Rapid Communication
Copyright ©2007 Baishideng Publishing Group Co.
World J Gastroenterol. May 28, 2007; 13(20): 2841-2845
Published online May 28, 2007. doi: 10.3748/wjg.v13.i20.2841
Table 1 Ablation protocol
Baselinepower outputStep up rateMaximum outputRestart outputApplication process
30 W10 W/minUntil roll-off70% of the maximum outputTwice
Table 2 Ablation time to roll-off, maximum output, and coagulation diameter
Group with RFA aloneGroup with RFA following LP-TAEP value
Ablation time to roll-off (s)445.1 ± 35.9320.6 ± 30.9P < 0.05
Maximum output (W)87.0 ± 4.8369.0 ± 7.38P < 0.05
Coagulation diameter (mm)33.2 ± 2.2841.7 ± 3.85P < 0.05