Table 1 Mean infusion rates for On-Q, Baxter and Ambu pain infusion devices over 10 h
Hour#1#2#3#4#5
On-Q11919.418.617.918.4
21818.418.416.818
317.618.417.916.117.5
417.81816.716.417.5
517.618.215.815.417.5
617.518.214.916.416.9
718.618.214.916.216.9
819.319.814.716.116.8
918.818.414.914.916.6
1018.51915.114.816.6
Baxter110.810.410.810.49.8
210.11010.610.310
3109.810.79.99.8
49.89.910.5109.7
59.79.510.11010
69.89.810.5109.8
79.69.610.49.89.9
810.59.810.4109.8
91010.110.3109.6
101010.110.710.19.6
Ambu19.59.2108.510.2
29.89.29.88.69.9
39.89.29.58.69.8
49.69.29.88.810.2
59.79.29.88.69.9
69.69.29.88.99.9
79.79.29.78.810.4
89.79.29.98.810.2
99.49.19.48.59.9
109.69.2108.79.8