World J Radiol. May 28, 2026; 18(5): 119048
Published online May 28, 2026. doi: 10.4329/wjr.v18.i5.119048
Table 4 Comparison of objective and subjective image quality parameters between abdominal photon-counting computed tomography and energy-integrating detector computed tomography, median (interquartile range)
Variable
PCCT
EID-CT
P value
meanliver65.95 (62.60-68.80)61.35 (55.77-65.08)0.001
SDliver11.60 (10.85-12.45)18.60 (16.62-20.32)< 0.001
meanfat-108.35 (-111.72 to -103.78)-106.15 (-110.78 to -99.08)0.192
SDfat11.75 (11.00-12.50)14.30 (13.03-16.35)< 0.001
SNR5.57 (5.02-6.32)3.26 (2.71-3.76)< 0.001
CNR10.26 (9.76-11.36)7.14 (6.21-7.67)< 0.001
EII71.99 (52.96-93.39)17.38 (12.71-20.75)< 0.001
SNR/(CDTI)1/22.90 (2.46-3.53)1.21 (0.97-1.47)< 0.001
CNR/(CDTI)1/25.46 (4.77-6.36)2.62 (2.36-2.98)< 0.001
NR1.00 (0.94-1.05)1.29 (1.10-1.48)< 0.001
Likert< 0.001
3129
487
5314