Brief Article
Copyright ©2014 Baishideng Publishing Group Co., Limited. All rights reserved.
World J Gastrointest Endosc. Mar 16, 2014; 6(3): 82-87
Published online Mar 16, 2014. doi: 10.4253/wjge.v6.i3.82
Feasibility of breath monitoring in patients undergoing elective colonoscopy under propofol sedation: A single-center pilot study
Gurpreet W Anand, Ludwig T Heuss
Gurpreet W Anand, Ludwig T Heuss, Division of Internal Medicine, Zollikerberg Hospital, 8125 Zollikerberg/Zurich, Switzerland
Author contributions: Anand GW acquired the data; Heuss LT participated in planning the study and performed the colonoscopies as the gastroenterologist.
Correspondence to: Ludwig T Heuss, MD, MBA, Division of Internal Medicine, Zollikerberg Hospital, Trichtenhauserstrasse 20, 8125 Zollikerberg/Zurich, Switzerland. ltheuss@hin.ch
Telephone: +41-44-3972012 Fax: +41-44-3972688
Received: November 21, 2013
Revised: January 26, 2014
Accepted: February 16, 2014
Published online: March 16, 2014
Abstract

AIM: To determine whether a newly developed respiratory rate monitor can practically and accurately monitor ventilation under propofol sedation in combination with standard monitoring.

METHODS: Patients [American Society of Anesthesiologists (ASA) Classification I-III] scheduled for elective colonoscopy under propofol sedation were monitored with a new device that measures the respiratory rate based on humidity in expired air. Patients with clinically significant cardiac disorders or pulmonary disease and patients requiring emergency procedures were excluded from study participation. All of the patients also received standard monitoring with pulse oximetry. This was a single-center study conducted in a community hospital in Switzerland. After obtaining written informed consent from all subjects, 76 patients (51 females and 25 males) were monitored during colonoscopy under propofol sedation. The primary endpoint was the occurrence of any respiratory event (apnea or hypopnea). Apnea was defined as the cessation of breathing for a minimum of 10 s. Significant apnea was defined as the cessation of breathing for more than 30 s. Hypopnea was defined as a reduction in the respiratory rate below 6/min for a minimum of 10 s. Any cases of significant apnea triggered interventions by the endoscopy team. The interventions included withholding propofol, verbal stimulation of the patients, and increased oxygen supplementation or the chin lift maneuver. A secondary endpoint was the correlation of apnea or hypopnea with hypoxemia (measured as a decrease in SaO2 of at least 5% from baseline or less than 90%).

RESULTS: At least one respiratory event was detected in thirty-seven patients (48.7%). In total, there were 73 respiratory events, ranging from one to six events in a single patient. Significant apnea (> 30 s) occurred in five patients (6%). Only one episode of apnea led to a relative SaO2 reduction (from 98% to 93%) after a 50 s lag time. No event requiring assisted ventilation was recorded. Our analysis revealed that the total propofol dose was an independent risk factor for respiratory events (P = 0.01). Artifacts developed with the same frequency with the new device as with conventional pulse oximetry. Compared with pulse oximetry alone, this new monitoring device detected more respiratory events and may provide earlier warning of impending respiratory abnormalities.

CONCLUSION: Apnea commonly occurs during endoscopy under sedation and may precede hypoxemia. We recommend this respiration rate monitor as an alternative to capnography to aid in detecting ventilatory problems.

Keywords: Apnea, Colonoscopy, Conscious sedation, Deep sedation, Propofol, Pulse oximetry, Respiratory monitoring

Core tip: Apnea monitoring is a useful adjunct in assessing the ventilatory status of patients undergoing sedation. Capnography is too expensive to be used during normal endoscopic procedures. A newly developed respiratory rate-monitoring device based on the humidity of expired air enables the real-time assessment of ventilation. Compared with pulse oximetry alone, this new monitoring device detected more respiratory events and may provide earlier warning of impending respiratory abnormalities.