Table 1 Illustrative example of a perfusion MRI sequence performed on a 1.5 T MR platform
MRI platformAvanto (Siemens, Erlangen, Germany)
Type of pulse sequence3D FLASH
TR2.72 ms
TE1 ms
Partition thickness8 mm
Slices per slab10
Matrix256 × 159
Phase encode directionAnterior to posterior
Number of averages1
Sensitivity encoding factor2
Flip angle before contrast2º and 14º
Flip angle after contrast14º
Bandwidth490 Hz
RF spoilingYes
Temporal resolution1.98 s per slab of 10 slices.
Precontrast scans10 measurements of each flip angle averaged for calculation of native T1
Gadolinium injection0.2 mmol/kg at 3 mL/s followed by 20 mL flush
Patient respirationQuiet breathing
Post contrast scansA total of 180 consecutive measurements. Inject contrast only when the 20th measurement has been completed
Scan sections to use for processingCenter 6 image sections only