised controlled study is required to fully evaluate the relative performance characteristics of these two systems.

COMMENTS
Background

Successful detection and removal of adenomas at colonoscopy is vital in reducing mortality from colorectal cancer. A number of optical image enhancement technologies have been developed and introduced into practical use with the aim of improving polyp detection from the background colonic mucosa. The i-scan technology is used in the Pentax HiLine (PHL) system of high-definition (HD+) colonoscopes and has been shown to produce better colonic polyp detection in comparison with the same system’s white light endoscopy (WLE).

Research frontiers

This study assessed and compared the performance of PHL with i-scan 1 (contrast/surface enhancement) mode and WLE with Olympus Lucera (OL), a high-resolution video-colonoscope system also widely and successfully used in modern endoscopy units.

Innovations and breakthroughs

This study featured a high-quality methodology to evaluate for potential differences in polyp detection, patient comfort and sedation requirements at colonoscopies performed with the two equipment systems, PHL with i-scan 1 and OL WLE in a homogenous group of 468 faecal occult blood test positive patients enrolled in the National Bowel Cancer Screening Programme at a single tertiary care centre in England. As compared with the only other similar published study (Banks et al, World J Gastroenterol 2011), this investigation is based on a larger series of patients (n = 468 vs n = 269), represents a single- (rather than single- compared with multiple-) colonoscopist performance, thus minimizing operator-dependent differences in polyp detection, and includes a detailed sub-group analysis by polyp size, type, location and advanced pre-malignant features.

Applications

Based on the results of this study, neither PHL with i-scan 1, nor OL WLE offer an advantage for the detection of colonic polyps of various sizes, type, location and advanced pre-malignant features. OL instruments are associated with lower degrees of patient discomfort and lower requirements for Midazolam during colonoscopy.

Terminology

I-scan - optical imaging technology invented by Pentax, Japan, based on the application of digital colour filters to achieve tone, contrast and surface enhancement of gastrointestinal mucosa as an alternative to the standard (white-light) endoscopy. PHL - a latest-generation system of HD+ colonoscopes equipped with the i-scan technology in addition to the standard white-light endoscopy mode.

Peer review

Publication was recommended by the reviewers as the study had demonstrated conflicting results compared to another similar but much smaller study (n = 269) published earlier, and statistical calculations were found to have been very well performed in this study.

Footnotes

P- Reviewer Galloro G S- Editor Song XX L- Editor A E- Editor Zhang DN

References
1.  Zauber AG, Winawer SJ, O’Brien MJ, Lansdorp-Vogelaar I, van Ballegooijen M, Hankey BF, Shi W, Bond JH, Schapiro M, Panish JF. Colonoscopic polypectomy and long-term prevention of colorectal-cancer deaths. N Engl J Med. 2012;366:687-696.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 1952]  [Cited by in F6Publishing: 2084]  [Article Influence: 173.7]  [Reference Citation Analysis (1)]
2.  Rex DK, Cutler CS, Lemmel GT, Rahmani EY, Clark DW, Helper DJ, Lehman GA, Mark DG. Colonoscopic miss rates of adenomas determined by back-to-back colonoscopies. Gastroenterology. 1997;112:24-28.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 1089]  [Cited by in F6Publishing: 1020]  [Article Influence: 37.8]  [Reference Citation Analysis (0)]
3.  Pohl H, Robertson DJ. Colorectal cancers detected after colonoscopy frequently result from missed lesions. Clin Gastroenterol Hepatol. 2010;8:858-864.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 180]  [Cited by in F6Publishing: 189]  [Article Influence: 13.5]  [Reference Citation Analysis (0)]
4.  Simmons DT, Harewood GC, Baron TH, Petersen BT, Wang KK, Boyd-Enders F, Ott BJ. Impact of endoscopist withdrawal speed on polyp yield: implications for optimal colonoscopy withdrawal time. Aliment Pharmacol Ther. 2006;24:965-971.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 153]  [Cited by in F6Publishing: 157]  [Article Influence: 8.7]  [Reference Citation Analysis (0)]
5.  East JE, Bassett P, Arebi N, Thomas-Gibson S, Guenther T, Saunders BP. Dynamic patient position changes during colonoscope withdrawal increase adenoma detection: a randomized, crossover trial. Gastrointest Endosc. 2011;73:456-463.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 78]  [Cited by in F6Publishing: 89]  [Article Influence: 6.8]  [Reference Citation Analysis (0)]
6.  Pohl J, Schneider A, Vogell H, Mayer G, Kaiser G, Ell C. Pancolonic chromoendoscopy with indigo carmine versus standard colonoscopy for detection of neoplastic lesions: a randomised two-centre trial. Gut. 2011;60:485-490.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 147]  [Cited by in F6Publishing: 150]  [Article Influence: 11.5]  [Reference Citation Analysis (0)]
7.  Hewett DG, Rex DK. Cap-fitted colonoscopy: a randomized, tandem colonoscopy study of adenoma miss rates. Gastrointest Endosc. 2010;72:775-781.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 90]  [Cited by in F6Publishing: 97]  [Article Influence: 6.9]  [Reference Citation Analysis (0)]
8.  Pasha SF, Leighton JA, Das A, Harrison ME, Gurudu SR, Ramirez FC, Fleischer DE, Sharma VK. Comparison of the yield and miss rate of narrow band imaging and white light endoscopy in patients undergoing screening or surveillance colonoscopy: a meta-analysis. Am J Gastroenterol. 2012;107:363-70; quiz 371.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 113]  [Cited by in F6Publishing: 120]  [Article Influence: 10.0]  [Reference Citation Analysis (0)]
9.  Kuiper T, van den Broek FJ, Naber AH, van Soest EJ, Scholten P, Mallant-Hent RCh, van den Brande J, Jansen JM, van Oijen AH, Marsman WA. Endoscopic trimodal imaging detects colonic neoplasia as well as standard video endoscopy. Gastroenterology. 2011;140:1887-1894.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 65]  [Cited by in F6Publishing: 74]  [Article Influence: 5.7]  [Reference Citation Analysis (0)]
10.  Kodashima S, Fujishiro M. Novel image-enhanced endoscopy with i-scan technology. World J Gastroenterol. 2010;16:1043-1049.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in CrossRef: 130]  [Cited by in F6Publishing: 140]  [Article Influence: 10.0]  [Reference Citation Analysis (0)]
11.  Hoffman A, Sar F, Goetz M, Tresch A, Mudter J, Biesterfeld S, Galle PR, Neurath MF, Kiesslich R. High definition colonoscopy combined with i-Scan is superior in the detection of colorectal neoplasias compared with standard video colonoscopy: a prospective randomized controlled trial. Endoscopy. 2010;42:827-833.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 116]  [Cited by in F6Publishing: 128]  [Article Influence: 9.1]  [Reference Citation Analysis (0)]
12.  Hong SN, Choe WH, Lee JH, Kim SI, Kim JH, Lee TY, Kim JH, Lee SY, Cheon YK, Sung IK. Prospective, randomized, back-to-back trial evaluating the usefulness of i-SCAN in screening colonoscopy. Gastrointest Endosc. 2012;75:1011-1021.e2.  [PubMed]  [DOI]  [Cited in This Article: ]
13.  Banks MR, Haidry R, Butt MA, Whitley L, Stein J, Langmead L, Bloom SL, O’Bichere A, McCartney S, Basherdas K. High resolution colonoscopy in a bowel cancer screening program improves polyp detection. World J Gastroenterol. 2011;17:4308-4313.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in CrossRef: 13]  [Cited by in F6Publishing: 11]  [Article Influence: 0.8]  [Reference Citation Analysis (0)]
14.  Governance Arrangements for Research Ethics Committees. A harmonised edition. Department of Health, 2011-05.  Available from: http://www.dh.gov.uk/en/Publicationsandstatistics/Publications/PublicationsPolicyAndGuidance/DH_126474.  [PubMed]  [DOI]  [Cited in This Article: ]
15.  Winawer SJ, Zauber AG, O’Brien MJ, Ho MN, Gottlieb L, Sternberg SS, Waye JD, Bond J, Schapiro M, Stewart ET. Randomized comparison of surveillance intervals after colonoscopic removal of newly diagnosed adenomatous polyps. The National Polyp Study Workgroup. N Engl J Med. 1993;328:901-906.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 701]  [Cited by in F6Publishing: 702]  [Article Influence: 22.6]  [Reference Citation Analysis (0)]
16.  Singh H, Nugent Z, Demers AA, Kliewer EV, Mahmud SM, Bernstein CN. The reduction in colorectal cancer mortality after colonoscopy varies by site of the cancer. Gastroenterology. 2010;139:1128-1137.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 341]  [Cited by in F6Publishing: 363]  [Article Influence: 25.9]  [Reference Citation Analysis (0)]
17.  Lakoff J, Paszat LF, Saskin R, Rabeneck L. Risk of developing proximal versus distal colorectal cancer after a negative colonoscopy: a population-based study. Clin Gastroenterol Hepatol. 2008;6:1117-121; quiz 1064.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 190]  [Cited by in F6Publishing: 181]  [Article Influence: 11.3]  [Reference Citation Analysis (0)]
18.  Baxter NN, Goldwasser MA, Paszat LF, Saskin R, Urbach DR, Rabeneck L. Association of colonoscopy and death from colorectal cancer. Ann Intern Med. 2009;150:1-8.  [PubMed]  [DOI]  [Cited in This Article: ]