Table 1 Artifact evaluation on cone-beam computed tomography with original reconstruction and streak metal artifact reduction technique reconstruction with full volume of interest
| Artifact type | Original | SMART-Full | χ² | P value | ||
| N | Y | Total | ||||
| Radial | N | 51 | 1 | 52 | ||
| Y | 19 | 2 | 21 | |||
| Total | 70 | 3 | 73 | 14.45 | < 0.001 | |
| Streak | N | 69 | 0 | 69 | ||
| Y | 4 | 0 | 4 | |||
| Total | 73 | 0 | 73 | 2.25 | < 0.001 | |
| Thickening | N | 42 | 6 | 48 | ||
| Y | 14 | 11 | 25 | |||
| Total | 56 | 17 | 73 | 2.450 | 0.118 | |
| Dark streak | N | 2 | 0 | 2 | ||
| Y | 45 | 26 | 71 | |||
| Total | 47 | 26 | 73 | 43.022 | < 0.001 | |
- Citation: Wang ZL, Wu G. Application of streak metal artifact reduction technique in cone-beam computed tomography guided percutaneous transthoracic needle biopsy. World J Radiol 2026; 18(2): 115249