Copyright
©The Author(s) 2025.
World J Radiol. Jun 28, 2025; 17(6): 105728
Published online Jun 28, 2025. doi: 10.4329/wjr.v17.i6.105728
Published online Jun 28, 2025. doi: 10.4329/wjr.v17.i6.105728
Table 4 Stopping power ratio modeling and testing mean error (%) using computed tomography number as presented in Chika[21]
SPR training ME(%) | Total | Lung | Soft | Bone | SPR testing ME(%) | Total | Soft | Bone |
TSPR,1,0;fr(μ) | 0.28 | 0.04 | 0.01 | 0.84 | 0.63 | 1.02 | 0.09 | |
TSPR,1,2;fr(μ) | 0.00 | 0.04 | 0.00 | 0.00 | -0.05 | 1.06 | -1.62 | |
TSPR,1,2;fm(μ) | -0.09 | -0.03 | -0.02 | -0.24 | -2.08 | -1.48 | -2.92 | |
TSPR,1,2;fL(μ) | 0.00 | 0.00 | -0.01 | 0.02 | -1.04 | -0.85 | -1.30 | |
Stochiometric | -0.01 | -0.01 | -0.01 | -0.01 | -1.15 | -0.93 | -1.46 | |
Taasti | -0.02 | -0.01 | -0.01 | -0.03 | 2.19 | 2.95 | 1.14 | |
H-S method | 0.17 | 0.45 | 2.25 | -3.84 | -0.23 | 2.66 | -4.28 | |
TSPR,1,2;fr(μ)4g | 0.00 | 0.04 | 0.00 | 0.00 | -0.31 | 0.62 | -1.62 |
- Citation: Chika CE. Advances in dual energy computed tomography approach for proton stopping power ratio computation in radiotherapy. World J Radiol 2025; 17(6): 105728
- URL: https://www.wjgnet.com/1949-8470/full/v17/i6/105728.htm
- DOI: https://dx.doi.org/10.4329/wjr.v17.i6.105728