World J Radiol. Feb 28, 2026; 18(2): 115741
Published online Feb 28, 2026. doi: 10.4329/wjr.v18.i2.115741
Published online Feb 28, 2026. doi: 10.4329/wjr.v18.i2.115741
Table 4 Quantitative image quality metrics and mean acquisition time for balanced turbo field echo, turbo field echo, and modified Dixon sequences, mean ± SD
| Parameter | BTFE | TFE | mDixon | Intergroups P value1 | P value (adjusted P value)2 | P value (adjusted P value)3 | P value (adjusted P value)4 |
| SNR | |||||||
| LM | 145.0 ± 12.9 | 88.68 ± 8.2 | 109.4 ± 12.2 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LAD-pro | 139.7 ± 13.6 | 83.2 ± 7.9 | 109.4 ± 12.6 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LCX-pro | 137.2 ± 14.5 | 80.4 ± 6.3 | 104.8 ± 12.5 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| RCA-pro | 135.4 ± 8.4 | 73.3 ± 9.2 | 94.3 ± 12.0 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| CNR | - | - | - | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LM | 83.7 ± 13.6 | 30.7 ± 6.4 | 50.2 ± 10.6 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LAD-pro | 78.2 ± 12.4 | 25.1 ± 5.8 | 50.1 ± 11.7 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LCX-pro | 75.7 ± 14.0 | 22.2 ± 7.1 | 45.5 ± 14.0 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| RCA-pro | 74.1 ± 12.1 | 15.4 ± 6.7 | 35.1 ± 8.6 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| VES5 | - | - | - | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LM | 257.3 ± 31.0 | 150.2 ± 19.3 | 80.5 ± 14.0 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LAD-pro | 266.1 ± 34.9 | 149.6 ± 14.2 | 82.1 ± 10.8 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| LCX-pro | 276.6 ± 33.9 | 141.1 ± 13.0 | 76.9 ± 7.4 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| RCA-pro | 284.1 ± 35.8 | 143.4 ± 13.5 | 79.7 ± 9.8 | < 0.001 | < 0.001 | < 0.001 | < 0.001 |
| Vessel length (mm) | |||||||
| LAD | 126 ± 10 | 114 ± 17 | 118 ± 17 | < 0.001 | < 0.001 | 0.016 | 0.017 |
| LCX | 77 ± 10 | 75 ± 8 | 76 ± 9 | 0.002 | 0.017 | 0.074 | 0.459 |
| RCA | 122 ± 7 | 120 ± 7 | 121 ± 7 | < 0.001 | 0.002 | 0.043 | 0.038 |
| Acquisition time (minute) | 6.13 ± 1.00 | 7.15 ± 2.02 | 7.17 ± 1.72 | 0.015 | 0.025 | 0.035 | 1.000 |