Case Report
Copyright ©The Author(s) 2015. Published by Baishideng Publishing Group Inc. All rights reserved.
World J Gastrointest Endosc. Apr 16, 2015; 7(4): 429-432
Published online Apr 16, 2015. doi: 10.4253/wjge.v7.i4.429
Bowel perforation due to break and distal passage of the safety ring of an adjustable intra-gastric balloon: A potentially life threatening situation
Ali M Al-Zubaidi, Hassan U Alghamdi, Abdu H Alzobydi, Irshad A Dhiloon, Laeeque A Qureshi
Ali M Al-Zubaidi, Hassan U Alghamdi, Abdu H Alzobydi, Irshad A Dhiloon, Laeeque A Qureshi, King Khalid Hospital, Najran 66262, Southern Province, Saudi Arabia
Laeeque A Qureshi, King Abdullah Medical City, Makkah 24246, Saudi Arabia
Author contributions: Al-Zubaidi AM placed and removed the balloon, the MRP who responsible for case and wrote the case report; Alghamdi HU did the laparotomy; Alzobydi AH participated in the laparotomy; Dhiloon IA done the case; Qureshi LA was peer reviewer.
Ethics approval: The case approved by the Hospital Board Ethical Committee.
Informed consent: Informed consent was taken from the patient.
Conflict-of-interest: The authors have no commercial associations that might be a conflict of interest in relation to this article.
Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Correspondence to: Dr. Ali M Al-zubaidi, Consultant Gastroenterologist, King Khalid Hospital, King Abdulaziz Road, Najran 66262, Southern Province, Saudi Arabia. dr_ali26@yahoo.com
Telephone: +966-59-8472448
Received: May 2, 2014
Peer-review started: May 3, 2014
First decision: May 23, 2014
Revised: January 1, 2015
Accepted: January 15, 2015
Article in press: January 19, 2015
Published online: April 16, 2015
Core Tip

Core tip: Because the rare reported unexpected complications that the balloon safety ring which designed to prevent its complication it was by it self the cause of serious complication. The u/s confirmation of balloon position was miss leading so radiographic images was essential when there is suspicious.