Copyright: ©Author(s) 2027.
World J Gastroenterol. Jan 7, 2027; 33(1): 122550
Published online Jan 7, 2027. doi: 10.3748/wjg.122550
Published online Jan 7, 2027. doi: 10.3748/wjg.122550
Table 2 Calculated gravitational force and magnetically induced displacement force on the tested video capsule endoscopes
| Video capsule endoscope | Gravitational force (mN) | Magnetically induced displacement force (mN) | F mag < F grav |
| PillCam SB3 | 29.14 | 288.27 | Not fulfilled |
| OMOM HD | 31.27 | 47.91 | Not fulfilled |
| MiroCam 1200 | 32.87 | 571.96 | Not fulfilled |
| CapsoCam Plus | 36.56 | 672.66 | Not fulfilled |
| MiroCam 2000 | 35.65 | 575.92 | Not fulfilled |
| PillCam Colon2 | 28.88 | 299.57 | Not fulfilled |
| OMOM CC | 29.00 | 127.5 | Not fulfilled |
| NaviCam Colon | 27.67 | 21.96 | Fulfilled |
| NaviCam | 45.65 | 2232 | Not fulfilled |
| OMOM RC | 57.29 | 2366 | Not fulfilled |
| MiroCam Navi | 46.66 | 2415 | Not fulfilled |
- Citation: Beck MJ, Keuchel M, Paschen J, Laser RM, Roling J, Bandorski D, Windhorst AC, Kurniawan N, Baltes P. Magnetic resonance imaging-induced forces on video capsule endoscopes: First systematic in vitro assessment. World J Gastroenterol 2027; 33(1): 122550
- URL: https://www.wjgnet.com/1007-9327/full/v33/i1/122550.htm
- DOI: https://dx.doi.org/10.3748/wjg.122550