Table 6 Changes in glycemic control before and after the A8+ TouchCare Nano automated insulin delivery system among prepubertal participants with type 1 diabetes mellitus (n = 9), mean ± SD
| Variable | Baseline (previous treatment) | After (A8+ TouchCare Nano AID) | Δ | SEM | P value1 |
| GMI% | 6.8 ± 0.2 | 6.5 ± 0.3 | -0.3 ± 0.4 | 0.1 | 0.0441 |
| HbA1c% | 7.4 ± 2.3 | 6.5 ± 0.2 | -1.0 ± 2.4 | 0.8 | 0.310 |
| TIR (70-180 mg/dL)% | 71.7 ± 9.4 | 82.4 ± 8.4 | 10.6 ± 5.9 | 2.2 | 0.0181 |
| TAR (> 180 mg/dL)% | 25.0 ± 7.5 | 14.6 ± 8.0 | -10.4 ± 3.5 | 1.3 | 0.0181 |
| Very high (> 250 mg/dL)% | 5.6 ± 5.0 | 4.1 ± 3.6 | -1.5 ± 3.1 | 1.2 | 0.271 |
| TBR (< 70 mg/dL)% | 3.3 ± 4.3 | 3.0 ± 2.2 | -0.3 ± 5.6 | 2.1 | 0.310 |
| Very low (< 56 mg/dL)% | 0.7 ± 1.3 | 0.3 ± 0.3 | -0.5 ± 1.4 | 0.5 | 0.892 |
| TITR (70-140 mg/dL)% | 47.8 ± 12.1 | 64.2 ± 10.3 | 16.4 ± 8.0 | 3 | 0.0181 |
| Blood glucose levels (mg/dL) | 148.4 ± 9.7 | 132.3 ± 13.2 | -16.1 ± 9.2 | 3.5 | 0.0271 |
| Average TDD (units) | 17.3 ± 9.5 | 23.3 ± 10.4 | 6.0 ± 11.9 | 6 | 0.273 |
- Citation: Kosta K, Grammatikopoulou MG, Toulia I, Smyrnaki P, Paschalidou EG, Vamvakis A, Bogiatzoglou A, Tzimos C, Tsiroukidou K. Patient satisfaction and performance of the A8+ TouchCare Nano in children and adolescents with type 1 diabetes mellitus. World J Clin Pediatr 2026; 15(3): 118236