Table 3 Comparison of measured respiratory rate between capnography and LRMD
SubjectCapnostream, mean ± SDLinshom, mean ± SDCapnostream–Linshom (95%CI)Pearson’s Correlation1 (95%CI)
114.9 ± 1.915.3 ± 2.1-0.46 (-0.50, -0.42)0.41 (0.40, 0.43)
415.6 ± 2.615.8 ± 3.3-0.21 (-0.25, -0.17)0.62 (0.61, 0.63)
525.1 ± 4.226.6 ± 2.9-1.4 (-1.5, -1.3)0.61 (0.60, 0.62)
67.9 ± 5.88.4 ± 5.2-0.52 (-0.56, -0.48)0.72 (0.72, 0.73)
723.1 ± 2.723.3 ± 4.4-0.21 (-0.28, -0.14)0.27 (0.25, 0.28)
913.2 ± 3.312.9 ± 3.40.37 (0.35, 0.39)0.84 (0.84, 0.84)
1015.7 ± 3.015.0 ± 3.60.76 (0.74, 0.78)0.76 (0.75, 0.76)
1115.8 ± 1.515.5 ± 1.70.26 (0.24, 0.27)0.65 (0.64, 0.65)