©The Author(s) 2016.
World J Methodol. Jun 26, 2016; 6(2): 133-142
Published online Jun 26, 2016. doi: 10.5662/wjm.v6.i2.133
Published online Jun 26, 2016. doi: 10.5662/wjm.v6.i2.133
Table 1 The methods commonly used for analyzing hemoglobin A1c according to the surveys conducted by Labquality Ltd. in 1994
| Numerical summary from a diabetic EDTA-blood sample 1/1994 | ||||
| Analyte/method group | Mean | SD | CV% | Number |
| GHbA1c % | ||||
| Ion exchange Ciba Corning | 8.7 | 0.4 | 4.3 | 4 |
| Ion exchange Diamat Biorad | 9.7 | 0.8 | 8.4 | 15 |
| Ion exchange Kyoto Daichii | 8.4 | 0.3 | 3.4 | 3 |
| Ion exchange Pharmacia MonoS | 9.0 | 0.8 | 8.7 | 17 |
| Ion exchange Shimadzu | 10.7 | 0.1 | 0.7 | 2 |
| Ion exchange others | 9.5 | 0.8 | 7.9 | 5 |
| Ion exchange minicolumn Biorad | 9.7 | 0.8 | 8.1 | 4 |
| Electrophoresis Beckman | 9.7 | 0.8 | 7.8 | 8 |
| Affininity method Abbott Imx | 8.4 | 0.6 | 7.5 | 15 |
| Affininity method Abbott Vision | 8.8 | 0.4 | 4.8 | 8 |
| LaInIA Ames DCA 2000 | 9.1 | 0.2 | 2.6 | 29 |
| TinIA Boehringer | 9.1 | 1.0 | 10.4 | 15 |
| Immunochemical others | 8.8 | 1.7 | 19.3 | 2 |
| All method groups | 9.2 | 0.8 | 8.9 | 127 |
| GHbA1 | ||||
| Ion exchange Ciba Corning | 11.8 | 0.3 | 2.4 | 2 |
| Ion exchange Diamat Bio-Rad | 12.1 | 0.4 | 3.2 | 7 |
| Ion exchange others | 11.5 | - | - | 1 |
| All methods groups | 12.0 | 0.4 | 3.3 | 10 |
| GHb | ||||
| Affinity method Abbott Imx | 13.4 | 0.4 | 2.7 | 2 |
| Affinity method mini col Isolab | 10.3 | 1.4 | 13.7 | 2 |
| All method groups | 11.8 | 2.0 | 16.5 | 4 |
- Citation: Penttilä I, Penttilä K, Holm P, Laitinen H, Ranta P, Törrönen J, Rauramaa R. Methods, units and quality requirements for the analysis of haemoglobin A1c in diabetes mellitus. World J Methodol 2016; 6(2): 133-142
- URL: https://www.wjgnet.com/2222-0682/full/v6/i2/133.htm
- DOI: https://dx.doi.org/10.5662/wjm.v6.i2.133