Table 2 Summary of clinical and perfromance metrics of different tomography-based microwave radiometric imaging systems
| System | Scan time | Breast size | Accuracy (true positive rate) | Sensitivity (true negative rate) | Specificity | Patients tested |
| Dartmouth college[20,21] | 2.2 minutes | 14.2 cm diameter | Detected a 4-cm tumor in a trial test with one patient | Not reported | Not reported | 1 |
| McMaster university[22] | 5 hours | 4.8 cm diameter | Reported a 247 mm resolution in simulation | No clinical trials | No clinical trials | 0 |
| Istanbul Technical University (SAFE)[23,24] | 20 minutes | Small and large | Detected lesions as small as 4 mm | 79% | 77% | 113 |
| University of Bristol (MARIA)[25] | 5 minutes | 32A and 32DD | Smallest recorded lesion was 5 mm | 76% | Not reported | 8 |
| University of Calgary (TSAR)[26] | < 30 minutes | B and C size | 10 mm lesion detected with SCR ~5.0 dB | No clinical trials | No clinical trials | 1 |
| McGill University[27] | 6 minutes | Amid A to D size | 10 mm spherical phantom tumor detected | No clinical trials | No clinical trials | 4 |
| Galway University (Wavelia)[28] | 6 minutes | Greater than 32B | Did not detect tumors < 10 mm in trials | 75% | Not reported | 5 |
| Chalmers University[29,30] | Not reported | 55.0-75.0 cm diameter | Used oil cups mimicking tumors (55-75 mm) | No clinical trials | No clinical trials | 0 |
| University of Perugia (Mammowave)[31] | 8 minutes | Various cup sizes | 73% | 82% | 353 | |
| Hiroshima University[32] | 15 minutes | Small (Japanese patients) | Minimum size 40 mm in trials (Japanese only) | 49% | 55% | 9 |
| University of São Paulo[33] | 28 minutes | 34B (imperial system) | 1 mm tumor in 150 mm breast phantom with SCR > 7.0 dB | No clinical trials | No clinical trials | 3 |
| University of Queensland[34] | 45 minutes | 12.4 cm | 20 mm × 30 mm water target in phantom | No clinical trials | No clinical trials | 1 |