Teoh AY, Ng EK, Chan SM, Lai M, Moran S, Binmoeller KF, Moon JH, Ho KY. Ex vivo comparison of the lumen-apposing properties of EUS-specific stents (with video).
Gastrointest Endosc 2016;
84:62-8. [PMID:
26684601 DOI:
10.1016/j.gie.2015.11.041]
[Citation(s) in RCA: 40] [Impact Index Per Article: 4.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/03/2015] [Accepted: 11/25/2015] [Indexed: 12/11/2022]
Abstract
BACKGROUND AND AIMS
Several EUS-specific stents have become available. It has been claimed that some of these stents have lumen-apposing properties, but objective data measuring such properties are not available. The aim of this study is to measure the lumen-apposing force (LAF) of these stents.
METHODS
The LAF of 3 EUS-specific metallic stents (stents A, N, and S) were compared in an ex vivo setting. Four types of anastomoses were performed with the stents including cholecysto-duodenal, cholecysto-gastric, gastro-gastric, and gastro-jejunal and compared with a hand-sewn (HS) equivalent of the anastomosis. The outcome parameter was the LAF created by each type of stent.
RESULTS
Sixty-four anastomoses were created. The overall mean (standard deviation) LAFs were significantly higher for stents A and S (P < .001). This difference persisted regardless of the type of anastomosis: gastro-gastric (P = .002), gastro-jejunal (P = .005), cholecysto-gastric (P = .002), and cholecysto-jejunal (P = .003). The differences in LAF created by each type of stent across different types of anastomoses were also compared. A trend to significance was observed in the anastomoses created by stent N (P = .064) and stent A (P =.052); a significant difference in LAF was observed among different anastomoses created by stent S (P = .015). The LAF created by HS anastomosis was significantly higher than that for all stents across all anastomoses.
CONCLUSIONS
Stents A and S had a higher LAF. The use of these stents should be considered when performing EUS-guided transmural luminal anastomoses in non-adherent organs. Further studies are required to confirm the clinical efficacies of these EUS-specific stents.
Collapse