Magnetic resonance imaging (MRI) is contraindicated in patients with an ingested video capsule endoscopy (VCE) due to the theoretical risk of mechanical and thermal harm. However, isolated cases of inadvertent scanning have not shown clear harm. Manufacturers advise against performing VCE in patients with an MRI scheduled within the next 14 days, yet asymptomatic retention for up to 12 years has been reported. With the increasing use of VCEs, the likelihood of incidental MRI exposure is rising. To date, no systematic data have been pub
To quantify magnetically induced forces on VCEs in a 3-Tesla (3-T) scanner to assess clinical safety risks.
Displacement forces on 11 commercially available VCE models were measured in a 3-T MRI scanner according to the American Society of Testing Materials (ASTM) F2052-21. Following ASTM F2213-17, torque was measured by the pulley method for eight VCEs and calculated for three magnet-containing gastric capsules with high dis
VCEs containing magnets showed significantly higher median displacement forces than those without magnets (2366 mN vs 293.92 mN, P = 0.012). Magnetically induced displacement forces were 50.7-fold higher than the individual gravitational force for magnet-containing devices (2366 mN vs 46.66 mN), compared to a 9.7-fold excess for capsules without a magnet (293.92 mN vs 30.20 mN). Median torque was 19.2 mNm for capsules without integrated magnets vs 289.02 mNm with magnets (P = 0.017). For all but one VCE, the forces exceeded the safety threshold defined by the capsule’s gravitational force. The mean length of internal metallic components was 21.3 mm (range: 18.2-25.4 mm); only two capsules slightly exceeded the 24-mm threshold for possible thermal effects.
In a 3-T MRI, all but one VCE, particularly magnet-containing capsules, exceeded accepted safety thresholds for displacement force and torque but remained below perforation-relevant levels.
Core Tip: With the increasing use of video capsule endoscopes (VCEs), accidental magnetic resonance imaging (MRI) in patients with an incorporated VCE is a growing risk. To date, no systematic data have been published on MRI-induced forces on VCEs. This in vitro study is the first to systematically quantify displacement forces and torque on 11 commercially available VCEs in a 3-Tesla MRI scanner. All but one VCE exceeded accepted safety thresholds, with magnet-containing capsules showing forces up to 50-fold above gravitational force. These findings highlight a potential safety concern relevant to both clinicians and engineers.