Table 2 Summary of clinical and perfromance metrics of different tomography-based microwave radiometric imaging systems
System
Scan time
Breast size
Accuracy (true positive rate)
Sensitivity (true negative rate)
Specificity
Patients tested
Dartmouth college[20,21]2.2 minutes14.2 cm diameterDetected a 4-cm tumor in a trial test with one patientNot reportedNot reported1
McMaster university[22]5 hours4.8 cm diameterReported a 247 mm resolution in simulationNo clinical trialsNo clinical trials0
Istanbul Technical University (SAFE)[23,24]20 minutesSmall and largeDetected lesions as small as 4 mm79%77%113
University of Bristol (MARIA)[25]5 minutes32A and 32DDSmallest recorded lesion was 5 mm76%Not reported8
University of Calgary (TSAR)[26]< 30 minutesB and C size10 mm lesion detected with SCR ~5.0 dBNo clinical trialsNo clinical trials1
McGill University[27]6 minutesAmid A to D size10 mm spherical phantom tumor detectedNo clinical trialsNo clinical trials4
Galway University (Wavelia)[28] 6 minutesGreater than 32BDid not detect tumors < 10 mm in trials75%Not reported5
Chalmers University[29,30]Not reported55.0-75.0 cm diameterUsed oil cups mimicking tumors (55-75 mm)No clinical trialsNo clinical trials0
University of Perugia (Mammowave)[31]8 minutesVarious cup sizes73%82%353
Hiroshima University[32]15 minutesSmall (Japanese patients)Minimum size 40 mm in trials (Japanese only)49%55%9
University of São Paulo[33]28 minutes34B (imperial system)1 mm tumor in 150 mm breast phantom with SCR > 7.0 dBNo clinical trialsNo clinical trials3
University of Queensland[34]45 minutes12.4 cm20 mm × 30 mm water target in phantomNo clinical trialsNo clinical trials1