Table 3 Mean absolute relative difference according to day of sensor wear
| Day | Paired values (n) | MARD (%), mean ± SD | ARD (%), median (Q1, Q3) | 95%CI |
| 1 | 67 | 15.7 ± 12.7 | 13.7 (5.8, 22.7) | 12.6-18.8 |
| 2 | 62 | 12.1 ± 8.4 | 11.0 (5.7, 16.5) | 10.0-14.3 |
| 3 | 58 | 9.6 ± 9.1 | 7.1 (3.2, 13.1) | 7.2-12.0 |
| 4 | 59 | 11.0 ± 11.4 | 8.3 (4.4, 14.2) | 8.1-14.0 |
| 5 | 44 | 12.6 ± 15.2 | 6.7 (3.1, 17.2) | 8.0-17.3 |
| 6 | 37 | 11.7 ± 12.4 | 7.6 (2.6, 14.3) | 7.5-15.8 |
| 7 | 28 | 13.5 ± 12.7 | 11.7 (3.9, 17.5) | 8.6-18.4 |
| 8 | 15 | 11.1 ± 11.1 | 8.1 (3.4, 11.4) | 5.0-17.3 |
| 9-11 | 27 | 11.3 ± 18.3 | 5.9 (1.7, 14.7) | 4.0-18.5 |
| Total | 397 | 12.2 ± 12.2 | 9.1 (3.7, 15.5) | 11.0-13.4 |
- Citation: Sirinvaravong S, Sakjirapapong C, Preechasuk L, Thongtang N. Accuracy and feasibility of initiating real-time continuous glucose monitoring in noncritically ill hospitalized patients with type 2 diabetes. World J Diabetes 2026; 17(11): 124291