Copyright: ©Author(s) 2026.
World J Cardiol. Nov 26, 2026; 18(11): 125622
Published online Nov 26, 2026. doi: 10.4330/wjc.125622
Published online Nov 26, 2026. doi: 10.4330/wjc.125622
Table 2 A summary of the quoted technical specifications of current leadless pacing systems as of September 2026 based on regulatory submissions and supplier technical submissions
| Micra VR | Micra AV | Aveir AR | Aveir VR | Empower | EBR wise | |
| Width (mm) | 6.7 | 6.7 | 6.5 | 6.5 | 6 | 2.7 |
| Length (mm) | 25.9 | 25.9 | 32 | 38 | 31.9 | 9.1 |
| Fixation | Passive tine | Passive tine | Active helix | Active helix | Passive tine | Active spike |
| Sensor | Accelerometer | Accelerometer | Temperature | Temperature | Accelerometer | N/A |
| Delivery catheter French size | 23 | 23 | 25 | 25 | 21 | 12 |
| Maximum mode | VVIR | VDD | AAIR | VVIR | VVIR | BIVP |
| Co-implant | No | No | Can be DR | Can be DR | With S-ICD | Mandatory |
| Designed to retrieve | No | No | Yes | Yes | Yes | No |
| Battery projected maximum | 16 years | 16 years | 10 years | 20 years | 10 years | 4.5 years |
| FDA Approved | Yes | Yes | Yes | Yes | No | Yes |
| Commercial use | Yes | Yes | Yes | Yes | No | Yes |
| Indication | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing and anti-tachycardia ventricular pacing | Resynchronisation therapy |
| Paediatric use | Yes | Yes | Yes | Yes | No | No |