BPG is committed to discovery and dissemination of knowledge
Minireviews
Copyright: ©Author(s) 2026.
World J Cardiol. Nov 26, 2026; 18(11): 125622
Published online Nov 26, 2026. doi: 10.4330/wjc.125622
Table 2 A summary of the quoted technical specifications of current leadless pacing systems as of September 2026 based on regulatory submissions and supplier technical submissions

Micra VR
Micra AV
Aveir AR
Aveir VR
Empower
EBR wise
Width (mm)6.76.76.56.562.7
Length (mm)25.925.9323831.99.1
FixationPassive tinePassive tineActive helixActive helixPassive tineActive spike
SensorAccelerometerAccelerometerTemperatureTemperature AccelerometerN/A
Delivery catheter French size232325252112
Maximum mode VVIRVDDAAIRVVIRVVIRBIVP
Co-implantNoNoCan be DRCan be DRWith S-ICDMandatory
Designed to retrieveNoNoYesYesYesNo
Battery projected maximum16 years16 years10 years20 years10 years4.5 years
FDA ApprovedYesYesYesYesNoYes
Commercial useYesYesYesYesNoYes
IndicationBradycardia pacingBradycardia pacingBradycardia pacingBradycardia pacingBradycardia pacing and anti-tachycardia ventricular pacingResynchronisation therapy
Paediatric useYesYesYesYesNoNo


Write to the Help Desk