Table 3 Steps in deployment of an over-the-scope clip
Steps
Ovesco clip (Ovesco Endoscopy)
Padlock (STERIS)
Set up the deviceAttach hand wheel to biopsy port; inspect cap/clip and release threadAttach external trigger/cable; inspect cap/clip; and confirm trigger function
Mount the cap on the endoscope and route actuationPreload clip on transparent cap; route release thread through the working channel to hand wheelPreload clip on cap; route actuation cable externally, leaving the working channel free
Prepare target siteIrrigate and adequately visualize the lesionSame as Ovesco
Orient correctlyCenter the lesion near the working-channel. Advance with cap over target; adjust torque or insufflation for a perpendicular approachSame as Ovesco
Capture tissueUse suction and/or Anchor or Twin Grasper; draw edges into capUse suction and grasping device via free channel to pull edges into cap more easily
Fire the clipRotate hand wheel to tension thread and deploy (similar to esophageal variceal band ligation mechanism)Squeeze external trigger to deploy; no channel thread
Verify closureConfirm apposition and hemostasis; irrigate or gently probe with a water jetSame as Ovesco
Know when to escalateIf closure or hemostasis incomplete, add thermal/adjunct clips or place a second OTSC; consider suturingSame as Ovesco