Copyright: ©Author(s) 2026.
World J Diabetes. Nov 15, 2026; 17(11): 124291
Published online Nov 15, 2026. doi: 10.4239/wjd.124291
Published online Nov 15, 2026. doi: 10.4239/wjd.124291
Table 4 Agreement between paired sensor glucose and capillary blood glucose measurements according to capillary blood glucose range
| Agreement level | ± 10/10% (mg/dL) | ± 15/15% (mg/dL) | ± 20/20% (mg/dL) | ± 30/30% (mg/dL) | ± 40/40% (mg/dL) |
| Total | 54.2% | 72.8% | 82.9% | 91.7% | 96.0% |
| 95%CI | 49.1-59.1 | 68.1-77.1 | 78.8-86.4 | 88.5-94.2 | 93.5-97.7 |
| < 70 mg/dL | 25.0% | 25.0% | 41.7% | 58.3% | 75% |
| 95%CI | 5.5-57.2 | 5.5-57.2 | 15.2-72.3 | 27.7-84.8 | 42.8-94.5 |
| 70-180 mg/dL | 52.5% | 72.1% | 81.6% | 90.5% | 96.1% |
| 95%CI | 44.9-60.0 | 64.9-78.5 | 75.1-87.0 | 85.2-94.4 | 92.1-98.4 |
| > 180 mg/dL | 57.3% | 76.2% | 86.4% | 94.7% | 97.1% |
| 95%CI | 50.2-64.1 | 69.8-81.9 | 81.0-90.8 | 90.6-97.3 | 93.8-98.9 |
- Citation: Sirinvaravong S, Sakjirapapong C, Preechasuk L, Thongtang N. Accuracy and feasibility of initiating real-time continuous glucose monitoring in noncritically ill hospitalized patients with type 2 diabetes. World J Diabetes 2026; 17(11): 124291
- URL: https://www.wjgnet.com/1948-9358/full/v17/i11/124291.htm
- DOI: https://dx.doi.org/10.4239/wjd.124291