Table 1 Comparisons of conventional categories of nebulizers
JetUltrasonicVibrating mesh
Features
Power sourceCompressed gas or electrical mainsElectrical mainsBatteries or electrical mains
PortabilityRestrictedRestrictedPortable
Treatment timeLongIntermediateShort
Output rateLowHigherHighest
Residual volume0.8-2.0 mLVariable but low< 0.2 mL
Environmental contamination
Continuous useHighHighHigh
Breath-activatedLowLowLow
Performance variabilityHighIntermediateLow
Formulation characteristics
TemperatureDecreasesIncreasesMinimum change
ConcentrationIncreasesVariablesMinimum change
SuspensionsLow efficiencyPoor efficiencyVariable efficiency
DenaturationPossibleProbablePossible
CleaningRequired, after single useRequired, after multiple useRequired, after single use
CostVery lowHighHigh