Brief Article
Copyright ©2012 Baishideng Publishing Group Co.
World J Radiol. Jul 28, 2012; 4(7): 328-334
Published online Jul 28, 2012. doi: 10.4329/wjr.v4.i7.328
Table 1 Means, standard deviations, calculated difference between the measured serum creatinine level (mg/dL) using StatSensor™ creatinine and the laboratory reference, Pearson's correlation coefficients, paired t-test and Maloney-Rastogi test, subdivided by study centre, pre- and post-offset correction
Centre 1Centre 2Mean (centre 1 +2)Pearson correlation coefficientP value (paired t-test)P value (Maloney-Rastogi-test)
StatSensor™
cSCrPre offsetDevice A0.80 ± 0.340.89 ± 0.270.85 ± 0.31Inter-device evaluation (Device A vs B)
Device B0.81 ± 0.340.89 ± 0.260.85 ± 0.30Center 1 = 0.97Center 1 = 0.404Center 1 = 0.480
Difference between Device A vs B-0.01 ± 0.090.00 ± 0.09Center 2 = 0.95Center 2 = 0.689Center 2 = 0.521
Mean (Device A + B)0.80 ± 0.340.89 ± 0.260.85 ± 0.31
Post offsetDevice A0.97 ± 0.50
Device B0.96 ± 0.50
Laboratory reference
vSCr0.96 ± 0.421.02 ± 0.260.99 ± 0.35Inter-method evaluation (cSCr vs vSCr)
Difference
cSCr - vSCrPre offsetMean (Device A + B)-0.16 ± 0.16a-0.14 ± 0.15b-0.15 ± 0.15c< 0.0001a,b,c< 0.0001a,c; 0.9962b
Device A-0.16 ± 0.160.93< 0.0001< 0.0001
Device B-0.15 ± 0.170.92< 0.0001< 0.0001
Post offsetDevice A0.01 ± 0.190.930.638< 0.0001
Device B0.00 ± 0.200.920.955< 0.0001