Table 5 The paired sensor glucose and capillary blood glucose measurements were stratified into five zones
| Zone | Percentage (%), n = 397 |
| Zone A (clinical accuracy) | 83.1 |
| Zone B (no or benign treatment) | 13.6 |
| Zone C (overcorrection) | 0.8 |
| Zone D (dangerous failure to detect and treat) | 2.5 |
| Failure to detect hyperglycemia | 1.0 |
| Failure to detect hypoglycemia | 1.5 |
| Zone E (erroneous measurement) | 0 |
- Citation: Sirinvaravong S, Sakjirapapong C, Preechasuk L, Thongtang N. Accuracy and feasibility of initiating real-time continuous glucose monitoring in noncritically ill hospitalized patients with type 2 diabetes. World J Diabetes 2026; 17(11): 124291