Table 3 Magnetically induced torque on the tested video capsule endoscopes
| Video capsule endoscope | Max magnetically induced torque (mNm) | Calculated upper bound of magnetically induced torque (mNm) | Worst-case torque due to gravity (mNm) | T mag < T grav |
| PillCam SB3 | 19.2 | Not determined | Approximately 0.76 | Not fulfilled |
| OMOM HD | 5.1 | Not determined | Approximately 0.82 | Not fulfilled |
| MiroCam 1200 | 41.7 | Not determined | Approximately 0.80 | Not fulfilled |
| CapsoCam Plus | 51.0 | Not determined | Approximately 1.11 | Not fulfilled |
| MiroCam 2000 | 54.0 | Not determined | Approximately 1.07 | Not fulfilled |
| PillCam Colon2 | 15.6 | Not determined | Approximately 0.93 | Not fulfilled |
| OMOM CC | 5.1 | Not determined | Approximately 0.97 | Not fulfilled |
| NaviCam Colon | Below measuring limit | Not determined | Approximately 0.86 | Fulfilled |
| NaviCam | Not determined | 271.95 | Approximately 1.23 | Not fulfilled |
| OMOM RC | Not determined | 289.02 | Approximately 1.75 | Not fulfilled |
| MiroCam Navi | Not determined | 293.90 | Approximately 1.19 | Not fulfilled |
- Citation: Beck MJ, Keuchel M, Paschen J, Laser RM, Roling J, Bandorski D, Windhorst AC, Kurniawan N, Baltes P. Magnetic resonance imaging-induced forces on video capsule endoscopes: First systematic in vitro assessment. World J Gastroenterol 2027; 33(1): 122550