Table 1 Optimized conventional imaging protocol at 1.5 Tesla (Magnetom Avanto B13, Siemens, Erlangen, Germany), configured with a 12 element designed head matrix coil
1.5 TPlaneSeq.TR (msec)TE/TI (msec)Flip angleSlice (mm)FOVNEXMatrix sizeScan time
NeonatesAxial1SE T1w450-650129032003192 × 2564.30
AxialIR840060/35018032302176 × 5124.30
Coronal AxialTSE T2w89009914532002176 × 2565.00
Axial2GRE T21w640183552002160 × 2563.30
Axial DWI (b0 and b800)EPI5700139-5200496 × 1281.00
Infants and adolescentsAxialFLAIR T2w7000110/200018032003154 × 2564.30
Axial1SE T1w450-650129032003192 × 2564.30
AxialIR840060/35018032302176 × 5124.30
CoronalTSE T2w89009914532002176 × 2565.00
Axial2GRE T21w640183552002160 × 2563.30
Axial DWI (b0 and b1000)EPI5700139-5200496 × 1281.00
Table 2 Optimized conventional imaging protocol at 3.0 Tesla (Magnetom Skyra, Siemens, Erlangen, Germany), configured with a 64 element designed Head Matrix coil
3.0 TPlaneSeq.TR (msec)TE/TI (msec)Flip angleSlice (mm)FOVNEXMatrix sizeScan time
NeonatesAxial1TSE T1w5506.713821804256 × 803.07
AxialIR200612/50015021802256 × 804.48
Coronal AxialTSE T2w1032012218021802384 × 903.18
Axial2GRE T21w600202021802256 × 853.35
Axial3FLAIR900088/250015021801320 × 703.56
Axial1 DWI (b0 and b800)EPI810076-22006192 × 1003.40
Infants and adolescentsAxialTSE T1w6006.415032003256 × 2561.36
Axial2IR250011/50015032201320 × 803.27
Coronal and AxialTSE T2w638010815032201384 × 751.29
Axial3GRE T21w630202042201320 × 752.33
AxialFLAIR Fat Sat900085/250015032201320 × 703.56
Axial1 (b0 and b1000)EPI900098-32203192 × 1003.00