Table 1 Stopping power ratio training and testing root mean square errors (%) using reference ρ e as presented in Chika[20]
SPR training RMSE (%) | Total | Lung | Soft | Bone | Testing RMSE (%) | Total | Soft | Bone |
SPR1,1 continuous | 0.32 | 0.70 | 0.35 | 0.13 | 0.73 | 0.80 | 0.62 | |
SPR1,1 piece-wise | 0.22 | 0.02 | 0.27 | 0.04 | 0.92 | 1.09 | 0.61 | |
SPR0,3 continuous | 0.31 | 0.14 | 0.38 | 0.07 | 0.71 | 0.77 | 0.62 | |
Kanematsu | 2.03 | 0.00 | 2.08 | 2.04 | 1.77 | 1.80 | 1.63 |