Table 3 Strengths, current limitations, and future priorities in corneal nerve imaging
| Domain | Strengths | Current limitations | Future priorities |
| Imaging technique[18,39,40] | Rapid, noninvasive, and repeatable visualization of the corneal subbasal nerve plexus | Requires dedicated equipment and trained operators; image quality may be affected by focus, illumination, motion, tissue compression, and limited sampling area | Wider access to the technique; standardized operator training, acquisition protocols, image-quality criteria, and anatomical sampling strategies |
| Biological significance[24,25,29,30] | Direct quantitative evaluation of small sensory nerve fibers in an easily accessible tissue | Corneal nerve abnormalities are not disease-specific and similar morphometric changes may occur across different systemic, neurological, and ocular conditions | Interpretation of CCM findings within the appropriate systemic, neurological, and ocular clinical context |
| Quantitative analysis[15,21,22,39,45] | Objective morphometric assessment of CNFL, CNFD, CNBD, CTBD, IWL, and other structural nerve parameters | Considerable heterogeneity in devices, sampling strategies, image-selection procedures, parameter definitions, and manual, semiautomated, or automated analysis methods | Harmonization of acquisition, sampling, parameter definitions, image selection, analysis software, and reporting standards |
| Clinical utility[5,47,48,51] | Potential role in detecting subclinical small-fiber abnormalities, longitudinal follow-up, patient phenotyping, and assessment of nerve regeneration | Group-level differences do not necessarily translate into accurate classification of individual patients; correlations with neuropathy severity and established reference tests are variable | Prospective multicenter diagnostic-accuracy and longitudinal studies using prespecified thresholds, representative populations, and clinically meaningful patient-level outcomes |
| Relationship with established tests[47,50,51] | May provide complementary structural information alongside clinical examination, quantitative sensory testing, neurophysiological assessment, and skin biopsy | Agreement with skin biopsy and functional small-fiber assessments is incomplete; CCM may capture partially overlapping but distinct aspects of nerve injury and cannot currently replace established diagnostic tests | Define the incremental clinical value of CCM and its optimal position within multimodal and multidisciplinary diagnostic pathways |
| Ocular confounding factors[24,25,52-55] | Ocular history, slit-lamp examination, and ocular-surface assessment allow many local modifiers of corneal nerve morphology to be identified | Dry eye disease, primary corneal disease, contact lens-related changes, previous ocular surgery, corneal trauma or infection, and topical treatments may independently alter CCM measurements and confound attribution to systemic neuropathy | Prespecified ocular eligibility criteria; standardized ocular-surface assessment; documentation of contact lens use and previous ocular procedures; exclusion of major active corneal disease when appropriate; and stratified, adjusted, or sensitivity analyses for relevant ocular confounders |
| Automation and artificial intelligence[16,17,41,46,56,57] | Faster image analysis, reduced dependence on manual quantification, improved repeatability, and encouraging diagnostic-classification performance | Current evidence is derived largely from selected datasets; limited cross-device and multicenter validation, potential patient-level data leakage, class imbalance, domain shift, insufficient calibration, and limited model interpretability restrict clinical generalizability | Participant-level data separation; multicenter and cross-device external validation; diverse populations; transparent reporting; calibration and uncertainty estimation; interpretable outputs; and demonstration of incremental clinical value beyond conventional CCM morphometry |
| Reference standards and diagnostic thresholds[20,47,48,51] | Growing availability of normative datasets and quantitative reference values for major CCM parameters | Diagnostic thresholds remain insufficiently validated for broad patient-level clinical use and may vary according to population, device, anatomical region, analytical method, and reference standard | Expansion of representative normative datasets and prospective validation of clinically meaningful, device- and population-appropriate thresholds |