BPG is committed to discovery and dissemination of knowledge
Retrospective Study
©The Author(s) 2025.
World J Gastrointest Surg. Oct 27, 2025; 17(10): 109407
Published online Oct 27, 2025. doi: 10.4240/wjgs.v17.i10.109407
Table 3 Comparison of catheter adjustment and radiation exposure between groups, n (%)
Parameter
Experimental group (n = 103)
Control group (n = 113)
Statistics
P value
Catheter adjustment
Number of adjustments (times), mean ± SD0.27 ± 0.461.45 ± 1.03t = 10.652< 0.001
Frequency of adjustmentsχ2 = 77.183< 0.001
0 time76 (73.8)21 (18.6)
1 time25 (24.3)40 (35.4)
≥ 2 times2 (1.9)52 (46.0)
Reasons for adjustment(n = 23)(n = 170)χ2 = 48.216< 0.001
Inaccurate position0 (0)128 (75.3)
Insufficient length15 (65.2)25 (14.7)
Catheter kinking8 (34.8)17 (10.0)
X-ray exposure
X-ray exposure time (seconds), mean ± SD5.3 ± 2.118.7 ± 4.5t = 26.419< 0.001
Number of fluoroscopies (times), mean ± SD1.1 ± 0.33.5 ± 0.8t = 29.783< 0.001
Patients requiring only 1 fluoroscopy93 (90.3)0 (0)χ2 = 168.297< 0.001
Patients requiring ≥ 3 fluoroscopies0 (0)59 (52.2)χ2 = 73.641< 0.001
Radiation dose
Average radiation dose (μGy·m2), mean ± SD14.8 ± 5.652.4 ± 12.3t = 28.937< 0.001
Operator scattered radiation dose (μSv/procedure), mean ± SD1.2 ± 0.54.3 ± 1.2t = 24.106< 0.001
Subgroup analysis (radiation-sensitive population)n = 15n = 17--
X-ray exposure time (seconds), mean ± SD4.7 ± 1.819.5 ± 5.2t = 10.342< 0.001
Radiation dose (μGy·m2), mean ± SD12.3 ± 4.754.8 ± 13.6t = 11.579< 0.001
Multivariate analysisOR95%CI-P value
Risk factors for multiple adjustments
Female sex2.131.25-3.64-0.005
BMI ≥ 25 kg/m22.861.67-4.91-< 0.001
Traditional anatomical landmark technique6.523.78-11.25-< 0.001
Protective factor for reduced X-ray exposure
Application of IC-ECG technique0.120.05-0.28-< 0.001


Write to the Help Desk