©Author(s) (or their employer(s)) 2026.
World J Radiol. Feb 28, 2026; 18(2): 115741
Published online Feb 28, 2026. doi: 10.4329/wjr.v18.i2.115741
Published online Feb 28, 2026. doi: 10.4329/wjr.v18.i2.115741
Table 6 Inter- and intra-observer agreements in the quantitative image quality and vessel length among three different sequences, mean ± SD
| Characteristic | Reader 1 | Reader 2 | Interobserver ICC (95%CI) | Reader 1 repeat | Intraobserver ICC (95%CI) |
| SNR | |||||
| LM | 114.4 ± 25.9 | 115.6 ± 23.0 | 0.957 (0.931-0.974) | 114.4 ± 24.0 | 0.989 (0.982-0.993) |
| LAD-pro | 110.8 ± 25.9 | 112.9 ± 30.7 | 0.966 (0.943-0.979) | 111.4 ± 23.7 | 0.985 (0.976-0.991) |
| LCX-pro | 107.4 ± 26.1 | 107.3 ± 22.4 | 0.956 (0.930-0.973) | 108.4 ± 23.6 | 0.986 (0.977-0.991) |
| RCA-pro | 103.6 ± 27.6 | 101.4 ± 26.6 | 0.975 (0.959-0.985) | 103.8 ± 24.6 | 0.989 (0.968-0.995) |
| CNR | |||||
| LM | 54.9 ± 24.2 | 57.2 ± 28.7 | 0.951 (0.918-0.970) | 55.9 ± 28.0 | 0.973 (0.956-0.983) |
| LAD-pro | 51.2 ± 24.1 | 51.1 ± 20.2 | 0.948 (0.916-0.968) | 52.1 ± 21.8 | 0.984 (0.973-0.990) |
| LCX-pro | 47.8 ± 25.1 | 50.1 ± 28.8 | 0.952 (0.920-0.971) | 48.8 ± 28.1 | 0.973 (0.957-0.984) |
| RCA-pro | 41.5 ± 26.3 | 43.3 ± 27.0 | 0.971 (0.951-0.982) | 43.3 ± 27.7 | 0.982 (0.967-0.990) |
| VES | |||||
| LM | 162.7 ± 76.6 | 155.7 ± 65.9 | 0.951 (0.917-0.971) | 159.6 ± 75.4 | 0.983 (0.972-0.989) |
| LAD-pro | 166.0 ± 79.8 | 159.0 ± 69.2 | 0.958 (0.927-0.975) | 162.9 ± 78.2 | 0.984 (0.974-0.990) |
| LCX-pro | 164.9 ± 86.5 | 161.5 ± 77.2 | 0.979 (0.966-0.987) | 161.2 ± 81.0 | 0.985 (0.975-0.991) |
| RCA-pro | 169.1 ± 88.9 | 162.0 ± 82.7 | 0.985 (0.964- 0.992) | 160.9 ± 84.6 | 0.989 (0.949-0.996) |
| Vessel length (mm) | |||||
| LAD | 119.2 ± 16.0 | 119.3 ± 16.0 | 0.997 (0.994-0.998) | 118.7 ± 16.2 | 0.999 (0.999-1.000) |
| LCX | 76.1 ± 9.1 | 75.9 ± 9.8 | 0.984 (0.975-0.990) | 76.2 ± 9.7 | 0.989 (0.982-0.993) |
| RCA | 121.4 ± 7.2 | 122.4 ± 7.2 | 0.978 (0.908-0.991) | 121.5 ± 7.2 | 0.996 (0.994-0.998) |
- Citation: Wei CW, Zhang RZ, Xu Y, Liu JY, Zhang N, Xu L, Wen ZY. Non-contrast 3.0T coronary magnetic resonance angiography: Image quality comparison of coronal balanced-turbo-field-echo, axial turbo-field-echo, and modified-Dixon sequences. World J Radiol 2026; 18(2): 115741
- URL: https://www.wjgnet.com/1949-8470/full/v18/i2/115741.htm
- DOI: https://dx.doi.org/10.4329/wjr.v18.i2.115741