Table 7 IMRT/VMAT/tomotherapy three-dimensional dosimetry
| Ref. | Dosimeter | Read-out method | Treatment site | Delivery type | Photon energy/number of fields | Comparison on the plane or in volume | Gamma index criteria % dose difference (global/local)/DTA (mm)/threshold (%) | Gamma index passing rate |
| [86] | MAGIC | MRI/T2 | Model | IMRT/GKSRS | 6MV/Co-60 | Volume | 3/3/ | 50.3% |
| [87] | FX gel | OCT/Vista | Head and Neck | IMRT | 6MV | Volume | 3/3/none | 84.1% |
| [88] | BANG3 | MRI/T2 | Prostate | Tomotherapy | 6MV/Arc | Volume | 3/3 | 53% |
| [89] | PRESAGE | DLOS | Brain | IMRT | 6MV | Volume | 3/3 | 95% |
| [90] | MAGIC-f | MRI/T2 | Prostate | Tomotherapy | 6MV/Arc | Plane | 3/3 | 88.4% |
| [91] | nPAG | OCTOPUS-IQ | Prostate | IMRT | /7fields | Volume | 95.3% | |
| [10] | PAGAT | MRI/T2 | Pituitary | IMRT | 6MV/7fields | Volume | 2/2 | 99.4% |
| [33] | BANG3 | MRI/T2 | Prostate | VMAT (Elekta) | 6MV/Arc | Volume | 3 (global)/3/25 | 95.7% |
| [92] | BANG3 | MRI/T2 | Prostate | VMAT (Varian) | 6MV/Arc | Volume | 3 (global)/3/50 | 90.0% |
| [93] | PRESAGE | DMOS | Brain | IMRT | 6MV/5fields | Volume | 3 (global)/3/10 | 99.4% |
| [94] | NIPAM | MRI/T2 | Eye | IMRT | 6MV/5fields | Plane | 3/3/ | 98.5% |