Editorial
Copyright ©2012 Baishideng Publishing Group Co.
World J Radiol. Aug 28, 2012; 4(8): 345-352
Published online Aug 28, 2012. doi: 10.4329/wjr.v4.i8.345
Table 3 Magnetic resonance imaging protocol
EntryFeet firstPositionSupine
Coil type8 cardiac or 8 body upper8 cardiac or 8 body upper8 cardiac or 8 body upper8 cardiac or 8 body upper8 cardiac or 8 body upper8 cardiac or 8 body upper8 cardiac or 8 body upper
Series desc(1) Axial T2(2) ASSET calib(3) Axial T1 asset(4) 3D LAVA XV Dyn(5) Diffusion(6) 3D 5 min LAVA XV(7) Fiesta
Scan planeAxialAxialAxial preAxialAxialAxialCoronal
Image mode2D2D2D in/out phase3D2D3D2D
Pulse seqFSE-XLFast SPGRFast SPGRFast SPGRDW EPIFast SPGRFiesta
TE1/TE2 (ms)85FWIP/FWOPMin FullMin (~ 50-60)Min
TR/#R-R (ms)4000-6000150-2201200-1800
Flip angle8515NA1550-75
ETL (echo train length)16
RCV BW1 (kHz)41.6762.5083.33(125)83.3383-125
FOV (cm)34-444834-4434-4238-4434-4434-44
SCAN THK (mm)61054545
Spacing/#Loc000-2.00/36 slices-2.00 or -1
Freq × phase256 × 192256 × 192320 × (160-192)100 × 160320 × (160-192)192 × 356
NEX3-41162
Phase FOV0.75-1.01.00.9-1.00.75-1.01.00.70-0.900.75 –1.0
Freq dirR/LA/PR/LR/LR/LR/LS/I